腹部血管造影中的图像质量与辐射暴露:在多种源-影像接收器距离和准直下常规探测器剂量驱动与对比噪声比驱动曝光控制的头对头体模与动物初步研究
该体模与动物初步研究比较了探测器剂量驱动曝光控制(DEC)与对比噪声比驱动曝光控制(CEC)在不同源-影像接收器距离(SID)和准直下的图像质量与辐射暴露。结果显示CEC在变化SID和更严格准直下维持图像质量,较长SID时不增加空气比释动能率,更紧准直时降低空气比释动能率;达到空气比释动能率限值后两者表现相似。
要点
- CEC在较长SID和更紧准直下维持图像质量,不增加或降低空气比释动能率。
- DEC下CNR和空气比释动能率随SID延长和准直增加而显著增加。
- 达到空气比释动能率限值后,CEC与DEC表现相似。
核心结论
- CEC在变化SID和比DEC更严格准直下维持图像质量,较长SID时不增加空气比释动能率,更紧准直时降低空气比释动能率;达到空气比释动能率限值后CEC与DEC表现相似。
研究设计与样本
- 体模与动物初步研究:体模用铁箔模拟导丝,置于代表15、25、35 cm患者厚度的聚甲基丙烯酸甲酯与铝板堆叠中,采用5个SID(100至130 cm)和2种准直(全视野、6×6 cm准直视野)采集透视图像;动物研究在2只麻醉猪中创建5种血管造影场景,采用2个SID(110和130 cm)和两种准直采集透视图像,由两名盲法有经验读者以全视野SID 110 cm的参考图像比较图像质量。
诊断标准与关键数值
- 体模研究:铁箔CNR通过计算获得,辐射剂量以空气比释动能率表示并用多变量回归评估。
- 动物研究:两名盲法有经验读者将图像质量与全视野SID 110 cm的参考图像比较;空气比释动能率用t检验比较。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: This phantom and animal pilot study aimed to compare image quality and radiation exposure between detector-dose-driven exposure control (DEC) and contrast-to-noise ratio (CNR)-driven exposure control (CEC) as functions of source-to-image receptor distance (SID) and collimation.
MATERIALS AND METHODS: First, an iron foil simulated a guide wire in a stack of polymethyl methacrylate and aluminum plates representing patient thicknesses of 15, 25, and 35 cm. Fluoroscopic images were acquired using 5 SIDs ranging from 100 to 130 cm and 2 collimations (full field of view, collimated field of view: 6 × 6 cm). The iron foil CNRs were calculated, and radiation doses in terms of air kerma rate were obtained and assessed using a multivariate regression. Second, 5 angiographic scenarios were created in 2 anesthetized pigs. Fluoroscopic images were acquired at 2 SIDs (110 and 130 cm) and both collimations. Two blinded experienced readers compared image quality to the reference image using full field of view at an SID of 110 cm. Air kerma rate was obtained and compared using t tests.
RESULTS: Using DEC, both CNR and air kerma rate increased significantly at longer SID and collimation below the air kerma rate limit. When using CEC, CNR was significantly less dependent of SID, collimation, and patient thickness. Air kerma rate decreased at longer SID and tighter collimation. After reaching the air kerma rate limit, CEC behaved similarly to DEC. In the animal study using DEC, image quality and air kerma rate increased with longer SID and collimation ( P < 0.005). Using CEC, image quality was not significantly different than using longer SID or tighter collimation. Air kerma rate was not significantly different at longer SID but lower using collimation ( P = 0.012).
CONCLUSIONS: CEC maintains the image quality with varying SID and collimation stricter than DEC, does not increase the air kerma rate at longer SID and reduces it with tighter collimation. After reaching the air kerma rate limit, CEC and DEC perform similarly.
引用
- Werncke T, Becker LS, Maschke SK, et al. Image Quality and Radiation Exposure in Abdominal Angiography: A Head-to-Head Comparison of Conventional Detector-Dose-Driven Versus Contrast-to-Noise Ratio-Driven Exposure Control at Various Source-to-Image Receptor Distances and Collimations in a Pilot Phantom and Animal Study. Invest Radiol. 2024. PMID: 38529924. doi:10.1097/RLI.0000000000001079