文献速递外周血管心血管技术与安全CT

低管电压全身CT血管成像联合极低碘剂量:混合迭代重建与深度学习图像重建算法的比较

Low-tube-voltage whole-body CT angiography with extremely low iodine dose: a comparison between hybrid-iterative reconstruction and deep-learning image-reconstruction algorithms.
PubMed 文献 · Clin Radiol · 2024-02-15

前瞻性研究纳入34例连续受试者,在80 kVp下行全身CTA,使用极低碘剂量(240 mg碘/ml对比剂与四分之一体积生理盐水同步给药,对应192 mg碘/ml),比较混合迭代重建(40%自适应统计迭代重建-Veo)、深度学习图像重建中等强度(DLIR-M)与高强度(DLIR-H)的图像质量。DLIR-H的动脉SNR最高,但DLIR-M在动脉显示与图像质量间平衡最佳,诊断可接受性优于混合迭代重建。

要点

  • 80 kVp全身CTA联合极低碘剂量下,DLIR-M在动脉显示与图像质量间平衡最佳。
  • 动脉SNR由高到低依次为DLIR-H、DLIR-M、混合迭代重建(p<0.001)。
  • DLIR-M与DLIR-H的诊断可接受性均显著优于混合迭代重建(p<0.001-0.005)。

核心结论

  • 与混合迭代重建和DLIR-H相比,DLIR-M显示出平衡良好的动脉显示与图像质量。

研究设计与样本

  • 前瞻性研究纳入34例连续受试者(27名男性;平均年龄74.2岁),于2020年1月至7月因评估主动脉疾病在80 kVp下行全身CTA。

诊断标准与关键数值

  • 对比剂为240 mg碘/ml,与四分之一体积生理盐水同步给药,对应192 mg碘/ml;平均对比剂用量64.1 ml,平均碘量15.4 g。
  • CT原始数据分别用40%自适应统计迭代重建-Veo(混合迭代重建)、DLIR中等强度(DLIR-M)和高强度(DLIR-H)重建。
  • 动脉SNR由高到低依次为DLIR-H、DLIR-M、混合迭代重建(p<0.001)。
  • 轴位图像上6条动脉、多平面重组图像上3条动脉、容积渲染图像上4条动脉的显示,DLIR-M或混合迭代重建显著优于DLIR-H(每项p≤0.009)。
  • 诊断可接受性DLIR-M和DLIR-H显著优于混合迭代重建(p<0.001-0.005)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

AIM: To evaluate arterial enhancement, its depiction, and image quality in low-tube potential whole-body computed tomography (CT) angiography (CTA) with extremely low iodine dose and compare the results with those obtained by hybrid-iterative reconstruction (IR) and deep-learning image-reconstruction (DLIR) methods.

MATERIALS AND METHODS: This prospective study included 34 consecutive participants (27 men; mean age, 74.2 years) who underwent whole-body CTA at 80 kVp for evaluating aortic diseases between January and July 2020. Contrast material (240 mg iodine/ml) with simultaneous administration of its quarter volume of saline, which corresponded to 192 mg iodine/ml, was administered. CT raw data were reconstructed using adaptive statistical IR-Veo of 40% (hybrid-IR), DLIR with medium- (DLIR-M), and high-strength level (DLIR-H). A radiologist measured CT attenuation of the arteries and background noise, and the signal-to-noise ratio (SNR) was then calculated. Two reviewers qualitatively evaluated the arterial depictions and diagnostic acceptability on axial, multiplanar-reformatted (MPR), and volume-rendered (VR) images.

RESULTS: Mean contrast material volume and iodine weight administered were 64.1 ml and 15.4 g, respectively. The SNRs of the arteries were significantly higher in the following order of the DLIR-H, DLIR-M, and hybrid-IR (p<0.001). Depictions of six arteries on axial, three arteries on MPR, and four arteries on VR images were significantly superior in the DLIR-M or hybrid-IR than in the DLIR-H (p≤0.009 for each). Diagnostic acceptability was significantly better in the DLIR-M and DLIR-H than in the hybrid-IR (p<0.001-0.005).

CONCLUSION: DLIR-M showed well-balanced arterial depictions and image quality compared with the hybrid-IR and DLIR-H.

来源:Clin Radiol,查看原文。版权归原作者所有。

引用

  1. Kawai N, Noda Y, Nakamura F, et al. Low-tube-voltage whole-body CT angiography with extremely low iodine dose: a comparison between hybrid-iterative reconstruction and deep-learning image-reconstruction algorithms. Clin Radiol. 2024. PMID: 38403540. doi:10.1016/j.crad.2024.02.002

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。