多能低千电子伏与单能低千伏图像在主动脉CT血管造影内漏检测中的比较
单中心前瞻性随机对照试验纳入125例EVAR术后CTA复查者,比较80kV单能图像(A组)与40/50keV多能虚拟单能图像(B组)的内漏检出能力;单能图像对比噪声比更高、主观图像质量更优,诊断准确性与50keV虚拟单能图像相当。
要点
- 80kV单能图像对比噪声比22.2,高于40keV的19.1(P=.048)和50keV的16.8(P<.001)。
- 内漏检出敏感度:80kV单能82.6%(19/23)、50keV 81.8%(18/22)高于40keV的77.3%(17/22)。
- 特异度组间相当(80kV 92.7%[38/41],两种虚拟单能能量均92.3%[35/38],P=.99),读者间一致度为1。
核心结论
- 与50keV虚拟单能图像相比,80kV单能图像在EVAR术后CTA内漏检测中图像质量更高、诊断准确性相当。
研究设计与样本
- 单中心前瞻性随机对照试验,2020年8月至2022年5月纳入EVAR术后CTA复查者125例(平均年龄76岁±8;103例男性),随机分为A组(80kV单能图像,n=64)和B组(多能CT 80/Sn150kV的40与50keV虚拟单能图像,n=61)。
诊断标准与关键数值
- 对比噪声比:80kV单能22.2 vs 40keV虚拟单能19.1(P=.048)vs 50keV虚拟单能16.8(P<.001);内漏检出敏感度:80kV单能82.6%(19/23)vs 40keV 77.3%(17/22)vs 50keV 81.8%(18/22);特异度:80kV单能92.7%(38/41)vs 两种虚拟单能能量均92.3%(35/38)(P=.99);读者间一致度为1;内漏明显度中位数:80kV单能2.99 vs 虚拟单能(两种能量)2.87(P=.04);整体主观图像质量中位数:80kV单能4(IQR 4-4)vs 40与50keV(两种能量)4(IQR 3-4)(P<.001)。
临床应用与阅片要点
- 作者结论认为80kV单能图像在EVAR术后CTA内漏检测中图像质量更高、诊断准确性与50keV虚拟单能图像相当。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Purpose To compare image quality, diagnostic performance, and conspicuity between single-energy and multi-energy images for endoleak detection at CT angiography (CTA) after endovascular aortic repair (EVAR). Materials and Methods In this single-center prospective randomized controlled trial, individuals undergoing CTA after EVAR between August 2020 and May 2022 were allocated to imaging using either low-kilovolt single-energy images (SEI; 80 kV, group A) or low-kiloelectron volt virtual monoenergetic images (VMI) at 40 and 50 keV from multi-energy CT (80/Sn150 kV, group B). Scan protocols were dose matched (volume CT dose index: mean, 4.5 mGy ± 1.8 [SD] vs 4.7 mGy ± 1.3, P = .41). Contrast-to-noise ratio (CNR) was measured. Two expert radiologists established the reference standard for the presence of endoleaks. Detection and conspicuity of endoleaks and subjective image quality were assessed by two different blinded radiologists. Interreader agreement was calculated. Nonparametric statistical tests were used. Results A total of 125 participants (mean age, 76 years ± 8; 103 men) were allocated to groups A (n = 64) and B (n = 61). CNR was significantly lower for 40-keV VMI (mean, 19.1; P = .048) and 50-keV VMI (mean, 16.8; P < .001) as compared with SEI (mean, 22.2). In total, 45 endoleaks were present (A: 23 vs B: 22). Sensitivity for endoleak detection was higher for SEI (82.6%, 19 of 23; P = .88) and 50-keV VMI (81.8%, 18 of 22; P = .90) as compared with 40-keV VMI (77.3%, 17 of 22). Specificity was comparable among groups (SEI: 92.7%, 38 of 41; both VMI energies: 92.3%, 35 of 38; P = .99), with an interreader agreement of 1. Conspicuity of endoleaks was comparable between SEI (median, 2.99) and VMI (both energies: median, 2.87; P = .04). Overall subjective image quality was rated significantly higher for SEI (median, 4 [IQR, 4-4) as compared with 40 and 50 keV (both energies: median, 4 [IQR, 3-4]; P < .001). Conclusion SEI demonstrated higher image quality and comparable diagnostic accuracy as compared with 50-keV VMI for endoleak detection at CTA after EVAR. Keywords: Aneurysms, CT, CT Angiography, Vascular, Aorta, Technology Assessment, Multidetector CT, Abdominal Aortic Aneurysms, Endoleaks, Perigraft Leak Supplemental material is available for this article. © RSNA, 2024.
引用
- Landsmann A, Sartoretti T, Mergen V, et al. Multi-Energy Low-Kiloelectron Volt versus Single-Energy Low-Kilovolt Images for Endoleak Detection at CT Angiography of the Aorta. Radiol Cardiothorac Imaging. 2024. PMID: 38451189. doi:10.1148/ryct.230217