乳腺自动三维超声弹性成像的体内验证研究
研究在临床用自动乳腺容积扫描仪(ABVS)上实现准静态三维弹性成像,采集82例患者共44个病灶(9个癌、23个囊肿、12个其他良性病灶)的射频数据。最大主应变比在恶性病灶(中位数1.7,范围1.0-3.2)与良性病灶(1.0,范围0.6-1.9)间有显著差异;轴向应变比在良恶性间无显著差异。
要点
- 最大主应变比在恶性病灶显著高于良性病灶(中位数1.7 vs 1.0)。
- 轴向应变比在良恶性病灶间无显著差异(0.8 vs 0.6)。
- 三维应变成像可在临床用ABVS上实现并三维显示与分析体内应变图像。
核心结论
- 三维应变成像成功在临床用ABVS上实现,可获得、显示和分析体内三维应变图像;最大主应变比在恶性病灶较良性病灶显著增高。
研究设计与样本
- 体内验证研究,采集82例患者的乳腺容积超声射频数据集(自动换能器抬升前后),病灶由人工标注并用自研应变算法计算应变,每病灶计算轴向和最大主应变比两种应变比。
关键影像表现/征象
- 共检出44个病灶:9个癌、23个囊肿、12个其他良性病灶。
- 最大主应变比:恶性病灶中位数1.7(范围1.0-3.2),良性病灶中位数1.0(范围0.6-1.9)。
- 轴向应变比:良性病灶中位数0.6(范围-12.7至4.9),恶性病灶中位数0.8(范围-3.5至5.1)。
诊断标准与关键数值
- 最大主应变比在恶性与良性病灶间差异有统计学意义(恶性中位数1.7,范围1.0-3.2;良性中位数1.0,范围0.6-1.9)。
- 轴向应变比在良性(中位数0.6,范围-12.7至4.9)与恶性病灶(中位数0.8,范围-3.5至5.1)间无显著差异。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: Studies have indicated that adding 2-D quasi-static elastography to B-mode ultrasound imaging improved the specificity for malignant lesion detection, as malignant lesions are often stiffer (increased strain ratio) compared with benign lesions. This method is limited by its user dependency and so unsuitable for breast screening. To overcome this limitation, we implemented quasi-static elastography in an automated breast volume scanner (ABVS), which is an operator-independent 3-D ultrasound system and is especially useful for screening women with dense breasts. The study aim was to investigate if 3-D quasi-static elastography implemented in a clinically used ABVS can discriminate between benign and malignant breast lesions.
METHODS: Volumetric breast ultrasound radiofrequency data sets of 82 patients were acquired before and after automated transducer lifting. Lesions were annotated and strain was calculated using an in-house-developed strain algorithm. Two strain ratio types were calculated per lesion: using axial and maximal principal strain (i.e., strain in dominant direction).
RESULTS: Forty-four lesions were detected: 9 carcinomas, 23 cysts and 12 other benign lesions. A significant difference was found between malignant (median: 1.7, range: [1.0-3.2]) and benign (1.0, [0.6-1.9]) using maximal principal strain ratios. Axial strain ratio did not reveal a significant difference between benign (0.6, [-12.7 to 4.9]) and malignant lesions (0.8, [-3.5 to 5.1]).
CONCLUSION: Three-dimensional strain imaging was successfully implemented on a clinically used ABVS to obtain, visualize and analyze in vivo strain images in three dimensions. Results revealed that maximal principal strain ratios are significantly increased in malignant compared with benign lesions.