深度学习重建的运动不敏感成像技术提升颈椎MR图像质量的临床效果
研究纳入2020年12月至2021年3月35例颈椎MRI患者,比较快速自旋回波、原始PROPELLER及PROPELLER联合深度学习重建50%与75%四组图像。结果显示PROPELLER联合深度学习重建可提高脊髓和胸锁乳突肌信噪比及脊髓对比噪声比,并显著降低图像噪声。
要点
- PROPELLER联合深度学习重建50%与75%较快速自旋回波和原始PROPELLER提高脊髓及胸锁乳突肌信噪比。
- PROPELLER联合深度学习重建50%与75%较快速自旋回波在C3-4和C4-5层面提高脊髓对比噪声比。
- PROPELLER联合深度学习重建50%与75%较快速自旋回波和原始PROPELLER在脊髓、胸锁乳突肌和背部肌肉显示显著降噪效果。
核心结论
- PROPELLER与深度学习重建联合应用可提高图像质量,获得高信噪比和对比噪声比并降低噪声。
研究设计与样本
- 研究纳入2020年12月至2021年3月35例接受颈椎MRI检查的患者,定量和定性评估快速自旋回波、原始PROPELLER、PROPELLER深度学习重建50%和PROPELLER深度学习重建75%四组图像。
关键影像表现/征象
- 在所有椎间盘层面,PROPELLER深度学习重建50%与75%的脊髓和胸锁乳突肌平均信噪比显著高于快速自旋回波和原始PROPELLER图像(P < .0083)。
- PROPELLER深度学习重建50%与75%在C3-4和C4-5层面的脊髓平均对比噪声比显著高于快速自旋回波(P < .0083),PROPELLER深度学习重建75%在C6-7层面的脊髓平均对比噪声比显著高于快速自旋回波(P < .0083)。
- 定性分析显示,PROPELLER深度学习重建50%与75%图像在脊髓、胸锁乳突肌和背部肌肉的噪声方面较快速自旋回波和原始PROPELLER图像显示显著降噪效果。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To demonstrate that a T2 periodically rotated overlapping parallel lines with enhanced reconstruction (PROPELLER) technique using deep learning reconstruction (DLR) will provide better image quality and decrease image noise.
METHODS: From December 2020 to March 2021, 35 patients examined cervical spine MRI were included in this study. Four sets of images including fast spin echo (FSE), original PROPELLER, PROPELLER DLR50%, and DLR75% were quantitatively and qualitatively reviewed. We calculated the signal-to-noise ratio (SNR) of the spinal cord and sternocleidomastoid (SCM) muscle and the contrast-to-noise ratio (CNR) of the spinal cord by applying region-of-interest at the spinal cord, SCM muscle, and background air. We evaluated image noise with regard to the spinal cord, SCM, and back muscles at each level from C2-3 to C6-7 in the 4 sets.
RESULTS: At all disc levels, the mean SNR values for the spinal cord and SCM muscles were significantly higher in PROPELLER DLR50% and DLR75% compared to FSE and original PROPELLER images (P < .0083). The mean CNR values of the spinal cord were significantly higher in PROPELLER DLR50% and DLR75% compared to FSE at the C3-4 and 4-5 levels and PROPELLER DLR75% compared to FSE at the C6-7 level (P < .0083). Qualitative analysis of image noise on the spinal cord, SCM, and back muscles showed that PROPELLER DLR50% and PROPELLER DLR75% images showed a significant denoising effect compared to the FSE and original PROPELLER images.
CONCLUSION: The combination of PROPELLER and DLR improved image quality with a high SNR and CNR and reduced noise.
ADVANCES IN KNOWLEDGE: Motion-insensitive imaging technique (PROPELLER) increased the image quality compared to conventional FSE images. PROPELLER technique with a DLR reduced image noise and improved image quality.
引用
- Song YS, Lee IS, Hwang M, et al. Clinical efficacy of motion-insensitive imaging technique with deep learning reconstruction to improve image quality in cervical spine MR imaging. Br J Radiol. 2024. PMID: 38366622. doi:10.1093/bjr/tqae026
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。