三维双回波稳态序列联合CAIPIRINHA两倍和四倍加速在膝关节软骨形态定量评估中的应用
64例受试者在3.0 T系统上接受两倍和四倍CAIPIRINHA加速三维双回波稳态序列及无并行加速技术三维双回波稳态序列扫描,两位肌骨放射科医师独立评估图像质量和诊断能力。两倍与四倍加速序列的软骨分级诊断一致性均良好,Kappa值分别为0.860和0.804(P<.001);软骨缺损敏感度分别为95.56%和91.49%,特异度分别为97.70%和97.65%;结论为CAIPIRINHA加速三维双回波稳态序列在扫描时间缩短60%后仍保持相当的诊断和分割性能。
要点
- 两倍加速三维CAIPIRINHA-DESS软骨分级诊断一致性Kappa值为0.860(P<.001),四倍加速为0.804(P<.001)。
- 两倍加速序列软骨缺损敏感度95.56%、特异度97.70%;四倍加速敏感度91.49%、特异度97.65%。
- 各序列软骨分割组内相关系数几乎均大于0.9。
核心结论
- CAIPIRINHA加速三维双回波稳态序列在扫描时间缩短60%后仍保持与无并行加速技术三维双回波稳态序列相当的膝关节软骨诊断和分割性能。
研究设计与样本
- 64例受试者在3.0 T系统上接受两倍和四倍CAIPIRINHA加速三维双回波稳态序列及无并行加速技术三维双回波稳态序列扫描,两位肌骨放射科医师在随机化和匿名化后独立评估图像质量和诊断能力,以无并行加速技术三维双回波稳态序列为软骨分级诊断标准。
关键影像表现/征象
- 图像信噪比和对比噪声比随扫描时间缩短而降低,同时伪影减少。
- 两倍和四倍CAIPIRINHA加速三维双回波稳态序列与无并行加速技术三维双回波稳态序列的软骨分级诊断一致性均良好。
诊断标准与关键数值
- 以无并行加速技术三维双回波稳态序列为软骨分级诊断标准,两倍加速三维CAIPIRINHA-DESS诊断一致性Kappa值为0.860(P<.001)。
- 四倍加速三维CAIPIRINHA-DESS诊断一致性Kappa值为0.804(P<.001)。
- 两倍加速三维CAIPIRINHA-DESS软骨缺损敏感度95.56%、特异度97.70%。
- 四倍加速三维CAIPIRINHA-DESS软骨缺损敏感度91.49%、特异度97.65%。
- 各序列软骨分割组内相关系数几乎均大于0.9。
临床应用与阅片要点
- CAIPIRINHA加速三维双回波稳态序列扫描时间缩短60%后仍可保持膝关节软骨的诊断和分割性能。
- 使用两倍与四倍加速序列时需注意图像信噪比和对比噪声比随扫描时间缩短而降低,同时伪影减少。
- 软骨分级诊断以无并行加速技术三维双回波稳态序列为标准时,两倍和四倍加速序列的一致性均良好。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: This study aimed to assess the diagnostic image quality and compare the knee cartilage segmentation results using a controlled aliasing in parallel imaging results in higher acceleration (CAIPIRINHA)-accelerated 3D-dual echo steady-state (DESS) research package sequence in the knee.
MATERIALS AND METHODS: A total of 64 subjects underwent both two- and fourfold CAIPIRINHA-accelerated 3D-DESS and DESS without parallel acceleration technique of the knee on a 3.0 T system. Two musculoskeletal radiologists evaluated the images independently for image quality and diagnostic capability following randomization and anonymization. The consistency of automatic segmentation results between sequences was explored using an automatic knee cartilage segmentation research application. The descriptive statistics and inter-observer and inter-method concordance of various acceleration sequences were investigated. P values < .05 were considered significant.
RESULTS: For image quality evaluation, the image signal-to-noise ratio and contrast-to-noise ratio decreased with the decrease in scanning time. However, it is accompanied by the reduction of artifacts. Using 3D-DESS without parallel acceleration technique as the standard for cartilage grading diagnosisand the diagnostic agreement of two- and fourfold CAIPIRINHA-accelerated 3D-DESS was good, kappa value was 0.860 (P < .001) and 0.804 (p < 0.001), respectively. Regarding cartilage defects, the sensitivity and specificity of the twofold acceleration 3D-CAIPIRINHA-DESS were 95.56% and 97.70%, and the fourfold CAIPIRINHA-accelerated 3D-DESS were 91.49% and 97.65%, respectively. The intraclass correlation coefficients of various sequences in cartilage segmentation were almost all greater than 0.9.
CONCLUSION: The CAIPIRINHA-accelerated 3D-DESS sequence maintained comparable diagnostic and segmentations performance of knee cartilage after a 60% scan time reduction.
引用
- Wen D, Zhou X, Hou B, et al. 3D-DESS MRI with CAIPIRINHA two- and fourfold acceleration for quantitatively assessing knee cartilage morphology. Skeletal Radiol. 2024. PMID: 38347270. doi:10.1007/s00256-024-04605-7
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。