2019-2023年英国乳腺筛查项目中的辐射剂量
该研究通过在线提交系统收集英国405台X线设备的111152例筛查和5113例评估曝光参数,结合输出与束质测量估算平均腺体剂量(MGD)。结果显示FFDM剂量较2016-2019年全国调查增加约10%而压迫乳腺厚度相似,DBT剂量总体较FFDM高34%-40%,50-60 mm厚乳腺高34%。
要点
- 英国乳腺筛查中FFDM的MGD较2016-2019年调查增加约10%,压迫乳腺厚度相似。
- DBT的MGD总体较FFDM高34%-40%,50-60 mm厚乳腺高34%。
- 低压迫乳腺厚度下PMMA等效乳腺厚度可能被高估,但证据不确凿。
核心结论
- 英国乳腺筛查中X线机型构成的变化导致FFDM乳腺剂量升高;NHSBSP中DBT剂量平均较FFDM高约34%-40%。
研究设计与样本
- 全国性剂量调查,收集英国405台X线设备的111152例筛查和5113例评估的匿名曝光参数,通过在线提交系统关联国家乳腺摄影质量控制数据库。
诊断标准与关键数值
- FFDM的MGD较2016-2019年全国调查增加约10%,压迫乳腺厚度保持相似。
- DBT的MGD总体较FFDM高34%-40%。
- 50-60 mm厚乳腺中DBT的MGD较FFDM高34%。
- 低压迫乳腺厚度下PMMA等效乳腺厚度可能被高估,但证据不确凿。
局限性
- 低压迫乳腺厚度下PMMA等效乳腺厚度可能被高估,但证据不确凿。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To report the latest UK mammography dose survey results and to compare radiation doses from digital breast tomosynthesis (DBT) and full-field digital mammography (FFDM) in UK breast screening.
METHODS: Anonymized exposure factors were collected for 111 152 screening cases and 5113 assessment cases from 405 x-ray sets across the United Kingdom using an online submission system linked to a national database of mammography quality control data. Output and beam quality measurements from each set were combined with exposure data to estimate mean glandular doses (MGD).
RESULTS: FFDM doses increased by ∼10% compared to the 2016-2019 national survey but compressed breast thicknesses (CBT) remained similar. DBT doses were 34%-40% higher than FFDM overall and 34% higher than FFDM for breasts 50-60 mm thick. We found a possible overestimation of PMMA breast equivalent thicknesses at low CBTs, but the evidence was not conclusive.
CONCLUSION: Recent changes to the mix of x-ray models in use in UK breast screening have resulted in higher FFDM breast doses. DBT doses in the NHSBSP are on average higher than FFDM by ∼34%-40%.
ADVANCES IN KNOWLEDGE: This is the first national study to report DBT and FFDM MGDs in UK breast screening.
引用
- Loveland J, Mackenzie A. Radiation doses received in the UK breast screening programmes 2019-2023. Br J Radiol. 2024. PMID: 38291906. doi:10.1093/bjr/tqad039