低剂量与常规CT尿路造影比较:双源CT不同时间电流乘积与相同管电压下的图像质量及多种疾病诊断效能
回顾性纳入357例双源CT尿路造影,比较常规剂量(280 mAs)与低剂量(70 mAs)在相同90 kVp及迭代重建下的图像质量与诊断效能。低剂量图像噪声显著更高、信噪比与对比噪声比更低,主观图像质量评分更低,但对泌尿系结石、炎症及肿块等疾病的诊断效能与常规剂量无差异。
要点
- 低剂量CT尿路造影(70 mAs、90 kVp、迭代重建)对泌尿系结石、炎症和肿块的诊断效能与常规剂量相当。
- 低剂量图像噪声显著增高、肝脏信噪比与对比噪声比显著降低,主观图像质量评分也显著低于常规剂量。
- 患者体重与噪声及有效辐射剂量呈正相关(P < .001)。
核心结论
- 尽管低剂量CT尿路造影图像质量较差,但采用70 mAs、90 kVp及迭代重建时,其对泌尿系多种疾病的诊断效能与常规CT等效。
研究设计与样本
- 回顾性研究,纳入2020年4月至2021年8月单中心连续357例第三代双源CT尿路造影,采用两期(平扫期、分团注射排泄期)扫描,常规剂量方案280 mAs、低剂量方案70 mAs,管电压均为90 kVp,两种方案均应用迭代重建。
关键影像表现/征象
- 低剂量图像噪声显著高于常规图像(P < .001),导致肝脏信噪比和对比噪声比显著降低(P < .001)。
- 两位放射科医师的主观定性分析均显示低剂量CT图像质量显著低于常规CT(P < .001)。
- 患者体重与噪声及有效辐射剂量呈正相关(P < .001)。
诊断标准与关键数值
- 低剂量方案为70 mAs、90 kVp并应用迭代重建,常规剂量方案为280 mAs、90 kVp并应用迭代重建。
- 对泌尿系结石、炎症和肿块等多种疾病的诊断效能,两种方案间无差异。
临床应用与阅片要点
- 低剂量CT尿路造影(70 mAs、90 kVp、迭代重建)对泌尿系结石、炎症和肿块的诊断效能与常规剂量相当,但图像噪声更高、主观图像质量更低。
- 患者体重与噪声及有效辐射剂量正相关(P < .001),阅片时需结合体重评估图像噪声对判读的潜在影响。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To compare the image quality and diagnostic performance of low-dose CT urography to that of concurrently acquired conventional CT using dual-source CT.
METHODS: This retrospective study included 357 consecutive CT urograms performed by third-generation dual-source CT in a single institution between April 2020 and August 2021. Two-phase CT images (unenhanced phase, excretory phase with split bolus) were obtained with two different tube current-time products (280 mAs for the conventional-dose protocol and 70 mAs for the low-dose protocol) and the same tube voltage (90 kVp) for the two X-ray tubes. Iterative reconstruction was applied for both protocols. Two radiologists independently performed quantitative and qualitative image quality analysis and made diagnoses. The correlation between the noise level or the effective radiation dose and the patients' body weight was evaluated.
RESULTS: Significantly higher noise levels resulting in a significantly lower liver signal-to-noise ratio and contrast-to-noise ratio were noted in low-dose images compared to conventional images (P < .001). Qualitative analysis by both radiologists showed significantly lower image quality in low-dose CT than in conventional CT images (P < .001). Patient's body weight was positively correlated with noise and effective radiation dose (P < .001). Diagnostic performance for various diseases, including urolithiasis, inflammation, and mass, was not different between the two protocols.
CONCLUSIONS: Despite inferior image quality, low-dose CT urography with 70 mAs and 90 kVp and iterative reconstruction demonstrated diagnostic performance equivalent to that of conventional CT for identifying various diseases of the urinary tract.
ADVANCES IN KNOWLEDGE: Low-dose CT (25% radiation dose) with low tube current demonstrated diagnostic performance comparable to that of conventional CT for a variety of urinary tract diseases.