多阅片者多病例受试者工作特征研究评估定量透射超声在乳腺病灶检出与召回方面相对于数字乳腺断层合成的非劣效性
本研究为多阅片者多病例观察者性能研究,24名放射科医生解读177例(66例含癌、异型性或实性肿块,111例正常或非实性良性异常)的定量透射超声与数字乳腺断层合成图像。定量透射超声的AUC为0.746±0.028,非劣效于数字乳腺断层合成的0.700±0.028(界值-0.05,P<.05);敏感度70.6±7.2% vs 85.2±6.4%,特异度60.1±12.3% vs 37.2±11.0%,差异均有统计学意义。
要点
- 定量透射超声AUC 0.746±0.028非劣效于数字乳腺断层合成0.700±0.028(非劣效界值-0.05,P<.05)。
- 定量透射超声敏感度70.6±7.2%低于数字乳腺断层合成85.2±6.4%,差异有统计学意义。
- 定量透射超声特异度60.1±12.3%高于数字乳腺断层合成37.2±11.0%,差异有统计学意义。
核心结论
- 定量透射超声可能是乳腺X线摄影的潜在替代方案,用于因年龄太小而不适合接受乳腺X线摄影的女性进行乳腺癌筛查。
研究设计与样本
- 多阅片者多病例观察者性能研究;前瞻性病例收集研究在四个临床中心进行;24名符合《乳腺X线摄影质量标准法案》资格的放射科医生解读177例(66例含癌、异型性或实性肿块,111例正常或非实性良性异常);先解读定量透射超声,2周后解读数字乳腺断层合成合成2D视图。
关键影像表现/征象
- 定量透射超声成像无乳腺X线摄影中所见的组织重叠。
- 在21例囊肿病例中,阅片者平均对1.1±1.4例在定量透射超声上建议召回而不在数字乳腺断层合成上建议召回。
- 在21例囊肿病例中,阅片者平均对10.6±2.2例在数字乳腺断层合成上建议召回而不在定量透射超声上建议召回。
诊断标准与关键数值
- 定量透射超声AUC为0.746±0.028,数字乳腺断层合成AUC为0.700±0.028,AUC差异为0.046±0.028(95%CI:[-0.008, 0.101]),非劣效界值为-0.05(P<.05)。
- 定量透射超声敏感度为70.6±7.2%,数字乳腺断层合成敏感度为85.2±6.4%,差异有统计学意义。
- 定量透射超声特异度为60.1±12.3%,数字乳腺断层合成特异度为37.2±11.0%,差异有统计学意义。
临床应用与阅片要点
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: Quantitative transmission (QT) imaging is an emerging volumetric ultrasound modality for women too young for mammography. QT images tissue without overlap seen in mammography, thereby can potentially improve breast mass detection and characterization and noncancer recall. We compared radiologists' interpretation of QT vs digital breast tomosynthesis (DBT) with a multireader multicase observer performance study.
MATERIALS AND METHODS: Study subjects received screening DBT and QT scans in HIPAA-compliant, institutional review board-approved prospective case-collection studies at four clinical sites. Twenty-four Mammography Quality Standards Act-qualified radiologists interpreted 177 cases (66 with cancer, atypia, or solid mass and 111 normal or with nonsolid benign abnormality), first QT, then 2 weeks later DBT synthesized 2D-views. Readers reported up to three findings per case and for each finding a recall or no recall decision and confidence of that decision. The study hypothesis was area under receiver operating characteristic curve (AUC) of QT was noninferior to DBT. Sensitivity and specificity were also compared.
RESULTS: AUC of QT (0.746 ± 0.028, mean ± SD) was noninferior to DBT (0.700 ± 0.028) for AUC difference margin of -0.05 (P < .05). AUC difference was 0.046 ± 0.028 (95% CI: [-0.008, 0.101]). Sensitivity was 70.6 ± 7.2% for QT and 85.2 ± 6.4% for DBT, specificity was 60.1 ± 12.3% vs 37.2 ± 11.0%, and both differences were statistically significant. Of a total of 21 cases of cysts, readers recommended recall, on average, in 1.1 ± 1.4 cases with QT, but not with DBT, and 10.6 ± 2.2 cases with DBT, but not with QT.
CONCLUSION: QT can be a potential alternative to mammography for breast cancer screening of women too young to undergo mammography.
引用
- Jiang Y, Iuanow E, Malik B, et al. A Multireader Multicase (MRMC) Receiver Operating Characteristic (ROC) Study Evaluating Noninferiority of Quantitative Transmission (QT) Ultrasound to Digital Breast Tomosynthesis (DBT) on Detection and Recall of Breast Lesions. Acad Radiol. 2024. PMID: 38290888. doi:10.1016/j.acra.2023.12.038