能量积分探测器CT与光子计数探测器CT脊髓造影在自发性颅内低压患者脑脊液-静脉瘘检测中的比较

Myelography Using Energy-Integrating Detector CT Versus Photon-Counting Detector CT for Detection of CSF-Venous Fistulas in Patients With Spontaneous Intracranial Hypotension.
PubMed 文献 · AJR Am J Roentgenol · 2024-01-31

该回顾性研究纳入38例自发性颅内低压患者,均行能量积分探测器CT脊髓造影(层厚0.625 mm)和光子计数探测器CT脊髓造影(层厚0.2 mm,超高分辨率模式)。三位盲法阅片者评估图像噪声、神经根袖 discernibility 和整体图像质量,并记录脑脊液-静脉瘘位置。光子计数探测器CT脊髓造影较能量积分探测器CT脊髓造影图像质量显著提高,对明确脑脊液-静脉瘘的敏感度在两位阅片者中显著更高,特异度无显著差异。

要点

  • 光子计数探测器CT脊髓造影(层厚0.2 mm,超高分辨率模式)较能量积分探测器CT脊髓造影(层厚0.625 mm)图像噪声更高、神经根袖 discernibility 和整体图像质量评分更高(所有阅片者 p < .05)。
  • 对明确脑脊液-静脉瘘的敏感度,光子计数探测器CT脊髓造影在阅片者1和2中显著高于能量积分探测器CT脊髓造影(均 p < .05),阅片者3无显著差异(p = .45)。
  • 特异度在两种技术间所有阅片者均无显著差异(均 p > .05)。

核心结论

  • 与能量积分探测器CT脊髓造影相比,光子计数探测器CT脊髓造影图像质量显著改善,对脑脊液-静脉瘘的敏感度显著更高(三位阅片者中的两位),且无显著特异度损失。

研究设计与样本

  • 回顾性研究,纳入38例自发性颅内低压患者(15男、23女,平均年龄55 ± 10岁),均行临床指示的能量积分探测器CT脊髓造影(层厚0.625 mm)和光子计数探测器CT脊髓造影(层厚0.2 mm,超高分辨率模式);三位盲法阅片者随机顺序阅片。

关键影像表现/征象

  • 图像噪声评分:光子计数探测器CT脊髓造影 vs 能量积分探测器CT脊髓造影,阅片者1为69.9 ± 18.5 vs 37.6 ± 15.2,阅片者2为59.5 ± 8.7 vs 49.3 ± 12.7,阅片者3为57.6 ± 13.2 vs 42.1 ± 16.6。
  • 神经根袖 discernibility 评分:光子计数探测器CT脊髓造影 vs 能量积分探测器CT脊髓造影,阅片者1为81.6 ± 21.7 vs 30.4 ± 13.6,阅片者2为83.6 ± 10 vs 70.1 ± 18.9,阅片者3为59.6 ± 13.5 vs 50.5 ± 14.4。
  • 整体图像质量评分:光子计数探测器CT脊髓造影 vs 能量积分探测器CT脊髓造影,阅片者1为83.2 ± 20.0 vs 38.1 ± 13.5,阅片者2为80.1 ± 10.1 vs 72.4 ± 19.8,阅片者3为57.8 ± 11.2 vs 51.9 ± 13.6。

诊断标准与关键数值

  • 明确脑脊液-静脉瘘定义:CT脊髓造影报告中用明确语言描述且衰减值大于70 HU的脑脊液-静脉瘘。
  • 11例患者存在明确脑脊液-静脉瘘。
  • 敏感度:能量积分探测器CT脊髓造影 vs 光子计数探测器CT脊髓造影,阅片者1为45% vs 64%,阅片者2为36% vs 55%,阅片者3为57% vs 55%。
  • 特异度:能量积分探测器CT脊髓造影 vs 光子计数探测器CT脊髓造影,阅片者1为96% vs 85%,阅片者2为100% vs 96%,阅片者3为100% vs 93%。
  • 敏感度比较:光子计数探测器CT脊髓造影显著高于能量积分探测器CT脊髓造影,阅片者1和2均 p < .05;阅片者3无显著差异(p = .45);特异度三位阅片者均无显著差异(均 p > .05)。

临床应用与阅片要点

  • 光子计数探测器CT脊髓造影采用层厚0.2 mm超高分辨率模式,较能量积分探测器CT脊髓造影(层厚0.625 mm)可提高神经根袖 discernibility 和整体图像质量评分。
  • 三位盲法阅片者中两位显示光子计数探测器CT脊髓造影对明确脑脊液-静脉瘘的敏感度显著更高,特异度无显著损失。
  • 研究结果支持光子计数探测器CT脊髓造影可能有助于促进自发性颅内低压的更早诊断和靶向治疗,避免潜在长期诊断检查过程中的高发病率。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND. CSF-venous fistulas (CVFs), which are an increasingly recognized cause of spontaneous intracranial hypotension (SIH), are often diminutive in size and exceedingly difficult to detect by conventional imaging. OBJECTIVE. This purpose of this study was to compare energy-integrating detector (EID) CT myelography and photon-counting detector (PCD) CT myelography in terms of image quality and diagnostic performance for detecting CVFs in patients with SIH. METHODS. This retrospective study included 38 patients (15 men and 23 women; mean age, 55 ± 10 [SD] years) with SIH who underwent both clinically indicated EID CT myelography (slice thickness, 0.625 mm) and PCD CT myelography (slice thickness, 0.2 mm; performed in ultrahigh-resolution mode) to assess for CSF leak. Three blinded radiologists reviewed examinations in random order, assessing image noise, discernibility of spinal nerve root sleeves, and overall image quality (each assessed using a scale of 0-100, with 100 denoting highest quality) and recording locations of the CVFs. Definite CVFs were defined as CVFs described in CT myelography reports using unequivocal language and having an attenuation value greater than 70 HU. RESULTS. For all readers, PCD CT myelography, in comparison with EID CT myelography, showed higher mean image noise (reader 1: 69.9 ± 18.5 [SD] vs 37.6 ± 15.2; reader 2: 59.5 ± 8.7 vs 49.3 ± 12.7; and reader 3: 57.6 ± 13.2 vs 42.1 ± 16.6), higher mean nerve root sleeve discernibility (reader 1: 81.6 ± 21.7 [SD] vs 30.4 ± 13.6; reader 2: 83.6 ± 10 vs 70.1 ± 18.9; and reader 3: 59.6 ± 13.5 vs 50.5 ± 14.4), and higher mean overall image quality (reader 1: 83.2 ± 20.0 [SD] vs 38.1 ± 13.5; reader 2: 80.1 ± 10.1 vs 72.4 ± 19.8; and reader 3: 57.8 ± 11.2 vs 51.9 ± 13.6) (all p < .05). Eleven patients had a definite CVF. Sensitivity and specificity of EID CT myelography and PCD CT myelography for the detection of definite CVF were 45% and 96% versus 64% and 85%, respectively, for reader 1; 36% and 100% versus 55% and 96%, respectively, for reader 2; and 57% and 100% versus 55% and 93%, respectively, for reader 3. The sensitivity was significantly higher for PCD CT myelography than for EID CT myelography for reader 1 and reader 2 (both p < .05) and was not significantly different between the two techniques for reader 3 (p = .45); for all three readers, specificity was not significantly different between the two modalities (all p > .05). CONCLUSION. In comparison with EID CT myelography, PCD CT myelography yielded significantly improved image quality with significantly higher sensitivity for CVFs (for two of three readers), without significant loss of specificity. CLINICAL IMPACT. The findings support a potential role for PCD CT myelography in facilitating earlier diagnosis and targeted treatment of SIH, avoiding high morbidity during potentially prolonged diagnostic workups.