文献速递技术与安全CT

临床双源光子计数探测器CT单能衰减测量在不同扫描范式下的一致性:体模研究

Consistency of Monoenergetic Attenuation Measurements for a Clinical Dual-Source Photon-Counting Detector CT System Across Scanning Paradigms: A Phantom Study.
PubMed 文献 · AJR Am J Roentgenol · 2024-02-07

该体模研究评估了临床双源光子计数探测器CT在不同管电压和图像质量水平下获得的虚拟单能图像衰减测量的一致性。结果显示所有扫描范式与参考扫描的组内相关系数均≥0.999,平均偏差在120 kVp和70 keV时不超过5 HU,但在90 kVp、50 keV时部分组合偏差超过5 HU甚至10 HU。结论认为光子计数CT总体上提供了一致的衰减测量,有助于定量应用。

要点

  • 光子计数CT虚拟单能图像在多数扫描范式下衰减测量一致性良好。
  • 管电压120 kVp和140 kVp时平均偏差均≤5 HU。
  • VMI 70 keV时平均偏差均≤5 HU,为较稳健的重建能级。

诊断标准与关键数值

  • 所有扫描范式与参考扫描的衰减测量组内相关系数≥0.999。
  • 90 kVp管电压下平均偏差范围为-14.9至13.6 HU;120 kVp为-2.7至1.7 HU;140 kVp为-3.9至3.8 HU。
  • VMI 50 keV平均偏差-14.9至13.6 HU;60 keV为-6.4至3.8 HU;70 keV为-3.7至1.4 HU;80 keV为-7.2至4.3 HU。
  • 120 kVp和140 kVp下所有插入物平均偏差均不超过5 HU;70 keV下所有插入物平均偏差均不超过5 HU。
  • 90 kVp联合50 keV时,30种插入物与扫描范式组合中14种平均偏差超过5 HU,4种超过10 HU;60和80 keV下仅2种组合(均在90 kVp)平均偏差超过5 HU。

局限性

  • 本研究为体模研究,结果不能直接外推到人体扫描。
  • 仅评估了有限管电压和图像质量水平组合,未涵盖所有临床扫描参数。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND. Use of virtual monoenergetic images (VMIs) from multienergy CT scans can mitigate inconsistencies in traditional attenuation measurements that result from variation in scan-related factors. Photon-counting detector (PCD) CT systems produce VMIs as standard image output under flexible scanning conditions. OBJECTIVE. The purpose of this article was to evaluate the consistency of monoenergetic attenuation measurements obtained from a clinical PCD CT scanner across a spectrum of scanning paradigms. METHODS. A phantom with 10 tissue-simulating inserts was imaged using a clinical dual-source PCD CT scanner. Nine scanning paradigms were obtained across combinations of tube voltages (90, 120, and 140 kVp) and image quality (IQ) levels (80, 145, and 180). Images were reconstructed at VMI levels of 50, 60, 70, and 80 keV. Consistency of attenuation measurements was assessed, using the 120 kVp with IQ level of 145 scanning paradigm as the reference scan. RESULTS. For all scanning paradigms, attenuation measurements showed intra-class correlation of 0.999 and higher with respect to the reference scan. Across inserts, mean bias relative to the reference scan ranged from -14.9 to 13.6 HU, -2.7 to 1.7 HU, and -3.9 to 3.8 HU at tube voltages of 90, 120, and 140 kVp, respectively; and from -14.9 to 13.6 HU, -6.4 to 3.8 HU, -3.7 to 1.4 HU, and -7.2 to 4.3 HU at VMI levels of 50, 60, 70, and 80 keV, respectively. Thus, mean bias did not exceed 5 HU for any insert at tube potentials of 120 kVp and 140 kVp, nor for any insert at a VMI level of 70 keV. At a VMI level of 50 keV and tube potential of 90 kVp, mean bias exceeded 5 HU for 14 of 30 possible combinations of inserts and scanning paradigms and exceeded 10 HU for four of 30 such combinations. At VMI levels of both 60 and 80 keV, mean bias exceeded 5 HU for only two combinations of inserts and scanning paradigms, all at a tube potential of 90 kVp. CONCLUSION. PCD CT generally provided consistent attenuation measurements across combinations of scanning paradigms and VMI levels. CLINICAL IMPACT. PCD CT may facilitate quantitative applications of CT data in clinical practice.

来源:AJR Am J Roentgenol,查看原文。版权归原作者所有。