广角数字乳腺断层合成中钙化信号优化:一项虚拟成像试验
该虚拟成像试验在8例患者的MLO DBT采集上加虚拟微钙化模型,评估三种采集方案对微钙化可检测性的影响。结果显示标准与凸形剂量分布方案在DBT中无显著差异,稀疏投影方案显著降低DBT和合成2D乳腺摄影的可检测性,DBT阅读时间约为合成2D乳腺摄影的四倍。
要点
- 标准与凸形剂量分布方案在DBT中微钙化可检测性无显著差异。
- 稀疏投影方案(13个投影)显著降低DBT及合成2D乳腺摄影的微钙化可检测性。
- DBT阅读时间约为合成2D乳腺摄影的四倍。
核心结论
- 凸形剂量分布方案较标准方案未显著提高DBT微钙化可检测性,稀疏投影方案在DBT和合成2D乳腺摄影中均显著降低可检测性。
研究设计与样本
- 虚拟成像试验,采用8例患者的内外侧斜位DBT采集(双倍自动曝光控制剂量),通过自主研发虚拟成像框架添加虚拟微钙化模型,由6名观察者进行JAFROC研究评估可检测性。
诊断标准与关键数值
- DBT微钙化可检测性AUC:标准方案0.97±0.01,凸形方案0.95±0.02,稀疏方案0.89±0.03。
- 标准与凸形剂量分布方案间无显著差异(p=0.309)。
- 稀疏投影方案较标准方案显著降低可检测性(p=0.001)。
- 合成2D乳腺摄影在凸形方案下AUC较高,但无显著差异(p=0.435)。
临床应用与阅片要点
- 在系统研究所用条件下,使用默认投影数时标准或凸形剂量设置均可获得最佳微钙化簇检测。
- 稀疏投影方案(13个投影)会显著降低微钙化可检测性。
- DBT阅读时间约为合成2D乳腺摄影的四倍。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Evaluate microcalcification detectability in digital breast tomosynthesis (DBT) and synthetic 2D mammography (SM) for different acquisition setups using a virtual imaging trial (VIT) approach.
MATERIALS AND METHODS: Medio-lateral oblique (MLO) DBT acquisitions on eight patients were performed at twice the automatic exposure controlled (AEC) dose. The noise was added to the projections to simulate a given dose trajectory. Virtual microcalcification models were added to a given projection set using an in-house VIT framework. Three setups were evaluated: (1) standard acquisition with 25 projections at AEC dose, (2) 25 projections with a convex dose distribution, and (3) sparse setup with 13 projections, every second one over the angular range. The total scan dose and angular range remained constant. DBT volume reconstruction and synthetic mammography image generation were performed using a Siemens prototype algorithm. Lesion detectability was assessed through a Jackknife-alternative free-response receiver operating characteristic (JAFROC) study with six observers.
RESULTS: For DBT, the area under the curve (AUC) was 0.97 ± 0.01 for the standard, 0.95 ± 0.02 for the convex, and 0.89 ± 0.03 for the sparse setup. There was no significant difference between standard and convex dose distributions (p = 0.309). Sparse projections significantly reduced detectability (p = 0.001). Synthetic images had a higher AUC with the convex setup, though not significantly (p = 0.435). DBT required four times more reading time than synthetic mammography.
DISCUSSION: A convex setup did not significantly improve detectability in DBT compared to the standard setup. Synthetic images exhibited a non-significant increase in detectability with the convex setup. Sparse setup significantly reduced detectability in both DBT and synthetic mammography.
CLINICAL RELEVANCE STATEMENT: This virtual imaging trial study allowed the design and efficient testing of different dose distribution trajectories with real mammography images, using a dose-neutral protocol.
KEY POINTS: • In DBT, a convex dose distribution did not increase the detectability of microcalcifications compared to the current standard setup but increased detectability for the SM images. • A sparse setup decreased microcalcification detectability in both DBT and SM images compared to the convex and current clinical setups. • Optimal microcalcification cluster detection in the system studied was achieved using either the standard or convex dose setting, with the default number of projections.