文献速递心血管MR

超短VIBE-Dixon序列与电影及延迟钆增强MRI序列检测心脏血栓的对比

Ultrashort VIBE-Dixon versus Cine and late gadolinium enhancement MRI sequences for cardiac thrombus detection.
PubMed 文献 · Eur J Radiol · 2024-02-09

该研究回顾性纳入2014至2022年间141例经独立裁决委员会确诊或排除心脏血栓的患者,所有患者均接受2D电影+2D延迟钆增强(Cine+LGE)及VIBE-Dixon序列检查,由两名盲法读者评估。VIBE-Dixon平均检查时间28.8±3.1秒,Cine+LGE为23.3±2.5分钟;血栓患病率35%(49例)。两名读者VIBE-Dixon敏感度均显著高于Cine+LGE(读者1:96% vs 73%,读者2:96% vs 78%,P均<0.01),特异度无显著差异;VIBE-Dixon总体诊断准确率高于Cine+LGE(95% vs 89%,P=0.02),且非劣于参考标准(Delta≤5%,概率>95%)。

要点

  • VIBE-Dixon检测心脏血栓敏感度96%,显著高于Cine+LGE的73%–78%。
  • VIBE-Dixon特异度92%–96%,与Cine+LGE的93%–98%无显著差异。
  • VIBE-Dixon总体准确率95%,高于Cine+LGE的89%(P=0.02)。

核心结论

  • 双平面VIBE-Dixon序列可在数秒内完成采集,对心脏血栓检测具有很高的诊断准确率,可作为独立序列用于无法耐受长时间采集的患者快速筛查心脏血栓。

研究设计与样本

  • 回顾性研究,从临床数据库连续纳入2014至2022年间141例经独立裁决委员会确诊或排除心脏血栓的患者(116例男性,平均年龄61岁),所有患者均接受2D-Cine+2D-LGE及VIBE-Dixon序列检查,两名盲法读者评估图像,以独立裁决委员会诊断为参考标准。

诊断标准与关键数值

  • VIBE-Dixon检测心脏血栓敏感度读者1为96%、读者2为96%,Cine+LGE读者1为73%、读者2为78%,两读者VIBE-Dixon均显著高于Cine+LGE(P均<0.01)。
  • VIBE-Dixon检测心脏血栓特异度读者1为96%、读者2为92%,Cine+LGE读者1为98%、读者2为93%,两读者两种序列间无显著差异(P>0.1)。
  • VIBE-Dixon总体诊断准确率95%,Cine+LGE为89%(P=0.02)。
  • VIBE-Dixon诊断准确率非劣于参考标准(Delta≤5%,概率>95%)。
  • 心脏血栓患病率为35%(49例)。

临床应用与阅片要点

  • VIBE-Dixon平均图像检查时间28.8±3.1秒,Cine+LGE为23.3±2.5分钟。
  • 作者结论认为VIBE-Dixon可作为独立序列用于无法耐受长时间采集的患者快速筛查心脏血栓。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To determine the diagnostic accuracy of volumetric interpolated breath-hold examination sequences with fat suppression in Dixon technique (VIBE-Dixon) for cardiac thrombus detection.

METHOD: From our clinical database, we retrospectively identified consecutive patients between 2014 and 2022 who had definite diagnosis or exclusion of cardiac thrombus confirmed by an independent adjudication committee, serving as the reference standard. All patients received 2D-Cine plus 2D-Late-Gadolinium-Enhancement (Cine + LGE) and VIBE-Dixon sequences. Two blinded readers assessed all images for the presence of cardiac thrombus. The diagnostic accuracy of Cine + LGE and VIBE-Dixon was determined and compared.

RESULTS: Among 141 MRI studies (116 male, mean age: 61 years) mean image examination time was 28.8 ± 3.1 s for VIBE-Dixon and 23.3 ± 2.5 min for Cine + LGE. Cardiac thrombus was present in 49 patients (prevalence: 35 %). For both readers sensitivity for thrombus detection was significantly higher in VIBE-Dixon compared with Cine + LGE (Reader 1: 96 % vs.73 %, Reader 2: 96 % vs. 78 %, p < 0.01 for both readers), whereas specificity did not differ significantly (Reader 1: 96 % vs. 98 %, Reader 2: 92 % vs. 93 %, p > 0.1). Overall diagnostic accuracy of VIBE-Dixon was higher than for Cine + LGE (95 % vs. 89 %, p = 0.02) and was non-inferior to the reference standard (Delta ≤ 5 % with probability > 95 %).

CONCLUSIONS: Biplanar VIBE-Dixon sequences, acquired within a few seconds, provided a very high diagnostic accuracy for cardiac thrombus detection. They could be used as stand-alone sequences to rapidly screen for cardiac thrombus in patients not amenable to lengthy acquisition times.

来源:Eur J Radiol,查看原文。版权归原作者所有。