运动补偿重建对双层探测器CT系统冠状动脉分析的影响:一项动态体模研究

The influence of motion-compensated reconstruction on coronary artery analysis for a dual-layer detector CT system: a dynamic phantom study.
PubMed 文献 · Eur Radiol · 2024-01-04

该动态体模研究用双层光谱探测器CT扫描置于仿真胸模中的5 mm人工中空冠状动脉,以0–80 mm/s水平速度模拟心脏运动,比较运动补偿重建(MCR)与标准重建的运动伪影面积。MCR使运动伪影面积随速度增加的斜率由0.44降至0.04,运动伪影减少11倍,作者认为可能改善心率偏高患者的CCTA评估。

要点

  • MCR在各运动速度下将运动伪影面积随时间变化的斜率由标准重建的0.44降至0.04,伪影减少11倍。
  • 标准重建在速度>20 mm/s时伪影面积显著大于参考值(最高57.2 mm²),MCR仅70 mm/s时仍显著(29.0 mm²)。
  • 作者认为MCR可能减少非诊断性扫描,尤其适用于心率偏高、无法用β受体阻滞剂或未达目标心率者。

核心结论

  • MCR可能通过减少冠状动脉运动伪影改善CCTA评估,尤其适用于心率偏高、不能接受β受体阻滞剂或未达到目标心率的患者。

研究设计与样本

  • 动态体模研究:双层光谱探测器CT扫描置于仿真胸模中央、内充碘对比剂(120 kVp下400 HU)的5 mm直径人工中空冠状动脉,以0–80 mm/s、每10 mm/s递增的恒定水平速度平移,每个速度重复5次临床协议CCTA扫描。

关键影像表现/征象

  • 参考平均(95%CI)运动伪影面积为24.9 mm²(23.8,26.0)。
  • 无MCR时,速度>20 mm/s的运动伪影面积显著大于参考值,最高达57.2 mm²(40.1,74.2)。
  • 有MCR时,除70 mm/s为29.0 mm²(27.0,31.0)外,其余速度与参考值无显著差异。

诊断标准与关键数值

  • 运动伪影以平面内运动面积(mm²)定量,参考均值为24.9 mm²(95%CI 23.8,26.0)。
  • 非MCR与MCR拟合直线斜率分别为0.44和0.04,据此计算运动伪影减少11倍。

临床应用与阅片要点

  • 作者认为MCR可能改善CCTA评估,尤其对心率偏高、不能接受β受体阻滞剂或未达目标心率者。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Cardiac motion artifacts hinder the assessment of coronary arteries in coronary computed tomography angiography (CCTA). We investigated the impact of motion compensation reconstruction (MCR) on motion artifacts in CCTA at various heart rates (HR) using a dynamic phantom.

MATERIALS AND METHODS: An artificial hollow coronary artery (5-mm diameter lumen) filled with iodinated contrast agent (400 HU at 120 kVp), positioned centrally in an anthropomorphic chest phantom, was scanned using a dual-layer spectral detector CT. The artery was translated at constant horizontal velocities (0-80 mm/s, increment of 10 mm/s). For each velocity, five CCTA scans were repeated using a clinical protocol. Motion artifacts were quantified using the in-plane motion area. Regression analysis was performed to calculate the reduction in motion artifacts provided by MCR, by division of the slopes of non-MCR and MCR fitted lines.

RESULTS: Reference mean (95% confidence interval) motion artifact area was 24.9 mm2 (23.8, 26.0). Without MCR, motion artifact areas for velocities exceeding 20 mm/s were significantly larger (up to 57.2 mm2 (40.1, 74.2)) than the reference. With MCR, no significant differences compared to the reference were shown for all velocities, except for 70 mm/s (29.0 mm2 (27.0, 31.0)). The slopes of the fitted data were 0.44 and 0.04 for standard and MCR reconstructions, respectively, resulting in an 11-time motion artifact reduction.

CONCLUSION: MCR may improve CCTA assessment in patients by reducing coronary artery motion artifacts, especially in those with elevated HR who cannot receive beta blockers or do not attain the targeted HR.

CLINICAL RELEVANCE STATEMENT: This vendor-specific motion compensation reconstruction may improve coronary computed tomography angiography assessment in patients by reduction of coronary artery motion artifacts, especially in those with elevated various heart rates (HR) who cannot receive beta blockers or do not attain the targeted HR.

KEY POINTS: • Motion artifacts are known to hinder the assessment of coronary arteries on coronary CT angiography (CCTA), leading to more non-diagnostic scans. • This dynamic phantom study shows that motion compensation reconstruction (MCR) reduces motion artifacts at various velocities, which may help to decrease the number of non-diagnostic scans. • MCR in this study showed to reduce motion artifacts 11-fold.

来源:Eur Radiol,查看原文。版权归原作者所有。