印度尼西亚计算机断层扫描仪性能评估:一项全国性研究的推荐意见

Performance evaluation of computed tomography scanners in Indonesia: recommendation from a nation-wide study.
指南 / 共识 · Radiat Prot Dosimetry · 2024-01-01

该研究评估了印度尼西亚684台CT扫描仪在2019-2022年测试周期的合规测试数据,旨在描述其性能概况并评价测试方案。87.8%的CT扫描仪无条件通过测试,8.8%有条件通过,3.4%未通过。有条件通过的前两位原因为水CT值准确性(45.2%)和激光定位准确性(41.9%);未通过设备的75.0%为未满足患者剂量测试标准。最新法规(BAPETEN第2/2022号法规)已纳入CT值和剂量测试参数的评估方法变更,并建议将激光定位准确性测试的通过标准改为绝对值1 mm。

要点

  • 87.8%的CT扫描仪无条件通过合规测试,8.8%有条件通过,3.4%未通过。
  • 有条件通过的前两位原因为水CT值准确性(45.2%)和激光定位准确性(41.9%)。
  • 未通过设备的75.0%因未满足患者剂量测试标准。

核心结论

  • 印度尼西亚CT扫描仪整体合规率较高,但水CT值准确性、激光定位准确性和患者剂量测试是主要失败环节,最新法规已调整前两者的评估方法,并建议将激光定位准确性通过标准改为绝对值1 mm。

研究设计与样本

  • 全国性合规测试数据评估,纳入印度尼西亚684台CT扫描仪,测试周期为2019-2022年。

诊断标准与关键数值

  • 87.8%的CT扫描仪无条件通过测试,8.8%有条件通过,3.4%未通过。
  • 有条件通过测试的设备中,水CT值准确性占45.2%,激光定位准确性占41.9%。
  • 未通过测试的设备中,75.0%为未满足患者剂量测试标准。
  • 水CT值准确性测试失败源于所用体模类型差异,部分体模的均匀性模块未使用水作为材料。
  • 激光定位准确性测试失败源于通过标准随最小层厚调整,导致小探测器尺寸和准直器的现代CT扫描仪往往无法通过。

临床应用与阅片要点

  • 最新法规(BAPETEN第2/2022号法规)已纳入CT值和剂量测试参数的评估方法变更。
  • 建议将激光定位准确性测试的通过标准改为绝对值1 mm。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

In this study, an evaluation of the compliance test data from 684 computed tomography (CT)-scanners in Indonesia for the 2019-22 test period was carried out. The study was aimed to describe the performance profile of CT-scanners in Indonesia and evaluate the testing protocol. A total of 87.8% of the CT-scanners unconditionally passed the tests, 8.8% passed the tests with conditions and 3.4% failed the tests. Of the devices conditionally passed the tests, the top two causes were water CT number accuracy (45.2%) and laser position accuracy (41.9%). Meanwhile, 75.0% of the failed devices were due to failing to meet the patient dose test criteria. The failure of the test for the water CT number accuracy parameter was caused by variations in the type of phantom used in the test, where several types of phantoms did not use water as material of the homogeneity module. Failures in laser position accuracy test were caused by the passing criteria that adjust to the minimum slice thickness, so that modern CT-scanner with small detector sizes and collimations tend not to pass. On the other hand, the failure on dose aspects was due to the frequent unavailability of baseline values for comparison. Of these top three failure causes, two of them, namely the CT number and dose test parameters, have been accommodated in the latest regulation (BAPETEN Regulation No. 2/2022) with a change in the evaluation method, while for the laser position accuracy test it is recommended to alter the passing criteria to an absolute value, namely 1 mm.