非慢性血流动力学条件下脑代谢的测量
在7头约克夏混合猪中模拟严重创伤性脑损伤(硬膜外血肿)与失血性休克(失血50%),用三维彩色血流成像(3DVF)测量颈动脉血流量并与有创流量计比较;3DVF相对有创流量计平均偏差-9.6%(均方根误差14.3%),结论提示可获取颈动脉绝对容积血流量,可能成为比现有替代标志物更特异的脑血流动力学生物标志物。
要点
- 3DVF测颈动脉血流量相对有创流量计平均偏差-9.6%,均方根误差14.3%。
- 颈动脉血流逆转期未检出显著的流量估计误差。
- 有创流量计的中位误差为-11.5%至39.7%。
核心结论
- 结果提示可获取通向大脑的颈动脉绝对容积血流量,并可能成为比现有替代标志物更特异的脑血流动力学生物标志物。
研究设计与样本
- 动物实验:7头约克夏混合猪(40-60 kg),模拟严重创伤性脑损伤(硬膜外血肿)及失血50%的失血性休克,用3DVF(4D16L,LOGIQ 9,GE HealthCare,Milwaukee,WI,USA)测量颈动脉血流量并与有创流量计(TS420,Transonic Systems Inc.,Ithaca,NY,USA)比较。
关键影像表现/征象
- 共实现6种生理状态:基线、血肿、基线2、失血性休克、失血性休克加血肿、失血后复苏。
- 平均脑氧摄取率从基线的40.6%±13.0%变化至失血性休克时的峰值68.4%±15.6%。
- 创伤条件下颈动脉可见血流逆转。
诊断标准与关键数值
- 3DVF相对有创流量计估计血流量的平均偏差为-9.6%,均方根误差为14.3%。
- 有创流量计的中位误差为-11.5%至39.7%。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Blood flow to the brain is a critical physiological function and is useful to monitor in critical care settings. Despite that, a surrogate is most likely measured instead of actual blood flow. Such surrogates include velocity measurements in the carotid artery and systemic blood pressure, even though true blood flow can actually be obtained using MRI and other modalities. Ultrasound is regularly used to measure blood flow and is, under certain conditions, able to provide quantitative volumetric blood flow in milliliters per minute. Unfortunately, most times the resulting flow data is not valid due to unmet assumptions (such as flow profile and angle correction). Color flow, acquired in three dimensions, has been shown to yield quantitative blood flow without any assumptions (3DVF).
METHODS: Here we are testing whether color flow can perform during physiological conditions common to severe injury. Specifically, we are simulating severe traumatic brain injury (epidural hematoma) as well as hemorrhagic shock with 50% blood loss. Blood flow was measured in the carotid artery of a cohort of 7 Yorkshire mix pigs (40-60 kg) using 3DVF (4D16L, LOGIQ 9, GE HealthCare, Milwaukee, WI, USA) and compared to an invasive flow meter (TS420, Transonic Systems Inc., Ithaca, NY, USA).
RESULTS: Six distinct physiological conditions were achieved: baseline, hematoma, baseline 2, hemorrhagic shock, hemorrhagic shock plus hematoma, and post-hemorrhage resuscitation. Mean cerebral oxygen extraction ratio varied from 40.6% ± 13.0% of baseline to a peak of 68.4% ± 15.6% during hemorrhagic shock. On average 3DVF estimated blood flow with a bias of -9.6% (-14.3% root mean squared error) relative to the invasive flow meter. No significant flow estimation error was detected during phases of flow reversal, that was seen in the carotid artery during traumatic conditions. The invasive flow meter showed a median error of -11.5% to 39.7%.
CONCLUSIONS: Results suggest that absolute volumetric carotid blood flow to the brain can be obtained and potentially become a more specific biomarker related to cerebral hemodynamics than current surrogate markers.