帧率对儿童视频透视吞咽研究中口咽部误吸检出的影响——一项观察性研究
该观察性研究由17名言语语言病理学家对13名婴幼儿/儿童的稀薄液体VFSS录像共2356个吞咽环进行盲法评级,比较15fps与30fps检测误吸的准确性。结果显示15fps与30fps的误吸检测准确性几乎相同(aROC 0.97 vs 0.96),评分者间与评分者内信度良好(ICC分别为0.82与0.89)。作者推荐儿童VFSS评估稀薄液体时使用15fps以降低辐射剂量。
要点
- 儿童VFSS评估稀薄液体误吸时,15fps与30fps的检测准确性几乎相同(aROC 0.97 vs 0.96)。
- 评分者间信度良好(ICC 0.82;95%CI 0.72-0.89),评分者内信度良好(ICC 0.89;95%CI 0.82-0.93)。
- 作者推荐儿童VFSS评估稀薄液体时使用15fps,以遵循ALARA原则降低辐射剂量。
核心结论
- 儿童VFSS评估稀薄液体误吸时,15fps与30fps的检测准确性几乎相同(aROC 0.97 vs 0.96),且15fps辐射剂量低于30fps,作者推荐使用15fps。
研究设计与样本
- 观察性研究,17名言语语言病理学家对13名婴幼儿/儿童的稀薄液体VFSS录像共2356个吞咽环进行盲法评级,其中76个吞咽环在15fps与30fps下随机呈现两次。
诊断标准与关键数值
- 误吸检测准确性:15fps的aROC为0.97(95%CI 0.96-0.97),30fps的aROC为0.96(95%CI 0.96-0.97);评分者间信度ICC为0.82(95%CI 0.72-0.89),评分者内信度ICC为0.89(95%CI 0.82-0.93)。
临床应用与阅片要点
- 作者推荐儿童VFSS评估稀薄液体时使用15fps,以在遵循ALARA原则的同时获得与30fps相似的误吸检测准确性。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: The videofluoroscopic swallow study (VFSS), currently the gold standard for assessing aspiration in children, incurs radiation. Adhering to the ALARA principle is crucial in minimising radiation dose whilst obtaining accurate diagnostic information in children. International adult VFSS guidelines recommend a capture rate of 30 frames per second (fps). Higher capture rates increase radiation yet there is limited evidence on best practice VFSS capture rates in children, particularly on thin fluid consistency-the fastest viscosity with the highest potential for missed aspiration on slower capture rates. We aimed to determine if image acquisition at 30fps versus 15fps alters the accuracy of detecting aspiration when assessing thin fluids during paediatric VFSS.
MATERIALS & METHODS: Seventeen speech language pathologists (SLPs) blindly rated a total of 2,356 swallow loops for the presence/absence of aspiration from VFSS recordings of 13 infants/children drinking thin fluids. 76 swallow loops were randomly presented at 15 versus 30fps, on two occasions. Area under receiver operating curve (aROCs) was used to compare the accuracy of aspiration ratings at 15 versus 30fps compared to a comparison set. The intraclass correlation coefficient (ICC) was used to examine rater reliability.
RESULTS: Accuracy for detecting aspiration was near-identical at 15fps (aROC:0.97; 95%CI:0.96-0.97) and 30fps (0.96; 95%CI 0.96-0.97). Good inter-rater (ICC:0.82; 95%CI:0.72-0.89) and intra-rater reliability among the raters (ICC:0.89; 95%CI:0.82-0.93) was found.
CONCLUSION: Using 15fps in paediatric VFSS when assessing thin fluid consistency aspiration provides a similar detection rate to using 30fps. As 15fps would have a lower radiation dose than 30fps, we recommend using 15fps when undertaking VFSS in children.
CLINICAL RELEVANCE STATEMENT: Adhering to the ALARA principles, a capture rate of 15fps should be used in paediatric VFSS for assessment on thin fluids.