文献速递心血管外周血管技术与安全US

校正颈动脉血流时间能否作为左心室射血时间的临床可接受替代血流动力学参数

Is the Corrected Carotid Flow Time a Clinically Acceptable Surrogate Hemodynamic Parameter for the Left Ventricular Ejection Time?
PubMed 文献 · Ultrasound Med Biol · 2024-01-19

本研究在25例接受经导管主动脉瓣置换术的重度主动脉瓣狭窄患者中,用脉冲波多普勒超声分别测量左心室流出道和颈总动脉获得校正左心室射血时间(cLVET)和校正颈动脉血流时间(ccFT),比较两者临床一致性;结果显示cLVET与ccFT相关性好(Bazett ρ=0.74,Wodey ρ=0.81),趋势一致性96%,作者认为ccFT可作为cLVET的替代指标。

要点

  • 在重度主动脉瓣狭窄行TAVR患者中,ccFT与cLVET的Bland-Altman偏倚为-39 ms(Bazett)和-37 ms(Wodey)。
  • cLVET与ccFT的Bazett法相关系数ρ=0.74、Wodey法ρ=0.81,均p<0.001。
  • cLVET与ccFT的趋势一致性为96%,作者认为ccFT可作为cLVET的替代指标。

核心结论

  • 在TAVR患者中,cLVET与ccFT的临床一致性可接受,提示ccFT可作为cLVET的替代指标。

研究设计与样本

  • 研究纳入25例拟行经导管主动脉瓣置换术的重度主动脉瓣狭窄患者,用脉冲波多普勒超声分别从左侧心室流出道和颈总动脉获取cLVET与ccFT,采用Bazett和Wodey方程计算校正值,并分别于TAVR术前(T1)和术后(T2)测量,进行相关性、Bland-Altman和趋势一致性分析。

诊断标准与关键数值

  • Bazett法:cLVET与ccFT相关系数ρ=0.74(p<0.001),偏倚-39 ms,一致性界限上限19 ms、下限-98 ms。
  • Wodey法:cLVET与ccFT相关系数ρ=0.81(p<0.001),偏倚-37 ms,一致性界限上限5 ms、下限-78 ms。
  • cLVET与ccFT的趋势一致性为96%(Bazett和Wodey法均同)。
  • cLVET从T1到T2下降(p<0.001),Bazett法相对降低11%、Wodey法相对降低9%;ccFT从T1到T2下降(p<0.001),Bazett法相对降低12%、Wodey法相对降低10%。

临床应用与阅片要点

  • 在TAVR患者中,ccFT与cLVET的偏倚约-37至-39 ms,报告时需注意ccFT绝对值较cLVET低估。
  • 作者认为ccFT可作为cLVET的替代指标,趋势一致性96%。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: The corrected left ventricular ejection time (cLVET) comprises the phase from aortic valve opening to aortic valve closure corrected for heart rate. As a surrogate measure for cLVET, the corrected carotid flow time (ccFT) has been proposed in previous research. The aim of this study was to assess the clinical agreement between cLVET and ccFT in a dynamic clinical setting.

METHODS: Twenty-five patients with severe aortic valve stenosis (AS) were selected for transcatheter aortic valve replacement (TAVR). The cLVET and ccFT were derived from the left ventricular outflow tract (LVOT) and the common carotid artery (CCA), respectively, using pulsed wave Doppler ultrasound. Bazett's (B) and Wodey's (W) equations were used to calculate cLVET and ccFT. Measurements were performed directly before (T1) and after (T2) TAVR. Correlation, Bland-Altman and concordance analyses were performed.

RESULTS: Corrected LVET decreased from T1 to T2 (p < 0.001), with relative reductions of 11% (B) and 9% (W). Corrected carotid flow time decreased (p < 0.001), with relative reductions of 12% (B) and 10% (W). The correlation between cLVET and ccFT was strong for B (ρ = 0.74, p < 0.001) and W (ρ = 0.81, p < 0.001). The bias was -39 ms (B) and -37 ms (W), and the upper and lower levels of agreement were 19 and -98 ms (B) and 5 and -78 ms (W), respectively. Trending ability between cLVET and ccFT was good (concordance 96%) for both B and W.

CONCLUSION: In TAVR patients, the clinical agreement between cLVET and ccFT was acceptable, indicating that ccFT could serve as a surrogate measure for cLVET.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. van Houte J, Eerdekens R, Manning F, et al. Is the Corrected Carotid Flow Time a Clinically Acceptable Surrogate Hemodynamic Parameter for the Left Ventricular Ejection Time? Ultrasound Med Biol. 2024. PMID: 38242742. doi:10.1016/j.ultrasmedbio.2023.12.013

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。