使用新型专用临床超声系统进行心脏二维剪切波成像:体模研究

Cardiac 2-D Shear Wave Imaging Using a New Dedicated Clinical Ultrasound System: A Phantom Study.
PubMed 文献 · Ultrasound Med Biol · 2024-03-12

本研究在弹性成像体模上评估新型MACH 30超声系统联合线阵与相控阵探头的心脏剪切波成像,并与已验证的Aixplorer系统比较。专用心脏序列在线阵探头下两系统平均百分比误差相似(8±14% vs 20±21%,p无显著差异),相控阵探头在4–10 cm深度内测量均匀,该区域外误差显著增大。

要点

  • 心脏剪切波成像最佳经胸深度为4–10 cm,此范围外平均百分比误差显著增大。
  • 线阵探头专用心脏序列的硬度测量精度在MACH 30与Aixplorer系统间无显著差异。
  • 相控阵探头在各测量区域均无平均百分比误差差异,无显著横向或轴向梯度。

核心结论

  • 配备相控阵探头和专用心脏设置的新型Supersonic MACH 30系统可提供均匀的硬度测量,尤其在4–10 cm深度操作时;这些体模结果可能有助于设计未来的在体研究。

研究设计与样本

  • 体模研究,使用Elasticity QA体模(Models 039和049,CIRS Inc.),比较新型MACH 30超声机(SuperSonic Imaging,Aix-en-Provence,France)联合线阵探头和相控阵探头与已验证的Aixplorer系统连接线阵探头(SuperSonic Imaging)的心脏剪切波成像。

诊断标准与关键数值

  • 线阵探头专用心脏序列平均百分比误差:MACH 30系统8±14% vs Aixplorer系统20±21%(p无显著差异)。
  • 经胸心脏探头最佳工作区位于4–10 cm,平均百分比误差29.5±25%,该区域外为93.3±130%(p<0.0001)。
  • 在4–10 cm区域内,硬度低于20 kPa与参考值有显著差异(p<0.0001)。
  • 相控阵探头在任何测量区域均无平均百分比误差差异,无显著横向或轴向梯度。

临床应用与阅片要点

  • 使用相控阵探头进行心脏剪切波成像时,应关注4–10 cm的最佳深度范围,该范围外测量误差显著增大。
  • 线阵探头专用心脏序列的硬度测量精度受目标硬度和测量点在视野中位置影响,但无随时间漂移。

局限性

  • 这些体模结果可能有助于设计未来的在体研究。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: The purpose of this study was to assess cardiac shear wave imaging implemented in a new MACH 30 ultrasound machine (SuperSonic Imaging, Aix-en-Provence, France) and interfaced with a linear probe and a phased array probe, in comparison with a previously validated Aixplorer system connected to a linear probe (SuperSonic Imaging) using Elasticity QA phantoms (Models 039 and 049, CIRS Inc., Norfolk, VA, USA).

METHODS: Quantile-quantile plots were used for distribution agreement. The accuracy of stiffness measurement was assessed by the percentage error and the mean percentage error (MPE), and its homogeneity, by the standard deviation of the MPE. A p value <0.01 was considered to indicate statistical significance.

RESULTS: The accuracy of dedicated cardiac sequences for linear probes was similar for the two systems with an MPE of 8 ± 14% versus 20 ± 21% (p = not significant) with the SuperSonic MACH 30 and Aixplorer, respectively, and was influenced by target stiffness and location of the measurement in the field of view, but without drift over time. The optimal transthoracic cardiac probe workspace was located between 4 and 10 cm, with an MPE of 29.5 ± 25% compared with 93.3 ± 130% outside this area (p < 0.0001). In this area, stiffness below 20 kPa was significantly different from the reference (p < 0.0001). The sectorial probe revealed no MPE difference in any of the measurement areas, with no significant lateral or axial gradient.

CONCLUSION: The new Supersonic MACH 30 system upgraded with a sectorial probe and specific cardiac settings provided homogenous stiffness measurements, especially when operating at depths between 4 and 10 cm. These phantom results may be useful in designing future in vivo studies.