原位校准珠作为背向散射系数计算参考的体内验证
研究在21只兔乳腺肿瘤中植入2 mm钛珠作为原位参考,评估其校准定量超声背向散射系数(BSC)并减轻衰减与传输损失的效果。结果显示,与外部参考体模相比,原位钛珠法在有无猪腹壁层覆盖时得到的有效散射体直径(ESD)和有效声浓度(EAC)估计偏差更小,提示该方法可提高QUS估计准确性。
要点
- 原位钛珠可同时作为影像标记和QUS校准参考,减少组织衰减与传输损失影响。
- 以钛珠校准较外部参考体模校准的ESD与EAC估计偏差更小。
- 该方法有望用于增强临床疾病状态分类的QUS估计准确性。
核心结论
- 原位钛珠作为放射性标记的同时可有效作为QUS校准靶,能考虑组织衰减与传输损失,获得更准确的QUS估计。
研究设计与样本
- 动物实验:在兔乳腺肿瘤(N=21)长至≥1 cm时植入2 mm钛珠,3天后用L-14/5 38阵列探头连接SonixOne扫描仪扫描,分别在有和无猪腹壁层覆盖条件下获取数据,比较外部参考体模与钛珠校准谱计算的ESD和EAC。
诊断标准与关键数值
- ESD估计在有无猪腹壁层间的95%CI:原位钛珠法为6.0至27.4,外部参考体模法为114至135.1;EAC估计95%CI:钛珠法为-8.1至0.5,体模法为-41.5至-32.2。
临床应用与阅片要点
- 作者认为原位钛珠作为校准靶可考虑用于提高QUS疾病状态分类的准确性。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: The study described here was aimed at assessing the capability of quantitative ultrasound (QUS) based on the backscatter coefficient (BSC) for classifying disease states, such as breast cancer response to neoadjuvant chemotherapy and quantification of fatty liver disease. We evaluated the effectiveness of an in situ titanium (Ti) bead as a reference target in calibrating the system and mitigating attenuation and transmission loss effects on BSC estimation.
METHODS: Traditional BSC estimation methods require external references for calibration, which do not account for ultrasound attenuation or transmission losses through tissues. To address this issue, we used an in situ Ti bead as a reference target, because it can be used to calibrate the system and mitigate the attenuation and transmission loss effects on estimation of the BSC. The capabilities of the in situ calibration approach were assessed by quantifying consistency of BSC estimates from rabbit mammary tumors (N = 21). Specifically, mammary tumors were grown in rabbits and when a tumor reached ≥1 cm in size, a 2 mm Ti bead was implanted in the tumor as a radiological marker and a calibration source for ultrasound. Three days later, the tumors were scanned with an L-14/5 38 array transducer connected to a SonixOne scanner with and without a slab of pork belly placed on top of the tumors. The pork belly acted as an additional source of attenuation and transmission loss. QUS parameters, specifically effective scatterer diameter (ESD) and effective acoustic concentration (EAC), were calculated using calibration spectra from both an external reference phantom and the Ti bead.
RESULTS: For ESD estimation, the 95% confidence interval between measurements with and without the pork belly layer was 6.0, 27.4 using the in situ bead and 114, 135.1 with the external reference phantom. For EAC estimation, the 95% confidence intervals were -8.1, 0.5 for the bead and -41.5, -32.2 for the phantom. These results indicate that the in situ bead method has reduced bias in QUS estimates because of intervening tissue losses.
CONCLUSION: The use of an in situ Ti bead as a radiological marker not only serves its traditional role but also effectively acts as a calibration target for QUS methods. This approach accounts for attenuation and transmission losses in tissue, resulting in more accurate QUS estimates and offering a promising method for enhanced disease state classification in clinical settings.