自由呼吸加速电影深度学习重建与标准屏气心脏MRI序列在缺血性心脏病患者中的前瞻性比较

Prospective Comparison of Free-Breathing Accelerated Cine Deep Learning Reconstruction Versus Standard Breath-Hold Cardiac MRI Sequences in Patients With Ischemic Heart Disease.
PubMed 文献 · AJR Am J Roentgenol · 2024-02-07

前瞻性纳入26例缺血性心脏病患者,比较自由呼吸深度学习重建电影序列与标准屏气平衡SSFP序列。 cine-DL序列采集时间更短(0.6±0.1 vs 2.4±0.6分钟,p<.001),LVEF无显著差异,主观图像质量更优,但模糊伪影更常见(42.3% vs 7.7%,p=.008),LVEDV在读者2中显著更大。

要点

  • 自由呼吸 cine-DL 序列采集时间显著短于标准屏气序列(0.6 vs 2.4 分钟,p<.001)。
  • 两序列 LVEF 无显著差异,cine-DL 序列主观图像质量更优。
  • cine-DL 序列模糊伪影更频繁(42.3% vs 7.7%,p=.008)。

核心结论

  • 自由呼吸 cine-DL 序列与标准屏气序列相比,LVEF 测量偏差很小且主观质量更好,但 cine-DL 序列测量的左心室容积更大。

研究设计与样本

  • 前瞻性研究,纳入2023年3月15日至2023年6月21日期间接受1.5T或3T心脏MRI评估缺血性心脏病的26例患者(平均年龄64.3±11.7岁,14男12女),比较研究性自由呼吸短轴 cine-DL 序列与标准屏气平衡SSFP序列,两名放射科医生盲法独立评估。

关键影像表现/征象

  • cine-DL 序列主观图像质量评分高于标准序列(读者1:2.3±0.5 vs 1.9±0.8,p=.02;读者2:2.2±0.4 vs 1.9±0.7,p=.02)。
  • 读者1报告 cine-DL 序列与标准序列的离共振伪影无显著差异(3.8% vs 23.1% examinations,p=.10),并行成像伪影亦无显著差异(3.8% vs 19.2%,p=.19)。
  • 读者1报告 cine-DL 序列模糊伪影较标准序列更频繁(42.3% vs 7.7% examinations,p=.008)。

诊断标准与关键数值

  • LVEF:读者1 cine-DL 序列 51.7%±14.3% vs 标准序列 51.3%±14.7%(p=.56);读者2 53.4%±14.9% vs 52.8%±14.6%(p=.53)。
  • LVEDV:读者1 cine-DL 序列 171.8±53.7 mL vs 标准序列 165.5±52.4 mL(p=.16);读者2 171.9±51.9 mL vs 160.6±49.4 mL(p=.01)。
  • LVESV:读者1 cine-DL 序列 88.1±49.3 mL vs 标准序列 86.0±50.5 mL(p=.45);读者2 85.2±48.1 mL vs 81.3±48.2 mL(p=.10)。
  • cine-DL 与标准序列的平均偏差:LVEF 读者1 0.4%、读者2 0.7%;LVEDV 读者1 6.3 mL、读者2 11.3 mL;LVESV 读者1 2.1 mL、读者2 3.9 mL。

临床应用与阅片要点

  • 对于无法反复屏气的缺血性心脏病患者,自由呼吸 cine-DL 序列可能提供可靠的 LVEF 测量。
  • cine-DL 序列测量的左心室容积较标准序列偏大,阅片时需注意。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND. Cine cardiac MRI sequences require repeated breath-holds, which can be difficult for patients with ischemic heart disease (IHD). OBJECTIVE. The purpose of the study was to compare a free-breathing accelerated cine sequence using deep learning (DL) reconstruction and a standard breath-hold cine sequence in terms of image quality and left ventricular (LV) measurements in patients with IHD undergoing cardiac MRI. METHODS. This prospective study included patients undergoing 1.5- or 3-T cardiac MRI for evaluation of IHD between March 15, 2023, and June 21, 2023. Examinations included an investigational free-breathing cine short-axis sequence with DL reconstruction (hereafter, cine-DL sequence). Two radiologists (reader 1 [R1] and reader 2 [R2]), in blinded fashion, independently assessed left ventricular ejection fraction (LVEF), left ventricular end-diastolic volume (LVEDV), left ventricular end-systolic volume (LVESV), and subjective image quality for the cine-DL sequence and a standard breath-hold balanced SSFP sequence; R1 assessed artifacts. RESULTS. The analysis included 26 patients (mean age, 64.3 ± 11.7 [SD] years; 14 men, 12 women). Acquisition was shorter for the cine-DL sequence than the standard sequence (mean ± SD, 0.6 ± 0.1 vs 2.4 ± 0.6 minutes; p < .001). The cine-DL sequence, in comparison with the standard sequence, showed no significant difference for LVEF for R1 (mean ± SD, 51.7% ± 14.3% vs 51.3% ± 14.7%; p = .56) or R2 (53.4% ± 14.9% vs 52.8% ± 14.6%; p = .53); significantly greater LVEDV for R2 (mean ± SD, 171.9 ± 51.9 vs 160.6 ± 49.4 mL; p = .01) but not R1 (171.8 ± 53.7 vs 165.5 ± 52.4 mL; p = .16); and no significant difference in LVESV for R1 (mean ± SD, 88.1 ± 49.3 vs 86.0 ± 50.5 mL; p = .45) or R2 (85.2 ± 48.1 vs 81.3 ± 48.2 mL; p = .10). The mean bias between the cine-DL and standard sequences by LV measurement was as follows: LVEF, 0.4% for R1 and 0.7% for R2; LVEDV, 6.3 mL for R1 and 11.3 mL for R2; and LVESV, 2.1 mL for R1 and 3.9 mL for R2. Subjective image quality was better for cine-DL sequence than the standard sequence for R1 (mean ± SD, 2.3 ± 0.5 vs 1.9 ± 0.8; p = .02) and R2 (2.2 ± 0.4 vs 1.9 ± 0.7; p = .02). R1 reported no significant difference between the cine-DL and standard sequences for off-resonance artifacts (3.8% vs 23.1% examinations; p = .10) and parallel imaging artifacts (3.8% vs 19.2%; p = .19); blurring artifacts were more frequent for the cine-DL sequence than the standard sequence (42.3% vs 7.7% examinations; p = .008). CONCLUSION. A free-breathing cine-DL sequence, in comparison with a standard breath-hold cine sequence, showed very small bias for LVEF measurements and better subjective quality. The cine-DL sequence yielded greater LV volumes than the standard sequence. CLINICAL IMPACT. A free-breathing cine-DL sequence may yield reliable LVEF measurements in patients with IHD unable to repeatedly breath-hold. TRIAL REGISTRATION. ClinicalTrials.gov NCT05105984.

来源:AJR Am J Roentgenol,查看原文。版权归原作者所有。