微泡与微波消融的相互作用:体模与兔模型研究
本研究采用自制体模(体外)和白新西兰兔(体内)探索微泡与微波消融(MWA)的相互作用,结果显示微泡对MWA的消融范围和温度升高无显著影响,但MWA以温度依赖方式导致微泡破坏,微泡耐热性差是实时CEUS监测消融范围的不可忽视障碍。
要点
- 微泡对MWA消融范围和温度升高无显著影响。
- MWA以温度依赖方式导致微泡破坏。
- 微泡耐热性差是实时CEUS监测MWA消融范围的障碍。
核心结论
- 微泡对MWA无显著影响,但MWA以温度依赖方式诱导微泡破坏;微泡耐热性差是实时CEUS监测MWA消融范围的不可忽视障碍。
研究设计与样本
- 研究采用自制体模(体外)和白新西兰兔(体内),比较有无微泡时行MWA的差异;体模微泡浓度为2×10^5 particles mL^-1,兔肝内静脉注射0.05 mL/kg SonoVue。
- MWA期间使用K型热电偶探针监测温度升高,CEUS监测和分析微泡信号强度,MWA后评估并比较有无微泡组的消融区体积。
关键影像表现/征象
- 在填充微泡的体模和兔肝中,MWA随消融时间形成逐渐扩大的微泡缺损区。
- 温度升高导致微泡破坏。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: This study aimed to explore the interactions between microbubbles and microwave ablation (MWA).
METHODS: The study employed custom-made phantoms (in vitro) and white New Zealand rabbits (in vivo). MWA was performed with or without microbubbles in the phantoms (2 × 105 particles mL-1) and rabbit livers (intravenous injection of 0.05 mL/kg SonoVue). During the MWA, K-type thermocouple probes were used to monitor the MWA-induced temperature increase. Contrast-enhanced ultrasound imaging (CEUS) was used to monitor and analyze the microbubbles signal intensity. After MWA, the ablation-zone volumes were evaluated and compared between the groups with and without microbubbles.
RESULTS: In both the phantom models and rabbits, microbubbles showed no significant influence on MWA, including the ablation range and MWA-induced temperature increase. In phantoms and rabbit livers filled with microbubbles, MWA caused the formation of a gradually expanding microbubble-defect region over the ablation time. An increase in the temperature caused microbubble destruction.
CONCLUSIONS: Microbubbles had no significant influence on MWA. However, MWA induced the destruction of microbubbles in a temperature-dependent manner. Thus, the poor thermotolerance of microbubbles is a non-negligible barrier when using CEUS to monitor the ablation range during MWA in real-time.