孤立性肺病变的体素内不相干运动扩散加权成像:梯度自旋回波序列的初步研究
本研究在3 T下对42例孤立性肺病变患者行TSE-IVIM、GRASE-IVIM和EPI-IVIM扫描,比较图像质量与参数可重复性。EPI-IVIM信号比更高、对比比更低、变形比更高(p<0.05),TSE-IVIM与GRASE-IVIM间无显著差异(p>0.05)。GRASE-IVIM降低变形比、提高ADC和D值重复扫描稳定性,且扫描时间最短。
要点
- GRASE-IVIM的变形比低于EPI-IVIM,图像变形更小
- TSE-IVIM的D和f值可重复性较好(ICC下限>0.90),GRASE-IVIM和EPI-IVIM可重复性较差(ICC下限<0.90)
- GRASE-IVIM的ADC和D值重复性(CV 2.56-3.12%和3.21-3.51%)优于EPI-IVIM(CV 10.03-10.2%和11.30-11.57%)
核心结论
- GRASE-IVIM降低变形比、改善ADC和D值重复扫描稳定性,且扫描时间最短。
研究设计与样本
- 前瞻性可行性研究,纳入42例孤立性肺病变患者,在3 T下行TSE-IVIM、GRASE-IVIM和EPI-IVIM扫描。
关键影像表现/征象
- EPI-IVIM的信号比高于TSE-IVIM和GRASE-IVIM,对比比低于二者,变形比高于二者(p<0.05)。
- TSE-IVIM与GRASE-IVIM在信号比、对比比和变形比上无显著差异(p>0.05)。
诊断标准与关键数值
- TSE-IVIM的D和f值可重复性较好(ICC下限>0.90),GRASE-IVIM和EPI-IVIM可重复性较差(ICC下限<0.90)。
- TSE-IVIM的ADC值CV为1.93-2.96%,D值CV为2.44-3.18%;GRASE-IVIM的ADC值CV为2.56-3.12%,D值CV为3.21-3.51%;EPI-IVIM的ADC值CV为10.03-10.2%,D值CV为11.30-11.57%。
- 所有序列的D*和f值重复性均较差。
- Bland-Altman分析显示所有序列的ADC和IVIM衍生参数间一致性界限较宽。
临床应用与阅片要点
- GRASE-IVIM可减少图像变形并提高ADC和D值重复扫描稳定性,且扫描时间最短,可能适合用于孤立性肺病变的IVIM评估。
- EPI-IVIM图像变形较大,ADC和D值重复性较差,解读其参数时需注意。
- TSE-IVIM的D和f值可重复性较好,但扫描时间较长。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
AIM: To evaluate the feasibility and image quality of intravoxel incoherent motion diffusion-weighted imaging (IVIM) using gradient- and spin-echo (GRASE) in solitary pulmonary lesions (SPLs) compared to echo planar imaging (EPI) and turbo spin-echo (TSE) at 3 T.
MATERIALS AND METHODS: Forty-two patients with SPLs underwent lung magnetic resonance imaging (MRI) using TSE-IVIM, GRASE-IVIM, and EPI-IVIM at 3 T. Signal ratio (SR), contrast ratio (CR), and image distortion ratio (DR) of three sequences were compared. The reproducibility and repeatability of the apparent diffusion coefficient (ADC) and IVIM-derived parameters were assessed using the interclass correlation coefficient (ICC) and coefficient of variation (CV). The repeatability of the ADC and IVIM-derived parameters between all sequences was evaluated using the Bland-Altman method.
RESULTS: EPI-IVIM had a higher SR, lower CR, and higher DR (p<0.05); however, there was no significant difference between TSE-IVIM and GRASE-IVIM (p>0.05). Compared to the D and f values of TSE-IVIM (ICC lower limit >0.90), GRASE-IVIM and EPI-IVIM showed poor reproducibility (ICC lower limit<0.90). The repeatability of the ADC and D values obtained by TSE-IVIM (CV, 1.93-2.96% and 2.44-3.18%, respectively) and GRASE-IVIM (CV, 2.56-3.12% and 3.21-3.51%, respectively) were superior to those of EPI-IVIM (CV, 10.03-10.2% and 11.30-11.57%). The repeatability of D∗ and f values for all sequences was poor. Bland-Altman analysis showed wide limits of agreement between the ADC and IVIM-derived parameters for all sequences.
CONCLUSION: GRASE-IVIM reduced the DR, improved the stability of the ADC and D values on repeated scans, and had the shortest scanning time.
引用
- Yin M, Cao G, Lv S, et al. Intravoxel incoherent motion diffusion-weighted imaging of solitary pulmonary lesions: initial study with gradient- and spin-echo sequences. Clin Radiol. 2024. PMID: 38307815. doi:10.1016/j.crad.2024.01.003
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。