放射科、心脏科和泌尿科的介入成像系统:能耗、碳排放和电费

Interventional Imaging Systems in Radiology, Cardiology, and Urology: Energy Consumption, Carbon Emissions, and Electricity Costs.
PubMed 文献 · AJR Am J Roentgenol · 2024-03-20

该研究为七个介入成像系统安装功率传感器,测量关机、空闲、活动及净成像状态功耗。结果显示净成像外能耗占预计总能耗93.3%,若将所有系统夜间和周末从空闲转为关机,每年可节省144,640 kWh、18.6吨CO2当量、37,896美元电费。

要点

  • 介入成像系统净成像外能耗占预计总能耗93.3%,为节能重点。
  • 夜间和周末将空闲系统关机,每年可节省144,640 kWh、18.6吨CO2当量、37,896美元。
  • 各系统关机状态较空闲状态平均功率降低0.2-6.9 kW(相对降低22.2-93.2%)。

核心结论

  • 介入成像系统能耗高,尤其在图像采集之外;策略性操作调整(如关闭空闲系统)可大幅降低碳排放和电费。

研究设计与样本

  • 对七个介入成像系统(介入放射学套房、神经介入套房、放射科透视单元、两个心脏科实验室、两个泌尿科透视单元)安装功率传感器;放射科和心脏科系统测量2022年6月1日至2022年11月28日间单次4周日志,泌尿科系统测量2023年7月31日至2023年8月13日间2周日志;从多种来源提取功率状态、手术时间戳和透视时间;使用瑞士已发表数值估算碳减排和电费节省。

临床应用与阅片要点

  • 原文指出策略性操作调整(如关闭空闲系统)可大幅降低介入成像系统的碳排放和电费。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND. The energy demand of interventional imaging systems has historically been estimated using manufacturer-provided specifications rather than directly measured. OBJECTIVE. The purpose of this study was to investigate the energy consumption of interventional imaging systems and estimate potential savings in the carbon emissions and electricity costs of such systems through hypothetical operational adjustments. METHODS. An interventional radiology suite, neurointerventional suite, radiology fluoroscopy unit, two cardiology laboratories, and two urology fluoroscopy units were equipped with power sensors. Power measurement logs were extracted for a single 4-week period for each radiology and cardiology system (all between June 1, 2022, and November 28, 2022) and for the 2-week period from July 31, 2023, to August 13, 2023, for each urology system. Power statuses, procedure time stamps, and fluoroscopy times were extracted from various sources. System activity was divided into off, idle (no patient in room), active (patient in room for procedure), and net-imaging (active fluoroscopic image acquisition) states. Projected annual energy consumption was calculated. Potential annual savings in carbon emissions and electricity costs through hypothetical operational adjustments were estimated using published values for Switzerland. RESULTS. Across the seven systems, the mean power draw was 0.3-1.1, 0.7-7.4, 0.9-7.6, and 1.9-12.5 kW in the off, idle, active, and net-imaging states, respectively. Across systems, the off state, in comparison with the idle state, showed a decrease in the mean power draw of 0.2-6.9 kW (relative decrease, 22.2-93.2%). The systems had a combined projected annual energy consumption of 115,684 kWh (range, 3646-26,576 kWh per system). The systems' combined projected energy consumption occurring outside the net-imaging state accounted for 93.3% (107,978/115,684 kWh) of projected total energy consumption (range, 89.2-99.4% per system). A hypothetical operational adjustment whereby all systems would be switched from the idle state to the off state overnight and on weekends (versus being operated in idle mode 24 hours a day, 7 days a week) would yield the following potential annual savings: for energy consumption, 144,640 kWh; for carbon emissions, 18.6 metric tons of CO2 equivalent; and for electricity costs, US$37,896. CONCLUSION. Interventional imaging systems are energy intensive, having high consumption outside of image acquisition periods. CLINICAL IMPACT. Strategic operational adjustments (e.g., powering down idle systems) can substantially decrease the carbon emissions and electricity costs of interventional imaging systems.

来源:AJR Am J Roentgenol,查看原文。版权归原作者所有。