商用0.55T系统腹部MRI的初步评估及与高场强的比较
研究纳入15名健康受试者和52例患者,在商用0.55T扫描仪行腹部MRI,并与1.5T/3T图像比较。0.55T所有序列在健康受试者中图像质量评为可接受,除部分序列外平均图像质量达诊断级;0.55T平均扫描时间54±10分钟 vs 高场强36±11分钟(P<.001)。结论认为0.55T可获得诊断质量腹部MR图像,但扫描时间更长,部分序列可能需进一步优化。
要点
- 0.55T腹部MRI所有序列在健康受试者中图像质量评为可接受,除动脉期FS 3D T1w GRE和3D MRCP对一位读者外平均图像质量达诊断级(>3分)。
- 与高场强相比,0.55T的IP 2D GRE和动脉期FS 3D T1w GRE评分显著较低,扩散加权EPI评分显著较高,其他序列等效。
- 0.55T平均扫描时间54±10分钟,高场强36±11分钟(P<.001)。
核心结论
- 商用0.55T扫描仪可获得诊断质量的腹部MR图像,但总采集时间较高场系统更长,部分序列可能受益于进一步优化。
研究设计与样本
- 前瞻性研究,纳入15名健康受试者和52例患者行0.55T腹部MRI,健康受试者另采集1.5T图像,52例患者中28例有1.5T/3T对比图像;两名放射科医师采用4分Likert量表评分图像质量并判断能否回答临床问题。
诊断标准与关键数值
- 图像质量采用4分Likert量表评分,平均图像质量>3分视为诊断级;0.55T所有序列在健康受试者中评为可接受;读者能回答临床问题的患者例数分别为读者1 52例、读者2 46例。
局限性
- 部分序列(动脉期FS 3D T1w GRE和3D MRCP)对一位读者平均图像质量未达诊断级(>3分),原文指出部分序列可能受益于进一步优化。
- 0.55T平均扫描时间54±10分钟较长,高于高场强36±11分钟(P<.001)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: This study aims to assess the quality of abdominal MR images acquired on a commercial 0.55T scanner and compare these images with those acquired on conventional 1.5T/3T scanners in both healthy subjects and patients.
MATERIALS AND METHODS: Fifteen healthy subjects and 52 patients underwent abdominal Magnetic Resonance Imaging at 0.55T. Images were also collected in healthy subjects at 1.5T, and comparison 1.5/3T images identified for 28 of the 52 patients. Image quality was rated by two radiologists on a 4-point Likert scale. Readers were asked whether they could answer the clinical question for patient studies. Wilcoxon signed-rank test was used to test for significant differences in image ratings and acquisition times, and inter-reader reliability was computed.
RESULTS: The overall image quality of all sequences at 0.55T were rated as acceptable in healthy subjects. Sequences were modified to improve signal-to-noise ratio and reduce artifacts and deployed for clinical use; 52 patients were enrolled in this study. Radiologists were able to answer the clinical question in 52 (reader 1) and 46 (reader 2) of the patient cases. Average image quality was considered to be diagnostic (>3) for all sequences except arterial phase FS 3D T1w gradient echo (GRE) and 3D magnetic resonance cholangiopancreatography for one reader. In comparison to higher field images, significantly lower scores were given to 0.55T IP 2D GRE and arterial phase FS 3D T1w GRE, and significantly higher scores to diffusion-weighted echo planar imaging at 0.55T; other sequences were equivalent. The average scan time at 0.55T was 54 ± 10 minutes vs 36 ± 11 minutes at higher field strengths (P < .001).
CONCLUSION: Diagnostic-quality abdominal MR images can be obtained on a commercial 0.55T scanner at a longer overall acquisition time compared to higher field systems, although some sequences may benefit from additional optimization.
引用
- Ramachandran A, Hussain HK, Gulani V, et al. Abdominal MRI on a Commercial 0.55T System: Initial Evaluation and Comparison to Higher Field Strengths. Acad Radiol. 2024. PMID: 38320946. doi:10.1016/j.acra.2024.01.018
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。