基于深度学习重建的高分辨率三维T2加权快速非对称自旋回波成像在桥小脑角肿瘤术前评估中的作用
本研究纳入13例连续患者,在3 T MRI上比较深度学习重建(DLR)与非DLR的高分辨率三维T2加权快速非对称自旋回波(HR-3D T2-FASE)成像对桥小脑角肿瘤的术前评估价值。DLR图像的肿瘤-三叉神经及脑脊液-三叉神经对比噪声比均显著高于非DLR图像,观察者对肿瘤、颅神经、血管及位置关系的显示偏好DLR更优。结论认为DLR HR-3D T2-FASE成像有助于桥小脑角肿瘤的术前评估。
要点
- DLR可提高HR-3D T2-FASE图像上肿瘤与三叉神经、脑脊液与三叉神经的对比噪声比。
- 观察者认为DLR对肿瘤、颅神经、血管及位置关系的显示优于非DLR图像。
- DLR与非DLR显示肿瘤周围面神经的准确例数分别为5例(38%)与4例(31%)。
核心结论
- DLR HR-3D T2-FASE成像有助于桥小脑角肿瘤的术前评估。
研究设计与样本
- 纳入13例连续患者,使用3 T MRI扫描仪行术前HR-3D T2-FASE成像,重建体素大小为0.23 × 0.23 × 0.5 mm,比较DLR与非DLR图像的对比噪声比,以术中所见为参考评估面神经与肿瘤的关系。
关键影像表现/征象
- 肿瘤与三叉神经的平均CNR在DLR图像为14.3 ± 8.9,高于非DLR图像的12.0 ± 7.6,P < 0.001。
- 脑脊液与三叉神经的平均CNR在DLR图像为66.4 ± 12.0,高于非DLR图像的53.9 ± 8.5,P < 0.001。
- 观察者对肿瘤显示与勾画的偏好DLR优于非DLR者占13例中的7例(54%)。
- 观察者对颅神经显示的偏好DLR优于非DLR者占13例中的6例(46%)。
- 观察者对血管显示及位置关系显示的偏好DLR优于非DLR者各占13例中的6例(46%)。
诊断标准与关键数值
- 肿瘤与三叉神经平均CNR:DLR图像14.3 ± 8.9 vs 非DLR图像12.0 ± 7.6,P < 0.001。
- 脑脊液与三叉神经平均CNR:DLR图像66.4 ± 12.0 vs 非DLR图像53.9 ± 8.5,P < 0.001。
- 以术中所见为参考,肿瘤周围面神经显示准确:DLR 5例(38%)vs 非DLR 4例(31%)。
临床应用与阅片要点
- 在桥小脑角肿瘤术前评估中,DLR HR-3D T2-FASE图像可提高肿瘤与三叉神经、脑脊液与三叉神经的对比噪声比。
- 观察者认为DLR图像对肿瘤、颅神经、血管及位置关系的显示优于非DLR图像。
- 作者认为DLR HR-3D T2-FASE成像有助于桥小脑角肿瘤的术前评估。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: We aimed to evaluate the effect of deep learning-based reconstruction (DLR) on high-spatial-resolution three-dimensional T2-weighted fast asymmetric spin-echo (HR-3D T2-FASE) imaging in the preoperative evaluation of cerebellopontine angle (CPA) tumors.
METHODS: This study included 13 consecutive patients who underwent preoperative HR-3D T2-FASE imaging using a 3 T MRI scanner. The reconstruction voxel size of HR-3D T2-FASE imaging was 0.23 × 0.23 × 0.5 mm. The contrast-to-noise ratios (CNRs) of the structures were compared between HR-3D T2-FASE images with and without DLR. The observers' preferences based on four categories on the tumor side on HR-3D T2-FASE images were evaluated. The facial nerve in relation to the tumor on HR-3D T2-FASE images was assessed with reference to intraoperative findings.
RESULTS: The mean CNR between the tumor and trigeminal nerve and between the cerebrospinal fluid and trigeminal nerve was significantly higher for DLR images than non-DLR-based images (14.3 ± 8.9 vs. 12.0 ± 7.6, and 66.4 ± 12.0 vs. 53.9 ± 8.5, P < 0.001, respectively). The observer's preference for the depiction and delineation of the tumor, cranial nerves, vessels, and location relation on DLR HR-3D T2FASE images was superior to that on non-DLR HR-3D T2FASE images in 7 (54%), 6 (46%), 6 (46%), and 6 (46%) of 13 cases, respectively. The facial nerves around the tumor on HR-3D T2-FASE images were visualized accurately in five (38%) cases with DLR and in four (31%) without DLR.
CONCLUSION: DLR HR-3D T2-FASE imaging is useful for the preoperative assessment of CPA tumors.
引用
- Hokamura M, Uetani H, Hamasaki T, et al. Effect of deep learning-based reconstruction on high-resolution three-dimensional T2-weighted fast asymmetric spin-echo imaging in the preoperative evaluation of cerebellopontine angle tumors. Neuroradiology. 2024. PMID: 38480538. doi:10.1007/s00234-024-03328-9
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。