双源光子计数探测器CT超高分辨率冠状动脉节段图像质量评估
研究纳入60例在双源光子计数探测器CT上行超高分辨率冠状动脉CT血管成像的患者,以0.2毫米层厚和Bv40、Bv44、Bv48、Bv56四种卷积核重建,按18节段模型由两名独立观察者评分。结果显示85%的冠状动脉节段图像质量为优秀或很好,组内相关系数0.88(95%CI:0.87-0.90),运动伪影见于5%节段、堆叠伪影见于7%节段、金属伪影见于1%节段。
要点
- 超高分辨率双源光子计数探测器CT冠状动脉成像在中等辐射剂量下85%节段图像质量为优秀或很好。
- 运动伪影见于45个(5%)节段,堆叠伪影见于60个(7%)节段,金属伪影见于9个(1%)节段。
- 两名观察者间图像质量评分组内相关系数为0.88(95%CI:0.87-0.90)。
核心结论
- 超高分辨率双源光子计数探测器CT冠状动脉CT血管成像在相对较高心率、中等辐射剂量下,85%的冠状动脉节段图像质量为优秀或很好,堆叠伪影有限。
研究设计与样本
- 研究纳入连续临床指征行光子计数探测器CT冠状动脉CT血管成像的患者60例,采用超高分辨率120×0.2毫米准直,以0.2毫米层厚和Bv40、Bv44、Bv48、Bv56卷积核及量子迭代重建4级重建,按18节段模型由两名独立观察者评分。
关键影像表现/征象
- 运动伪影见于45个(5%)冠状动脉节段。
- 堆叠伪影见于60个(7%)冠状动脉节段。
- 金属伪影见于9个(1%)冠状动脉节段。
诊断标准与关键数值
- 两名观察者分别有255个和256个节段因血管缺如或管径过小无法评估(总潜在节段1080个)。
- 两名观察者均将85%的冠状动脉节段评为优秀或很好。
- 图像质量评分的组内相关系数为0.88(95%CI:0.87-0.90)。
- 信噪比为9.3±2.3,对比噪声比为11.7±2.6。
临床应用与阅片要点
- 扫描时平均心率为71±11次/分,中位CTDIvol为19(16-22)mGy,中位DLP为243(198-327)mGy·cm。
- 患者中位年龄62(25th-75th百分位:53-67)岁,45%为男性,中位钙化评分为62(0-217)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: The study is intended to assess the image quality of ultra-high resolution (UHR) coronary computed tomography angiography (CCTA) performed on dual source photon-counting detector CT (PCD-CT).
METHOD: Consecutive patients, who underwent clinically indicated CCTA on PCD-CT (UHR 120x 0.2 mm collimation), were included. CCTA images were reconstructed at 0.2 mm slice thickness with Bv40, Bv44, Bv48 and Bv56 kernels and quantum iterative reconstruction level 4. Contrast-to-noise (CNR) and signal-to-noise ratios (SNR) were quantified from contrast-enhanced blood and subcutaneous adipose tissue. All reconstructions were scored per coronary segment (18-segment model) for presence, image quality, motion artefacts, stack artefacts, plaque presence and composition, and stenosis degree. Image quality was scored by two independent observers.
RESULTS: Sixty patients were included (median age 62 [25th - 75th percentile: 53-67] years, 45% male, median calcium score 62 [0-217]). The mean heart rate during scanning was 71 ± 11 bpm. Median CTDIvol was 19 [16-22]mGy and median DLP 243 [198-327]mGy.cm. The SNR was 9.3 ± 2.3 and the CNR was 11.7 ± 2.6. Of the potential 1080 coronary segments (60 patients x 18 segments), 255/256 (reader1/reader2) segments could not be assessed for being absent or non-evaluable due to size. Both readers scored 85% of the segments as excellent or very good (Intraclass Correlation Coefficient: 0.88 (95% CI: 0.87-0.90). Motion artefacts were present in 45(5%) segments, stack artefacts in 60(7%) segments and metal artefacts in 9(1%) segments.
CONCLUSION: UHR dual-source PCD-CT CCTA provides excellent or very good image quality in 85% of coronary segments at relatively high heart rates at moderate radiation dose with only limited stack artefacts.
引用
- van der Bie J, Sharma SP, van Straten M, et al. Image quality assessment of coronary artery segments using ultra-high resolution dual source photon-counting detector computed tomography. Eur J Radiol. 2024. PMID: 38190778. doi:10.1016/j.ejrad.2023.111282
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。