文献速递儿科影像胸部CT

儿科胸部计算机断层扫描成像中z轴过扫描的射束准直对晶状体和甲状腺剂量的影响

Impact of beam collimation of z-overscanning on dose to the lens and thyroid gland in paediatric thoracic computed tomography imaging.
PubMed 文献 · Pediatr Radiol · 2024-02-03

研究使用新生儿和5岁儿童拟人 phantom,在儿科胸部螺旋CT中比较40 mm与80 mm准直对晶状体和甲状腺所致成像体积外剂量的影响。结果显示两种 phantom 中两种准直的成像体积外剂量(晶状体和甲状腺)差异均有统计学意义(P < 0.01)。结论认为,与80 mm准直相比,40 mm准直下新生儿和5岁儿童的成像体积外剂量(晶状体和甲状腺)显著更低。

要点

  • 儿科胸部螺旋CT中,40 mm准直较80 mm准直降低晶状体和甲状腺的成像体积外剂量。
  • 新生儿与5岁儿童 phantom 中,两种准直的成像体积外剂量差异均有统计学意义(P < 0.01)。
  • 自适应准直可减少沿z轴成像体积外沉积的剂量。

核心结论

  • 与80 mm准直相比,40 mm准直下新生儿和5岁儿童胸部螺旋CT扫描的成像体积外剂量(晶状体和甲状腺)显著更低。

研究设计与样本

  • 采用新生儿和5岁儿童拟人 phantom,在螺旋CT下比较40 mm与80 mm准直,使用光释光剂量计(OSLD)测量晶状体和甲状腺表面成像体积外剂量并比较图像噪声。

诊断标准与关键数值

  • 新生儿和5岁儿童 phantom 中,40 mm与80 mm准直的成像体积外剂量(晶状体和甲状腺)差异均有统计学意义(P < 0.01)。

临床应用与阅片要点

  • 在儿科胸部螺旋CT中,可参考40 mm准直以降低晶状体和甲状腺的成像体积外剂量。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Adaptive collimation reduces the dose deposited outside the imaged volume along the z-axis. An increase in the dose deposited outside the imaged volume (to the lens and thyroid) in the z-axis direction is a concern in paediatric computed tomography (CT).

OBJECTIVE: To compare the dose deposited outside the imaged volume (to the lens and thyroid) between 40-mm and 80-mm collimation during thoracic paediatric helical CT.

MATERIALS AND METHODS: We used anthropomorphic phantoms of newborns and 5-year-olds with 40-mm and 80-mm collimation during helical CT. We compared the measured dose deposited outside the imaged volume using optically stimulated luminescence dosimeters (OSLD) at the surfaces of the lens and thyroid and the image noise between the 40-mm and 80-mm collimations.

RESULTS: There were significant differences in the dose deposited outside the imaged volume (to the lens and thyroid) between the 40-mm and 80-mm collimations for both phantoms (P < 0.01).

CONCLUSION: Compared with that observed for 80-mm collimation in helical CT scans of the paediatric thorax, the dose deposited outside the imaged volume (to the lens and thyroid) was significantly lower in newborns and 5-year-olds with 40-mm collimation.

来源:Pediatr Radiol,查看原文。版权归原作者所有。

引用

  1. Masuda T, Kiguchi M, Fujioka C, et al. Impact of beam collimation of z-overscanning on dose to the lens and thyroid gland in paediatric thoracic computed tomography imaging. Pediatr Radiol. 2024. PMID: 38308740. doi:10.1007/s00247-024-05862-3

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。