介入性锥形束CT扫描中运动校正原型图像重建算法的可行性
本研究通过肺部模体实验和导航支气管镜临床CBCT数据,评估一种运动校正原型重建算法。模体实验中算法在多数连续正弦运动及模拟屏气扫描中改善锐利度,27例临床扫描显示运动校正重建边缘锐利度显著优于标准重建(2.81(2.24-6.46)vs. 2.80(2.16-4.75),p=0.003),但整体图像质量评分较低。
要点
- 运动校正重建较标准重建显著提高临床CBCT边缘锐利度(2.81 vs. 2.80,p=0.003)。
- 模体实验中算法在14次连续正弦运动中的13次、7次模拟屏气扫描中的5次改善锐利度。
- 运动校正重建的支气管镜-组织界面和导管-组织界面锐利度评分更高。
核心结论
- 应用该运动校正原型算法于CBCT图像是可行的,算法在诊断性导航支气管镜CBCT扫描中定量和定性均改善了医疗器械的锐利度。
研究设计与样本
- 模体实验使用Xsight肺部模体及含直或弯曲导管的定制插入物,随后将算法应用于导航支气管镜操作的CBCT数据,共分析27例临床扫描,由三名专家定性评估。
关键影像表现/征象
- 运动校正重建较标准重建边缘更锐利(2.81(2.24-6.46)vs. 2.80(2.16-4.75),p=0.003)。
- 定性评估显示运动校正重建的支气管镜-组织界面和导管-组织界面锐利度评分更高。
- 肿瘤边界评分在评估者之间不一致,且新重建的整体图像质量评分较低。
局限性
- 肿瘤边界评分在评估者之间不一致。
- 新重建的整体图像质量评分较低。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: Assess the feasibility of a prototype image reconstruction algorithm in correcting motion artifacts in cone-beam computed tomography (CBCT) scans of interventional instruments in the lung.
MATERIALS AND METHODS: First, phantom experiments were performed to assess the algorithm, using the Xsight lung phantom with custom inserts containing straight or curved catheters. During scanning, the inserts moved in a continuous sinusoidal or breath-hold mimicking pattern, with varying amplitudes and frequencies. Subsequently, the algorithm was applied to CBCT data from navigation bronchoscopy procedures. The algorithm's performance was assessed quantitatively via edge-sharpness measurements and qualitatively by three specialists.
RESULTS: In the phantom study, the algorithm improved sharpness in 13 out of 14 continuous sinusoidal motion and five out of seven breath-hold mimicking scans, with more significant effects at larger motion amplitudes. Analysis of 27 clinical scans showed that the motion corrected reconstructions had significantly sharper edges than standard reconstructions (2.81 (2.24-6.46) vs. 2.80 (2.16-4.75), p = 0.003). These results were consistent with the qualitative assessment, which showed higher scores in the sharpness of bronchoscope-tissue interface and catheter-tissue interface in the motion-corrected reconstructions. However, the tumor demarcation ratings were inconsistent between raters, and the overall image quality of the new reconstructions was rated lower.
CONCLUSION: Our findings suggest that applying the new prototype algorithm for motion correction in CBCT images is feasible. The algorithm improved the sharpness of medical instruments in CBCT scans obtained during diagnostic navigation bronchoscopy procedures, which was demonstrated both quantitatively and qualitatively.
引用
- Spenkelink IM, Heidkamp J, Verhoeven RLJ, et al. Feasibility of a Prototype Image Reconstruction Algorithm for Motion Correction in Interventional Cone-Beam CT Scans. Acad Radiol. 2024. PMID: 38220570. doi:10.1016/j.acra.2023.12.030
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。