前列腺动脉栓塞术前规划:CT与CBCT的辐射暴露和图像质量比较

PAE planning: Radiation exposure and image quality of CT and CBCT.
PubMed 文献 · Eur J Radiol · 2024-01-23

研究比较CT与锥形束CT(CBCT)在前列腺动脉栓塞术前规划中的器官剂量、有效剂量和图像质量。结果显示CT有效剂量2.7 mSv低于CBCT的21.8 mSv,除睾丸外各器官剂量CT均低于CBCT,CT对前列腺动脉的对比噪声比和检出能力更优。

要点

  • CT有效剂量2.7 mSv,CBCT为21.8 mSv,CT明显更低。
  • 除睾丸外,CT各盆腔器官剂量均低于CBCT。
  • CT对前列腺动脉的CNR和检出能力优于CBCT。

核心结论

  • 术前CT较CBCT器官剂量更低、图像质量更好,应考虑用于前列腺动脉的正确识别。

研究设计与样本

  • 使用双能CT扫描仪和平板血管造影系统,在含65个热释光剂量计的人体模中测量器官剂量,并选取体型与体模接近的患者计算盆腔动脉CNR。

关键影像表现/征象

  • CT对盆腔动脉的CNR优于CBCT,但一支前列腺动脉在CT上显示狭窄时例外。
  • 有经验的放射科医生评估也证实CT对前列腺动脉的可检出性更好。

诊断标准与关键数值

  • CT有效剂量2.7 mSv,CBCT有效剂量21.8 mSv。
  • 体模器官剂量CT vs CBCT:胆囊1.2 mGy vs 3.3 mGy,肠道5.8 mGy vs 23.9 mGy,膀胱6.9 mGy vs 19.4 mGy,前列腺6.4 mGy vs 13.2 mGy,睾丸4.7 mGy vs 2.4 mGy,骨盆活性骨髓5.1 mGy vs 18.2 mGy,股骨活性骨髓3.3 mGy vs 6.6 mGy。

临床应用与阅片要点

  • 作者结论认为术前CT器官剂量更低、图像质量更好,应考虑用于前列腺动脉的正确识别。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To determine accurate organ doses, effective doses, and image quality of computed tomography (CT) compared with cone beam CT (CBCT) for correct identification of prostatic arteries.

METHOD: A dual-energy CT scanner and a flat-panel angiography system were used. Dose measurements (gallbladder (g), intestine (i), bladder (b), prostate (p), testes (t), active bone marrow of pelvis (bmp) and femura (bmf)) were performed using an anthropomorphic phantom with 65 thermoluminescent dosimeters in the pelvis and abdomen region. For the calculation of the contrast-to-noise ratio (CNR) of the pelvic arteries, a patient whose weight and height were almost identical to those of the phantom was selected for each examination type.

RESULTS: The effective dose of CT was 2.7 mSv and that of CBCT was 21.8 mSv. Phantom organ doses were lower for CT than for CBCT in all organs except the testes (g: 1.2 mGy vs. 3.3 mGy, i: 5.8 mGy vs. 23.9 mGy, b: 6.9 mGy vs. 19.4 mGy, p: 6.4 mGy vs. 13.2 mGy, t: 4.7 mGy vs. 2.4 mGy, bmp: 5.1 mGy vs. 18.2 mGy, bmf: 3.3 mGy vs. 6.6 mGy). For human pelvic arteries, the CNR of CT was better than that of CBCT, with the exception of one prostate artery that showed stenosis on CT. Evaluation by experienced radiologists also confirmed the better detectability of prostate arteries on CT examination.

CONCLUSIONS: In our study preprocedural CT had lower organ doses and better image quality comparedd with CBCT and should be considered for the correct identification of prostatic arteries.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Steiniger B, Fiebich M, Grimm MO, et al. PAE planning: Radiation exposure and image quality of CT and CBCT. Eur J Radiol. 2024. PMID: 38278010. doi:10.1016/j.ejrad.2024.111329

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。