全肩关节置换术后灾难性失效导致严重金属沉着症一例报告
本病例报告描述一例解剖型全肩关节置换术后因聚乙烯组件失效导致严重金属沉着症的患者,使用光子计数探测器CT联合金属伪影减少后处理算法明确诊断,该先进影像方法有效识别了植入物失效来源及严重金属沉着症的表现,包括骨溶解和假性肿瘤形成。
要点
- 肩关节非负重关节,金属沉着症发生率极低,但聚乙烯组件磨损、骨折或分离可致异常金属对金属接触。
- 光子计数探测器CT联合金属伪影减少算法可有效识别植入物失效来源及金属沉着症的表现。
- 金属沉着症可表现为骨溶解和假性肿瘤形成,先进影像有助于指导手术规划。
核心结论
- 光子计数探测器CT联合金属伪影减少后处理算法能准确描述金属沉着症的严重程度和范围,有助于指导手术规划以减少相关严重并发症。
研究设计与样本
- 病例报告,一例解剖型全肩关节置换术后患者,使用光子计数探测器CT并辅以金属伪影减少后处理算法进行诊断。
关键影像表现/征象
- 影像学表现包括骨溶解和假性肿瘤形成。
- 先进影像方法有效识别了植入物失效的来源。
临床应用与阅片要点
- 对于全肩关节置换术后疑似金属沉着症的患者,使用光子计数探测器CT联合金属伪影减少算法可帮助识别植入物失效来源及金属沉着症的范围。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Metallosis is an unusual but consequential complication arising from orthopedic hardware implantation, characterized by the deposition of metallic particles in the periprosthetic soft tissues. The incidence of metallosis associated with shoulder arthroplasties is exceptionally rare since the shoulder is not a weight-bearing joint, making it less susceptible to mechanical wear and, consequently, to conditions like particle disease and metallosis. Nevertheless, anomalous metal-on-metal interactions can develop in total shoulder arthroplasties if the polyethylene component fails due to wear, fracture, or dissociation. If left unaddressed, metallosis can incite an adverse immune-mediated local tissue response, culminating in joint destruction and adjacent soft tissues and muscle necrosis. In this case report, the diagnosis of metallosis was made in a patient with an anatomic total shoulder arthroplasty using a state-of-the-art photon counting detector CT supplemented by post-processing metal artifact reduction algorithms. This advanced imaging approach was effective in discerning the source of implant failure and in identifying manifestations of severe metallosis including osteolysis and pseudotumor formation. Advanced imaging methods can accurately characterize the severity and extent of metallosis, thereby helping guide surgical planning to mitigate serious complications associated with this condition.
引用
- Ogunfuwa FO, Needell S, Simovitch RW, et al. Severe metallosis following catastrophic failure of total shoulder arthroplasty - a case report. Skeletal Radiol. 2024. PMID: 38236295. doi:10.1007/s00256-024-04575-w
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。