更新希腊计算机断层扫描国家诊断参考水平:患者保护优化面临的挑战

Updating national diagnostic reference levels for computed tomography in Greece: Challenges on patient protection optimisation.
PubMed 文献 · Eur J Radiol · 2024-03-16

该文报告2019至2022年希腊全国CT剂量调查结果,收集120家医疗机构127台扫描仪(占全国CT扫描仪25%)的剂量学数据,用于修订成人CT检查的诊断参考水平(DRL)。修订后DRL的CTDIvol平均降低15%、扫描DLP平均降低7%,并新建立颈部、颈椎、骨盆-髋部、冠状动脉钙化评分及心脏CT血管造影的DRL,修订版DRL已在希腊正式采用。

要点

  • 修订后DRL的CTDIvol较初始DRL平均降低15%,扫描DLP平均降低7%。
  • 新建立颈部、颈椎、骨盆-髋部、冠状动脉钙化评分及CCTA的DRL。
  • 不同扫描仪间剂量学值和扫描长度差异大,需标准化与优化。

核心结论

  • 修订DRL的剂量学值降低幅度相对其他国家级研究较低,不同扫描仪间剂量学值和扫描长度差异大,强调标准化与优化的必要性。
  • 迭代重建(IR)等先进技术展示进一步降低剂量的潜力,但统一实施仍存挑战。
  • 研究强调在CT使用增加和技术演进背景下持续优化的重要性,修订版DRL已在希腊正式采用。

研究设计与样本

  • national survey conducted from 2019 to 2022,聚焦修订成人CT检查的DRL。
  • 收集120家医疗机构127台扫描仪(占全国CT扫描仪25%)的剂量学数据,强调地理分布和技术代表性。
  • DRL参数为感兴趣区典型采集的CTDIvol和DLP(扫描DLP)。

诊断标准与关键数值

  • 修订DRL的CTDIvol较初始DRL平均降低15%,扫描DLP平均降低7%。
  • 除修订7项CT检查的DRL值外,新建立颈部、颈椎、骨盆-髋部、冠状动脉钙化(Ca)评分及心脏CT血管造影(CCTA)检查的DRL。

临床应用与阅片要点

  • 修订版DRL已在希腊正式采用,体现对安全有效CT实践的承诺。
  • 不同扫描仪间剂量学值和扫描长度差异大,需标准化与优化。
  • 迭代重建(IR)等先进技术展示进一步降低剂量的潜力,但统一实施仍存挑战。

局限性

  • 修订DRL的剂量学值降低幅度相对其他国家级研究较低。
  • 不同扫描仪间剂量学值和扫描长度差异大。
  • 迭代重建(IR)等先进技术统一实施仍存挑战。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

The escalating use of Computed Tomography (CT) imaging necessitates establishment and periodic revision of Diagnostic Reference Levels (DRLs) to ensure patient protection optimization. This paper presents the outcomes of a national survey conducted from 2019 to 2022, focusing on revising DRLs for adult CT examinations. Dosimetric data from 127 scanners in 120 medical facilities, representing 25% of the country's CT scanners, were collected, emphasizing geographic distribution and technology representation. Τhe parameters used for DRLs were the CTDIvol and the DLP of a typical acquisition of the region of interest (scan DLP). In addition to the 7 CT examination for which the DRL values were revised, establishment of DRLs for neck, cervical spine, pelvic bones-hips, coronary artery calcium (Ca) score and cardiac computed tomography angiography (CCTA) examinations was performed. Revised DRLs exhibited a 15 % average decrease in CTDIvol and a 7 % average decrease in scan DLP from the initial DRLs. This reduction of dosimetric values is relatively low compared to other national studies. The findings revealed wide variations in dosimetric values and scan lengths among scanners, emphasizing the need for standardization and optimization. Incorporation of advanced technologies like Iterative Reconstruction (IR) showcased potential for further dose reduction, yet challenges in uniform implementation persist. The study underscores the importance of ongoing optimisation efforts, particularly in the context of increased CT utilization and evolving technology. The revised DRLs have been officially adopted in Greece, emphasizing the commitment to safe and effective CT practices.

来源:Eur J Radiol,查看原文。版权归原作者所有。