金属脊柱植入物患者0.55 T与1.5 T MRI金属伪影严重程度及近金属解剖评估能力的比较——扫描仪比较研究
本研究纳入50例后路融合术后患者,比较0.55 T与1.5 T MRI的金属伪影及近金属解剖评估能力。0.55 T信号/对比与分辨率评分较低,但磁敏感伪影更轻,对椎管和神经孔的评估能力与1.5 T相当;在多节段植入物患者中,0.55 T对椎管和神经孔的评估优于1.5 T。
要点
- 0.55 T磁敏感伪影评分1.8±0.5,显著低于1.5 T的3.0±0.6(p<.001)。
- 椎管评估0.55 T 4.4±0.5 vs 1.5 T 4.2±0.9(p=.69),神经孔3.8±0.5 vs 3.8±0.9(p=.19),两组相当。
- 多节段植入物患者中,0.55 T对椎管和神经孔的评估优于1.5 T(4.4±0.5 vs 3.6±1.1,p<.001)。
核心结论
- 脊柱植入物金属伪影在0.55 T MRI上明显较轻;在多节段后路融合患者中,这转化为对近金属解剖更优的评估能力,而1.5 T MRI在此达到诊断极限。
研究设计与样本
- 扫描仪比较研究,纳入50例连续患者(平均年龄69±12岁),均为腰椎或胸腰椎后路融合术后并接受0.55 T与1.5 T MRI检查,采用临床常规金属伪影减少方案,由两名经过专科培训的骨肌放射科医生以5点Likert量表评分。
关键影像表现/征象
- 0.55 T信号/对比评分为4.0±0.3,低于1.5 T的4.4±0.6(p<.001)。
- 0.55 T分辨率评分为3.8±0.5,低于1.5 T的4.2±0.7(p<.001)。
- 0.55 T磁敏感伪影评分为1.8±0.5,轻于1.5 T的3.0±0.6(p<.001)。
- 按植入物长度亚组分析,1.5 T近金属解剖可见度随植入节段增加而下降,0.55 T则保持不变。
诊断标准与关键数值
- 椎管评估能力评分:0.55 T为4.4±0.5 vs 1.5 T为4.2±0.9(p=.69),两组相当。
- 神经孔评估能力评分:0.55 T为3.8±0.5 vs 1.5 T为3.8±0.9(p=.19),两组相当。
- 多节段植入物患者椎管和神经孔评估:0.55 T为4.4±0.5 vs 1.5 T为3.6±1.1(p<.001),0.55 T更优。
- 两名阅片者间一致性极好,范围0.84-0.94。
临床应用与阅片要点
- 后路融合术后患者行MRI时,0.55 T可减轻金属伪影,对椎管和神经孔的评估能力与1.5 T相当。
- 多节段植入物患者中,0.55 T对近金属解剖的评估优于1.5 T,而1.5 T在此达到诊断极限。
- 0.55 T信号/对比与分辨率评分低于1.5 T,阅片时需结合伪影减轻与图像质量下降权衡。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: To compare image quality and metal artifact severity at 0.55 T and 1.5 T MRI in patients with spinal implants following posterior fusion surgery.
MATERIALS AND METHODS: 50 consecutive patients (mean age: 69 ± 12 years) who underwent 0.55 T and 1.5 T MRI following posterior fusion surgery of the lumbar or thoracolumbar spine were included. Examinations used metal artifact reduction protocols from clinical routine. Images were rated by two fellowship-trained musculoskeletal radiologists for image quality, ability to assess the spinal canal and the neural foramina, and artifact severity on 5-point Likert scales. Additionally, differences in artifact severity and visibility of near-metal anatomy among implant sizes (1-level vs. 2-level vs. >2-levels) were evaluated.
RESULTS: Signal/contrast (mean: 4.0 ± 0.3 [0.55 T] vs. 4.4 ± 0.6 [1.5 T]; p < .001) and resolution (3.8 ± 0.5 vs. 4.2 ± 0.7; p < .001) were rated lower at 0.55 T. The ability to assess the spinal canal (4.4 ± 0.5 vs. 4.2 ± 0.9; p = .69) and the neural foramina (3.8 ± 0.5 vs. 3.8 ± 0.9; p = .19) were however rated equally good with excellent interrater agreement (range: 0.84-0.94). Susceptibility artifacts were rated milder at 0.55 T (1.8 ± 0.5 vs. 3.0 ± 0.6; p < .001). For implant size-based subgroups, the visibility of near-metal anatomy decreased with implant length at 1.5 T, but remained unchanged at 0.55 T. In consequence, the spinal canal and neural foramina could be better assessed at 0.55 T in patients with multi-level implants (4.4 ± 0.5 vs. 3.6 ± 1.1; p < .001).
CONCLUSION: Metal artifacts of spinal implants are substantially less pronounced at 0.55 T MRI. When examining patients with multi-level posterior fusion, this translates into a superior ability to assess near-metal anatomy, where 1.5 T MRI reaches diagnostic limitations.
引用
- Seifert AC, Breit HC, Schlicht F, et al. Comparing Metal Artifact Severity and Ability to Assess Near-Metal Anatomy Between 0.55 T and 1.5 T MRI in Patients with Metallic Spinal Implants-A Scanner Comparison Study. Acad Radiol. 2024. PMID: 38242732. doi:10.1016/j.acra.2023.12.048