成人患者中植入MR兼容、MR不兼容及其他有源植入式医疗器械的MRI检查
本文综述成人有源植入式医疗器械(AIMD)患者接受MRI检查的安全性管理,涉及MRI安全物理学概念、MRI安全人员职责、厂商手册信息获取、协议修改与风险获益权衡,并概述推荐工作流程及导线断裂、多器械植入等进阶议题。
要点
- AIMD包括心脏植入式电子器械、脊髓/脑深部/骨生长刺激器、神经刺激器及药物输注泵等。
- 需掌握峰值空间梯度磁场、比吸收率、射频脉冲有效磁场分量均方根值及梯度切换率等安全参数。
- 应建立算法化MRI安全流程,含安全信息审查、患者筛查、优化成像及检查前后监测。
核心结论
- 放射科医生应能修改扫描协议以满足安全要求,考虑MRI替代方案,并权衡MRI潜在获益与风险。
- 实施算法化MRI安全流程很重要,包括审查MRI安全信息、患者筛查、优化成像及检查前中后监测患者。
研究设计与样本
- 本文为综述,由作者复习MRI安全相关物理学概念、国际磁共振医学学会(ISMRM)设定的MRI安全人员职责及厂商手册信息,并概述推荐工作流程。
诊断标准与关键数值
- MRI安全相关参数包括峰值空间梯度磁场、比吸收率、发射射频脉冲有效磁场分量的均方根值及梯度切换率。
- 维持安全限值所需的参数及获取最新信息的途径源自厂商手册。
临床应用与阅片要点
- 按照国际磁共振医学学会(ISMRM)的规定明确MRI安全人员的职责。
- 从厂商手册获取相关最新信息,并据此修改协议以满足安全要求。
- 评估MRI可能的替代方案,并权衡MRI潜在获益与潜在风险。
- 对导线断裂或废弃导线及多枚植入器械患者等进阶议题予以处理。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Active implanted medical devices (AIMDs) enable therapy and patient monitoring by way of electrical activity and typically have a battery and electrical leads. The most common types of AIMDs include cardiac implantable electronic devices (CIEDs), spinal cord stimulators, deep brain stimulators, bone growth or fusion stimulators, other neurostimulators, and drug infusion pumps. As more patients with AIMDs undergo MRI, it is important to consider the safety of patients who have these implanted devices during MRI. The authors review the physics concepts related to MRI safety, such as peak spatial gradient magnetic field, specific absorption rate, root mean square value of the effective magnetic component of the transmitted RF pulse, and gradient slew rate, as well as the parameters necessary to remain within safety limits. The roles of MRI safety personnel, as set forth by the International Society of Magnetic Resonance in Medicine, are emphasized. In addition, the relevant information provided in vendor manuals is reviewed, with a focus on how to obtain relevant up-to-date information. The radiologist should be able to modify protocols to meet safety requirements, address possible alternatives to MRI, and weigh the potential benefits of MRI against the potential risks. A few more advanced topics, such as fractured or abandoned device leads and patients with multiple implanted medical devices, also are addressed. Recommended workflows for MRI in patients with implanted medical devices are outlined. It is important to implement an algorithmic MRI safety process, including a review of the MRI safety information; patient screening; optimal imaging; and monitoring patients before, during, and after the examination. ©RSNA, 2024 Test Your Knowledge questions for this article are available in the supplemental material. See the invited commentary by Shetty et al in this issue.
引用
- Harwood M, Fahrenholtz SJ, Wellnitz CV, et al. MRI in Adult Patients with Active and Inactive Implanted MR-conditional, MR-nonconditional, and Other Devices. Radiographics. 2024. PMID: 38421911. doi:10.1148/rg.230102
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。