光谱光子计数CT对人工心脏瓣膜的超高分辨与K边缘成像:体模研究
该体模研究将2枚金属瓣和3枚生物瓣置于含稀释碘对比剂的管中,在胸部体模内用光子计数探测器CT(PCD-CT)与双层双能CT(DEDL-CT)重复扫描,比较图像质量。客观分析显示PCD-CT下金属结构更薄、 Blooms伪影更少(金属伪影体素占比11% vs 40%);主观评分显示PCD-CT对多数结构显示更清晰,但瓣叶上约2 mm病变仍具挑战;钨K边缘成像可进一步减少伪影。
要点
- PCD-CT较DEDL-CT减少金属瓣结构Blooms伪影,金属伪影体素占比11% vs 40%。
- PCD-CT对瓣膜结构清晰度和锐利度评分更优(除机械瓣叶清晰度两者均4分)。
- 钨K边缘成像可显示含钨瓣叶并进一步减少Blooms和金属伪影。
核心结论
- PCD-CT较DEDL-CT客观和主观图像质量均改善,但瓣膜结构上毫米级病变仍是PCD-CT的挑战;钨K边缘成像可进一步减少伪影。
研究设计与样本
- 体模研究:2枚金属瓣和3枚生物瓣置于含稀释碘对比剂的管中,于胸部体模内以匹配参数在DEDL-CT和PCD-CT重复多角度扫描;4枚瓣膜结构上附着两个约2 mm厚病变(分别含肌肉和脂肪),加与不加延长环再次扫描;金属瓣另以钨K边缘成像参数扫描;3名专家读者对常规图像和钨K边缘图像行4分制主观评分。
关键影像表现/征象
- PCD-CT下瓣膜金属结构更薄且Blooms伪影更少。
- PCD-CT金属伪影体素占比11%(四分位距6%)vs DEDL-CT 40%(四分位距13%)。
- 部分结构仅在PCD-CT上可见或清晰可见。
- 仅PCD-CT可检测到两个病变之间的某些差异。
- 钨K边缘成像可显示含钨的瓣叶,图像Blooms和金属伪影较常规PCD-CT图像更少(P<0.01)。
诊断标准与关键数值
- 主观评分采用4分制,评估瓣膜各部分、病变、金属伪影和Blooms伪影的清晰度和锐利度。
- PCD-CT对全部结构清晰度和锐利度评分更优(均P<0.006),但机械瓣叶清晰度两者均为4分(四分位距3%)。
- PCD-CT对Blooms和条纹伪影评分均更优(P≤0.01)。
- PCD-CT对病变清晰度和锐利度均优于DEDL-CT(均P<0.02)。
- 加延长环后DEDL-CT清晰度和锐利度降低(P=0.04和P=0.02),PCD-CT仅锐利度降低(P=0.04)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: The contribution of cardiac computed tomography (CT) for the detection and characterization of prosthetic heart valve (PHV) complications is still limited due mainly to artifacts. Computed tomography systems equipped with photon-counting detectors (PCDs) have the potential to overcome these limitations. Therefore, the aim of the study was to compare image quality of PHV with PCD-CT and dual-energy dual-layer CT (DEDL-CT).
MATERIALS AND METHODS: Two metallic and 3 biological PHVs were placed in a tube containing diluted iodinated contrast inside a thoracic phantom and scanned repeatedly at different angles on a DEDL-CT and PCD-CT. Two small lesions (~2 mm thickness; containing muscle and fat, respectively) were attached to the structure of 4 valves, placed inside the thoracic phantom, with and without an extension ring, and scanned again. Acquisition parameters were matched for the 2 CT systems and used for all scans. Metallic valves were scanned again with parameters adapted for tungsten K-edge imaging. For all valves, different metallic parts were measured on conventional images to assess their thickness and blooming artifacts. In addition, 6 parallelepipeds per metallic valve were drawn, and all voxels with density <3 times the standard deviation of the contrast media were recorded as an estimate of streak artifacts. For subjective analysis, 3 expert readers assessed conventional images of the valves, with and without lesions, and tungsten K-edge images. Conspicuity and sharpness of the different parts of the valve, the lesions, metallic, and blooming artifacts were scored on a 4-point scale. Measurements and scores were compared with the paired t test or Wilcoxon test.
RESULTS: The objective analysis showed that, with PCD-CT, valvular metallic structures were thinner and presented less blooming artifacts. Metallic artifacts were also reduced with PCD-CT (11 [interquartile (IQ) = 6] vs 40 [IQ = 13] % of voxels). Subjective analysis allowed noticing that some structures were visible or clearly visible only with PCD-CT. In addition, PCD-CT yielded better scores for the conspicuity and for the sharpness of all structures (all P s < 0.006), except for the conspicuity of the leaflets of the mechanical valves, which were well visible with either technique (4 [IQ = 3] for both). Both blooming and streak artifacts were reduced with PCD-CT ( P ≤ 0.01). Overall, the use of PCD-CT resulted in better conspicuity and sharpness of the lesions compared with DEDL-CT (both P s < 0.02). In addition, only with PCD-CT some differences between the 2 lesions were detectable. Adding the extension ring resulted in reduced conspicuity and sharpness with DEDL-CT ( P = 0.04 and P = 0.02, respectively) and only in reduced sharpness with PCD-CT ( P = 0.04). Tungsten K-edge imaging allowed for the visualization of the only dense structure containing it, the leaflets, and it resulted in images judged having less blooming and metallic artifacts as compared with conventional PCD-CT images ( P < 0.01).
CONCLUSIONS: With PCD-CT, objective and subjective image quality of metallic and biological PHVs is improved compared with DEDL-CT. Notwithstanding the improvements in image quality, millimetric lesions attached to the structure of the valves remain a challenge for PCD-CT. Tungsten K-edge imaging allows for even further reduction of artifacts.