文献速递心血管CT

光子计数探测器CT钙去除图像重建算法对钙化冠状动脉病变诊断性能的提升

Diagnostic improvements of calcium-removal image reconstruction algorithm using photon-counting detector CT for calcified coronary lesions.
PubMed 文献 · Eur J Radiol · 2024-01-30

本研究回顾性纳入17例在60天内完成光子计数探测器冠状动脉CT血管成像(PCD-CCTA)和侵入性冠状动脉造影(ICA)的患者,评估162个钙化斑块节段。钙去除图像较常规图像主观图像质量更高(3.2±0.9 vs 2.7±0.5,p<0.001),基于节段的ROC曲线下面积(AUC)为0.87 vs 0.79(p=0.006),特异度(80.5% vs 69.5%,p=0.002)和阳性预测值(64.1% vs 52.0%,p<0.001)显著提高,但对比噪声比无显著差异。

要点

  • 钙去除图像可提高钙化斑块节段的主观图像质量并减少非诊断图像比例。
  • 钙去除图像诊断钙化冠状动脉病变的AUC、特异度和阳性预测值均优于常规图像。
  • 钙去除图像与常规图像的对比噪声比无显著差异。

核心结论

  • 光子计数探测器CT钙去除图像可能用于评估钙化冠状动脉病变的诊断性能。

研究设计与样本

  • 回顾性研究,纳入17例在CCTA后60天内行ICA的患者,评估162个钙化斑块节段,以ICA为参考标准。

关键影像表现/征象

  • 钙去除图像主观图像质量评分为3.2±0.9,常规图像为2.7±0.5(p<0.001)。
  • 钙去除图像非诊断图像质量节段百分比为28.3%,常规图像为32.7%(p<0.001)。
  • 钙去除图像平均对比噪声比为13.9±3.6,常规图像为13.3±3.4(p=0.356),两者无显著差异。

诊断标准与关键数值

  • 基于节段的ROC曲线下面积:钙去除图像0.87,常规图像0.79(p=0.006)。
  • 特异度:钙去除图像80.5%,常规图像69.5%(p=0.002)。
  • 阳性预测值:钙去除图像64.1%,常规图像52.0%(p<0.001)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To investigate the diagnostic performance of a calcium-removal image reconstruction algorithm with photon-counting detector-computed tomography (PCD-CT), a technology that hides only the calcified plaque from the spectral data in coronary calcified lesions.

METHODS: This retrospective study included 17 patients who underwent PCD-coronary CT angiography (CCTA) with at least one significant coronary stenosis (≥50 %) with calcified plaque by CCTA and invasive coronary angiography (ICA) performed within 60 days of CCTA. A total of 162 segments with calcified plaque were evaluated for subjective image quality using a 4-point scale. Their calcium-removal images were reconstructed from conventional images, and both images were compared with ICA images as the reference standard. The contrast-to noise ratios for both images were calculated.

RESULTS: Conventional and calcium-removal images had a subjective image quality of 2.7 ± 0.5 and 3.2 ± 0.9, respectively (p < 0.001). The percentage of segments with a non-diagnostic image quality was 32.7 % for conventional images and 28.3 % for calcium-removal images (p < 0.001). The segment-based diagnostic accuracy revealed an area under the receiver operating characteristic curve of 0.87 for calcium-removal images and 0.79 for conventional images (p = 0.006). Regarding accuracy, the specificity and positive predictive value of calcium-removal images were significantly improved compared with those of conventional images (80.5 % vs. 69.5 %, p = 0.002 and 64.1 % vs. 52.0 %, p < 0.001, respectively). The objective image quality of the mean contrast-to-noise ratio did not differ between the images (13.9 ± 3.6 vs 13.3 ± 3.4, p = 0.356) CONCLUSIONS: Calcium-removal images with PCD-CT can potentially be used to evaluate diagnostic performance for calcified coronary artery lesions.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Nishihara T, Miyoshi T, Nakashima M, et al. Diagnostic improvements of calcium-removal image reconstruction algorithm using photon-counting detector CT for calcified coronary lesions. Eur J Radiol. 2024. PMID: 38309215. doi:10.1016/j.ejrad.2024.111354

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。