颅内动脉狭窄的CT衍生血流储备分数:基于计算流体力学的初步研究
本研究回顾性纳入19例中重度(≥50%)颅内动脉狭窄患者,以导丝压力测量FFR为参考标准,在狭窄末端(D0)、狭窄远端1 cm(D1)和2 cm(D2)处测量CT-FFR。CT-FFRD1表现出完美预测(AUC 1.000),CT-FFRD1≤0.80检测FFR≤0.80的敏感度0.94、特异度1.00、AUC 0.969,CT-FFRD1与FFR相关性最强(r=0.915)。结论认为狭窄远端1 cm处的CT-FFR与有创FFR结果最可比。
要点
- CT-FFR测量在狭窄末端及远端1 cm、2 cm处均具有优良的观察者内和观察者间一致性(ICC范围0.930-0.992)。
- CT-FFRD1≤0.80检测FFR≤0.80的敏感度为0.94(95%CI 0.68-1.00),特异度为1.00(95%CI 0.31-1.00),AUC为0.969(95%CI 0.772-1.000)。
- 狭窄远端1 cm处测量的CT-FFR与有创FFR相关性最强(r=0.915,p<0.001),平均差异0.02,95%一致性界限-0.16至0.19。
核心结论
- 狭窄远端1 cm处测量的脑CT-FFR与有创FFR结果最可比,提示其可能具有评估颅内动脉狭窄功能相关性的临床应用前景。
研究设计与样本
- 回顾性纳入19例中重度(≥50%)颅内动脉狭窄患者(平均年龄58.6岁±1.9[SD],13例男性),均接受导丝压力测量和术前头部CTA,以有创FFR为参考标准,在狭窄末端(D0)、远端1 cm(D1)和远端2 cm(D2)处测量CT-FFR。
诊断标准与关键数值
- 以有创FFR≤0.80为血流动力学显著狭窄时,CT-FFRD1≤0.80的敏感度为0.94(95%CI 0.68-1.00),特异度为1.00(95%CI 0.31-1.00),AUC为0.969(95%CI 0.772-1.000)。
- 在不同FFR阈值下,CT-FFRD1均表现出完美预测,AUC值为1.000(95%CI 0.824-1.000)。
- CT-FFRD1≤0.75识别FFR≤0.75的AUC为0.964。
临床应用与阅片要点
- 在颅内动脉狭窄的功能评估中,可考虑将CT-FFR测量位置标准化为狭窄远端1 cm处,该位置与有创FFR的一致性和相关性最佳。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: CT-derived fractional flow reserve (CT-FFR) has been widely applied in coronary hemodynamic assessment. However, the feasieablity and standardization measurement in intracranial artery stenosis (ICAS) remains to be defined.
PURPOSE: To demonstrate the feasibility of CT-FFR in ICAS functional assessment and explore the optimal CT-FFR measurement position with invasive FFR as reference standard.
MATERIALS AND METHODS: Nineteen patients (mean age, 58.6 years ± 1.9 [SD]; 13 men) with moderate to severe (≥50 %) ICAS undergoing guidewire-based pressure measurement and preoperative head CT angiography (CTA) were retrospectively enrolled. CT-FFR was measured in the following standard measurement positions, including the end of stenosis (D0), 1 cm distal to the stenosis (D1) and 2 cm distal to the stenosis (D2). Diagnostic performance of CT-FFR was assessed by the area under the curve (AUC) of receiver operating characteristic curves by assuming invasive FFR ≤ 0.80 or 0.75 as hemodynamically significant stenosis.
RESULTS: Excellent intra- and inter-observer agreement (ICC range, 0.930-0.992) was observed for CT-FFR measurement in different positions. Under different FFR thresholds, the diagnostic performance of CT-FFRD1 showed perfect prediction with AUC values of 1.000 (95 % CI: 0.824, 1.000). The sensitivity, specificity and AUC of CT-FFRD1 ≤ 0.80 in detecting FFR ≤ 0.80 was 0.94 (95 % CI: 0.68, 1.00), 1.00 (95 % CI: 0.31, 1.00) and 0.969 (95 % CI: 0.772, 1.000), respectively. Similar performance of CT-FFRD1 ≤ 0.75 was obtained for identifying FFR ≤ 0.75 with the AUC of 0.964. The strongest correlation (r = 0.915, p < 0.001) and agreement (mean difference: 0.02, 95 % limits of agreement: -0.16 to 0.19) were observed between CT-FFRD1 and FFR.
CONCLUSION: Cerebral CT-derived fractional flow reserve (CT-FFR) measured 1 cm distal to stenosis achieved the most comparable results with invasive FFR, which indicated its potentially promising clinical application for evaluating the functional relevance of intracranial artery stenosis.
引用
- Yin Z, Zhou C, Guo J, et al. CT-derived fractional flow reserve in intracranial arterial stenosis: A pilot study based on computational fluid dynamics. Eur J Radiol. 2024. PMID: 38181628. doi:10.1016/j.ejrad.2024.111285
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。