儿童喂养管定位低剂量X线摄影的研发与临床可行性
本研究研发了针对儿童喂养管定位的低剂量喂养管X线摄影(RDFTR),采用0.32 mAs最低设置和气管隆突至髂嵴的视野,经线对板和体模测试后回顾性比较临床剂量。23,789例检查显示RDFTR较腹部、胸部及婴幼儿胸腹部摄影中位DAP降低55%-93%(p<0.001);3,286份报告显示管尖可见率99.1%,异位0.2%在气道、2.3%在食管,0.9%未显示管尖,7例(0.2%)报告气胸或气腹但5例为预期或伪影。
要点
- RDFTR采用0.32 mAs最低设置和气管隆突至髂嵴视野,中位DAP较常规摄影降低55%-93%。
- RDFTR下线对板空间分辨率从2.2线对降至0.6线对,但喂养管尖可见性保留。
- 3,286份报告中管尖可见率99.1%,气道异位0.2%、食管异位2.3%、管尖未显示0.9%。
核心结论
- RDFTR在维持管尖可见性的同时显著降低儿童临时喂养管患者的辐射剂量。
研究设计与样本
- 研发与临床可行性研究,包括线对板和体模测试及回顾性临床使用审查,比较23,789例检查的剂量面积乘积(DAP),并审查3,286份RDFTR报告。
关键影像表现/征象
- 线对板测试显示RDFTR空间分辨率从2.2线对降至0.6线对,但喂养管尖可见性保留。
- 3,286份报告中管尖可见3,256份(99.1%),管尖未显示30份(0.9%)。
- 管尖异位至气道8次(0.2%)、食管74次(2.3%)。
- 气胸或气腹报告7次(0.2%),其中5例为预期或伪影。
诊断标准与关键数值
- RDFTR采用年龄适配腹部摄影,最低mAs设置0.32 mAs,视野介于气管隆突与髂嵴之间。
- 中位DAP比较:RDFTR较腹部摄影降低72%-93%,较胸部摄影降低55%-78%,较婴幼儿胸腹部摄影降低76%-79%(p<0.001)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Temporary feeding tubes are commonly used but may lead to complications if malpositioned. Radiographs are the gold standard for assessing tube position, but clinician concern over radiation risks may curtail their use.
OBJECTIVE: We describe development and use of a reduced dose feeding tube radiograph (RDFTR) targeted for evaluation of feeding tube position.
MATERIALS AND METHODS: Age-based abdominal radiograph was adapted to use the lowest mAs setting of 0.32 mAs with field of view between carina and iliac crests. The protocol was tested in DIGI-13 line-pair plates and anthropomorphic phantoms. Retrospective review of initial clinical use compared dose area product (DAP) for RDFTR and routine abdomen, chest, or infant chest and abdomen. Review of RDFTR reports assessed tube visibility, malpositioning, and incidental critical findings.
RESULTS: Testing through a line-pair phantom showed loss of spatial resolution from 2.2 line pairs to 0.6 line pairs but preserved visibility of feeding tube tip in RDFTR protocol. DAP comparisons across 23,789 exams showed RDFTR reduced median DAP 72-93% compared to abdomen, 55-78% compared to chest, and 76-79% compared to infant chest and abdomen (p<0.001). Review of 3286 reports showed tube was visible in 3256 (99.1%), malpositioned in airway 8 times (0.2%) and in the esophagus 74 times (2.3%). The tip was not visualized in 30 (0.9%). Pneumothorax or pneumoperitoneum was noted seven times (0.2%) but was expected or spurious in five of these cases.
CONCLUSION: RDFTR significantly reduces radiation dose in children with temporary feeding tubes while maintaining visibility of tube tip.
引用
- Kaplan SL, Jalloul M, Akbari E, et al. Development and clinical feasibility of a reduced-dose radiograph in children for feeding tube placement. Pediatr Radiol. 2023. PMID: 38141080. doi:10.1007/s00247-023-05829-w