经直肠高频超声系统适配前列腺光声成像:实现与临床前验证
研究将临床经直肠高频微超声系统(≥15 MHz)与经尿道侧射出光光纤结合,构建前列腺光声成像平台,并在体模和无壁通道中评估分辨率与造影剂显像,同时在大鼠肿瘤模型中经静脉注射卟啉体进行在体成像,显示注射后3小时光声信号较血液信号升高。作者认为高分辨率光声成像可能为前列腺诊断和治疗应用提供有价值补充信息。
要点
- 经直肠微超声(≥15 MHz)联合经尿道侧射出光可实现前列腺光声成像,14 mm深度轴向分辨率74 μm、横向261 μm。
- 聚氯乙烯-塑溶胶体模中声速1383 m/s、20 MHz衰减4 dB/mm,可用于无壁通道内光声造影剂成像。
- 肿瘤模型静脉注射卟啉体后3小时可见光声信号较血液信号升高,并可通过光谱解混分离。
核心结论
- 成功构建并表征了经直肠微超声与光声成像联合平台,并在肿瘤模型中完成在体成像;高分辨率光声成像可能为前列腺诊断和治疗应用提供有价值补充信息。
研究设计与样本
- 临床经直肠微超声系统改装为与可调谐激光脉冲同步采集光声图像,配经尿道侧射出光光纤;采用聚氯乙烯-塑溶胶体模在无壁通道中成像染料和可生物降解纳米颗粒造影剂卟啉体;水中测分辨率后在肿瘤模型经静脉注射卟啉体行在体成像。
关键影像表现/征象
- 光声图像以5 Hz采集,图像重建离线完成。
- 14 mm深度处光声图像分辨率轴向74 μm、横向261 μm。
- 聚氯乙烯-塑溶胶中声速1383 m/s,20 MHz衰减4 dB/mm。
- 肿瘤内卟啉体光声信号经光谱解混与血液信号分离,注射后3小时观察到信号升高。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: High-frequency, high-resolution transrectal micro-ultrasound (micro-US: ≥15 MHz) imaging of the prostate is emerging as a beneficial tool for scoring disease risk and accurately targeting biopsies. Adding photoacoustic (PA) imaging to visualize abnormal vascularization and accumulation of contrast agents in tumors has potential for guiding focal therapies. In this work, we describe a new imaging platform that combines a transrectal micro-US system with transurethral light delivery for PA imaging.
METHODS: A clinical transrectal micro-US system was adapted to acquire PA images synchronous to a tunable laser pulse. A transurethral side-firing optical fiber was developed for light delivery. A polyvinyl chloride (PVC)-plastisol phantom was developed and characterized to image PA contrast agents in wall-less channels. After resolution measurement in water, PA imaging was demonstrated in phantom channels with dyes and biodegradable nanoparticle contrast agents called porphysomes. In vivo imaging of a tumor model was performed, with porphysomes administered intravenously.
RESULTS: Photoacoustic imaging data were acquired at 5 Hz, and image reconstruction was performed offline. PA image resolution at a 14-mm depth was 74 and 261 μm in the axial and lateral directions, respectively. The speed of sound in PVC-plastisol was 1383 m/s, and the attenuation was 4 dB/mm at 20 MHz. PA signal from porphysomes was spectrally unmixed from blood signals in the tumor, and a signal increase was observed 3 h after porphysome injection.
CONCLUSION: A combined transrectal micro-US and PA imaging system was developed and characterized, and in vivo imaging demonstrated. High-resolution PA imaging may provide valuable additional information for diagnostic and therapeutic applications in the prostate.
引用
- Singh N, Chérin E, Roa CF, et al. Adaptation of a Clinical High-Frequency Transrectal Ultrasound System for Prostate Photoacoustic Imaging: Implementation and Pre-clinical Demonstration. Ultrasound Med Biol. 2024. PMID: 38238200. doi:10.1016/j.ultrasmedbio.2023.11.010