三种多层螺旋CT高分辨率骨成像的定量与定性评价
研究比较三种多层螺旋CT系统的高分辨率采集策略,用体模评估几何精度、均匀性、扫描层几何和空间分辨率,并用人体干骨测试不同协议,由训练有素的CT用户进行盲法视觉分析。结果显示所有系统几何精度和均匀性满意,5%处的平面内MTF分别为13.4、15.9和18.1 lp/cm,干骨评估确认采集#3最佳。结论认为小焦点、高采样探测器的增量采集在高分辨率成像上超越了低螺距螺旋采集的参考。
要点
- 三种系统5%处平面内MTF分别为13.4、15.9和18.1 lp/cm。
- 干骨盲法评估中采集#3被认为最佳。
- 小焦点高采样探测器的增量采集在高分辨率成像上超越低螺距螺旋采集。
核心结论
- 小焦点、高采样探测器的增量采集在高分辨率成像上超越了低螺距螺旋采集的参考。
研究设计与样本
- 比较三种多层螺旋CT系统,用体模评估几何精度、均匀性和扫描层几何及空间分辨率,用人体干骨测试不同协议,并由训练有素的CT用户进行盲法视觉分析(p值)。
诊断标准与关键数值
- 5%处的平面内MTF分别为13.4、15.9和18.1 lp/cm。
- 干骨评估确认采集#3被认为最佳。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
INTRODUCTION: Strategies for achieving high resolution varies between manufacturers. In CT, the helical mode with narrow collimation has long been considered as the gold standard for high-resolution imaging. More recently, incremental modes with small dexels and focal spot, have been developed but have not been compared with helical acquisitions under optimal conditions. The aim of this work is to compare the high-resolution acquisition strategies currently proposed by recent MSCT.
METHODS: Three CT systems were compared. A phantom was used to evaluate geometric accuracy, uniformity, scan slice geometry, and spatial resolution. Human dry bones were used to test different protocols on real bone architecture. A blind visual analysis was conducted by trained CT users for classifying the different acquisitions (p-values).
RESULTS: All systems give satisfactory results in terms of geometric accuracy and uniformity. The in-plane MTF at 5% were respectively 13.4, 15.9 and 18.1 lp/cm. Dry-bones evaluation confirms that acquisition#3 is considered as the best.
CONCLUSIONS: The incremental acquisition coupled with à small focal spot, and a high-sampling detector, overpasses the reference of low-pitch helical acquisitions for high-resolution imaging. Cortical bone, bony vessels, and tumoral matrix analysis are the very next challenges that will have to be managed to improve normal and pathologic bone imaging thanks to the availability UHR-CT systems.