光子计数探测器CT:虚拟单能成像联合锡预滤过减少踝关节术后金属伪影的临床效用
该回顾性研究纳入48例踝关节内固定患者,比较光子计数探测器CT有锡预滤过(n=30)与无锡预滤过(n=18)采集下虚拟单能图像(60–190 keV)对金属伪影的减少效果。锡预滤过采集的伪影水平更低,120 keV虚拟单能图像较常规多能图像显著减少金属伪影,并提高骨愈合可见度和图像可判读性。作者认为锡预滤过光子计数探测器CT联合虚拟单能重建可能是有益补充。
要点
- 锡预滤过采集较无锡预滤过采集在多能图像和虚拟单能图像上均显示更低伪影水平(P≤0.043)。
- 锡预滤过虚拟单能图像≥120 keV较常规多能图像显著减少高密度和低密度金属伪影。
- 锡预滤过120 keV虚拟单能图像在骨愈合可见度、图像可判读性上优于常规多能图像,伪影范围评分更低。
核心结论
- 锡预滤过虚拟单能图像在120 keV时较常规多能图像显著减少金属伪影,并改善骨愈合可见度和图像可判读性;作者认为锡预滤过光子计数探测器CT联合虚拟单能重建可能是有益补充。
研究设计与样本
- 回顾性研究,纳入2023年3月至10月在光子计数探测器CT上行踝关节内固定扫描的患者48例(女21例,男27例,平均年龄55.1±19.4岁),分别有锡预滤过采集30例和无锡预滤过采集18例;在骨算法下以10 keV递增重建60–190 keV虚拟单能图像(分别为VMI Sn和VMI Std),比较噪声测量及3名阅读者对5个重建级别的骨愈合可见度、可判读性和伪影范围评分。
关键影像表现/征象
- 锡预滤过采集(n=30)较无锡预滤过采集(n=18)在多能图像和虚拟单能图像上伪影水平更低(P≤0.043)。
- VMI Sn在≥120 keV时较常规多能图像显著减少金属伪影:高密度伪影40.2 HU(IQR 39.8)vs 14.0 HU(IQR 11.1),P≤0.01;低密度伪影91.2 HU(IQR 82.4)vs 29.7 HU(IQR 39.6),P≤0.001。
- VMI Std在≥100 keV时较常规多能图像显著减少金属伪影:高密度伪影57.7 HU(IQR 33.4)vs 19.4 HU(IQR 27.6),P≤0.001;低密度伪影106.9 HU(IQR 76.1)vs 57.4 HU(IQR 55.7),P≤0.021。
诊断标准与关键数值
- VMI Sn在120 keV时骨愈合可见度评分高于常规多能图像(P≤0.001)。
- VMI Sn在120 keV时图像可判读性评分优于常规多能图像(P=0.023),伪影范围评分低于常规多能图像(P≤0.001)。
临床应用与阅片要点
- 作者认为锡预滤过光子计数探测器CT联合虚拟单能重建可能是踝关节金属植入物术后CT成像的有益补充。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: The aim of this study was to compare the effectiveness and clinical utility of virtual monoenergetic image (VMI) reconstructions in computed tomography (CT) scans with and without tin prefiltration on a photon-counting detector (PCD) CT system to reduce metal implant artifacts in the postoperative ankle.
MATERIALS AND METHODS: This retrospective study included patients with internal fixation of the ankle scanned with and without tin prefiltration (Sn) on a PCD CT scanner between March and October 2023. Virtual monoenergetic images between 60 and 190 keV were reconstructed with a 10-keV increment in a bone kernel for both acquisitions (VMI Sn and VMI Std , respectively). Noise measurements assessed artifact reduction in the most prominent near-metal image distortions and were compared between acquisitions modes as well as between polychromatic images and VMIs. Three readers assessed the visibility of osseous healing along with interpretability and artifact extent for 5 reconstruction levels.
RESULTS: A total of 48 patients (21 females, 27 males; mean age, 55.1 ± 19.4 years) were included in this study. Tin-prefiltered acquisitions (n = 30) had a lower artifact level for polychromatic images and VMIs compared with non-tin-prefiltered acquisitions (n = 18; P ≤ 0.043). A significant reduction of metal artifacts was observed for VMI Sn ≥120 keV compared with polychromatic images (hyperdense artifacts: 40.2 HU [interquartile range (IQR) 39.8] vs 14.0 HU [IQR 11.1]; P ≤ 0.01 and hypodense artifacts: 91.2 HU [IQR 82.4] vs 29.7 HU [IQR 39.6]; P ≤ 0.001). For VMI Std , this applied to reconstructions ≥100 keV (hyperdense artifacts: 57.7 HU [IQR 33.4] vs 19.4 HU [IQR 27.6]; P ≤ 0.001 and hypodense artifacts: 106.9 HU [IQR 76.1] vs 57.4 HU [IQR 55.7]; P ≤ 0.021). For visibility of osseous healing, VMI Sn at 120 keV yielded higher ratings compared with polychromatic images ( P ≤ 0.001), whereas image interpretability was rated better ( P = 0.023), and artifact extent was rated lower ( P ≤ 0.001) compared with polychromatic images.
CONCLUSIONS: Tin-prefiltered VMI at 120 keV showed a significant reduction in metal artifacts compared with polychromatic images, whereas visibility of osseous healing and image interpretability was improved. Therefore, tin-prefiltration PCD CT with VMI reconstructions may be a helpful complement to postsurgical CT imaging of the ankle in patients with metal implants.
引用
- Marth AA, Goller SS, Kajdi GW, et al. Photon-Counting Detector CT: Clinical Utility of Virtual Monoenergetic Imaging Combined With Tin Prefiltration to Reduce Metal Artifacts in the Postoperative Ankle. Invest Radiol. 2024. PMID: 38214560. doi:10.1097/RLI.0000000000001058
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。