利用3D打印结构评估彩色多普勒闪烁伪像的潜在成因
本研究用5微米分辨率3D打印六种不同沟槽几何形状的半圆柱体(螺距87.5–343微米、振幅41.5–209微米),置于脱气水中用彩色多普勒(3.1–6.3 MHz)扫描,发现两种标记出现闪烁:一种仅在标记上有预存气泡时闪烁(螺距/振幅比174/210微米/微米,0.8),另一种无预存气泡也闪烁(比值87/87微米/微米,1.00)。结论提示预存气泡与空化或超声波与图案化/粗糙表面的相互作用均是产生闪烁伪像的重要因素。
要点
- 螺距/振幅比为174/210微米/微米(0.8)的标记仅在存在预存气泡时出现闪烁。
- 螺距/振幅比为87/87微米/微米(1.00)的标记无预存气泡也出现闪烁。
- 预存气泡与空化或超声波与粗糙表面的相互作用均为闪烁伪像的重要成因。
核心结论
- 现有气泡与空化或超声波与图案化或粗糙表面的相互作用均是产生闪烁伪像的重要因素。
研究设计与样本
- 体外实验:用5微米分辨率微尺度3D打印机制备六种半圆柱体,每种标记含非闪烁对照共打印四份,螺距87.5–343微米、振幅41.5–209微米;脱气水中用GE Logiq E9扫描仪9L线阵探头以3.1–6.3 MHz彩色多普勒模式扫描。
关键影像表现/征象
- 螺距/振幅比为174/210微米/微米(0.8)的标记仅在标记上有预存气泡时出现闪烁。
- 螺距/振幅比为87/87微米/微米(1.00)的标记无预存气泡也出现闪烁。
- 光学成像用于在超声照射前后观察圆柱体上的气泡。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: The color Doppler twinkling artifact has been attributed to existing microbubbles or cavitation occurring on rough objects such as kidney stones, some breast biopsy clips, catheter guidewires and sandpaper. The objective was to investigate the correlation between the surface characteristics of helical constructs with different groove geometries and the occurrence of twinkling, as well as to identify locations conducive to bubble retention and/or cavitation.
METHODS: Six half-cylinders were created with a microscale 3-D printer with 5 µm resolution to replicate the geometry of twinkling helical constructs resembling catheter guidewires. Four copies of each marker including a non-twinkling control were printed. The half-cylinders had pitch (peak-to-peak distance) values ranging from 87.5 to 343 µm and amplitude (groove depth) values ranging from 41.5 to 209 µm. The half-cylinders were submerged in degassed water and optically imaged before and after ultrasound insonification to visualize bubbles on the cylinders. The cylinders remained submerged while scanning with the color Doppler mode at frequencies from 3.1 to 6.3 MHz using a GE Logiq E9 scanner and 9L linear array transducer.
RESULTS: Two markers exhibited twinkling: one with pitch-to-amplitude ratio of 174/210 µm/µm (0.8) that twinkled only with pre-existing bubbles on the marker; the other had a ratio of 87/87 µm/µm (1.00) that twinkled without pre-existing bubbles on the marker.
CONCLUSION: This work provides strong evidence that both existing bubbles and either cavitation or ultrasound wave interactions with patterned or rough surfaces are significant factors in producing the twinkling signature.