利用治疗中皮肤表面及呼气末与吸气末计划CT图像预测实时肿瘤表面的模型

A model that predicts a real-time tumour surface using intra-treatment skin surface and end-of-expiration and end-of-inhalation planning CT images.
PubMed 文献 · Br J Radiol · 2024-01-01

研究利用呼气末(EOE)和吸气末(EOI)三维CT图像构建皮肤表面运动与内部肿瘤运动及形变之间的映射模型,并在15例患者的4D CT图像上验证。无模拟噪声时肿瘤表面平均中心位置误差1.29 mm、DSC 0.924、95%-Hausdorff距离2.76 mm、平均一致距离1.13 mm;有模拟噪声时分别为1.33 mm、0.920、2.79 mm、1.15 mm。作者提出并验证了一种仅用EOE和EOI三维CT图像及实时皮肤表面图像构建的患者特异性模型,可预测呼吸运动中的内部肿瘤运动与形变。

要点

  • 仅用EOE和EOI两套三维CT图像及实时皮肤表面图像即可构建患者特异性映射模型。
  • 无模拟噪声时肿瘤表面平均中心位置误差1.29 mm、DSC 0.924。
  • 有模拟噪声时平均中心位置误差1.33 mm、DSC 0.920。

核心结论

  • 提出并验证了一种患者特异性模型,仅用EOE和EOI三维CT图像及实时皮肤表面图像构建,可预测呼吸运动中的内部肿瘤运动与形变。

研究设计与样本

  • 治疗前从EOE和EOI三维CT图像分割重建皮肤与肿瘤表面,用非刚性配准算法将EOE皮肤和肿瘤表面配准到EOI得到位移矢量场并构建映射模型。

关键影像表现/征象

  • 治疗中将EOE皮肤表面配准到实时图像,得到实时皮肤表面位移矢量场。
  • 利用生成的映射模型,输入实时皮肤表面即可计算实时肿瘤表面。

诊断标准与关键数值

  • 无模拟噪声时肿瘤表面平均中心位置误差1.29 mm、DSC 0.924、95%-Hausdorff距离2.76 mm、平均一致距离1.13 mm。
  • 有模拟噪声时肿瘤表面平均中心位置误差1.33 mm、DSC 0.920、95%-Hausdorff距离2.79 mm、平均一致距离1.15 mm。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To develop a mapping model between skin surface motion and internal tumour motion and deformation using end-of-exhalation (EOE) and end-of-inhalation (EOI) 3D CT images for tracking lung tumours during respiration.

METHODS: Before treatment, skin and tumour surfaces were segmented and reconstructed from the EOE and the EOI 3D CT images. A non-rigid registration algorithm was used to register the EOE skin and tumour surfaces to the EOI, resulting in a displacement vector field that was then used to construct a mapping model. During treatment, the EOE skin surface was registered to the real-time, yielding a real-time skin surface displacement vector field. Using the mapping model generated, the input of a real-time skin surface can be used to calculate the real-time tumour surface. The proposed method was validated with and without simulated noise on 4D CT images from 15 patients at Léon Bérard Cancer Center and the 4D-lung dataset.

RESULTS: The average centre position error, dice similarity coefficient (DSC), 95%-Hausdorff distance and mean distance to agreement of the tumour surfaces were 1.29 mm, 0.924, 2.76 mm, and 1.13 mm without simulated noise, respectively. With simulated noise, these values were 1.33 mm, 0.920, 2.79 mm, and 1.15 mm, respectively.

CONCLUSIONS: A patient-specific model was proposed and validated that was constructed using only EOE and EOI 3D CT images and real-time skin surface images to predict internal tumour motion and deformation during respiratory motion.

ADVANCES IN KNOWLEDGE: The proposed method achieves comparable accuracy to state-of-the-art methods with fewer pre-treatment planning CT images, which holds potential for application in precise image-guided radiation therapy.