文献速递心血管CT

光子计数探测器与能量积分探测器CT对冠状动脉管腔狭窄百分比视觉评估的比较

Comparison of Photon-counting Detector and Energy-integrating Detector CT for Visual Estimation of Coronary Percent Luminal Stenosis.
PubMed 文献 · Radiology · 2023-01-01

前瞻性研究纳入25例连续成年门诊患者,先后行临床双源能量积分探测器(EID)冠状动脉CT血管成像和研究性双源光子计数探测器(PCD)CT,PCD CT重建50和100 keV虚拟单能量图像。两名阅片者独立评估并定量测量狭窄百分比,结果显示PCD CT的视觉狭窄百分比估计低于EID CT,而定量测量两者无差异。

要点

  • PCD CT视觉评估的冠状动脉狭窄百分比低于EID CT(阅片者1中位20.6% vs 31.8%,P < .001)。
  • 阅片者2视觉狭窄百分比PCD CT中位6.5% vs EID CT 22.9%(P = .002)。
  • 定量狭窄百分比EID CT与PCD CT无差异(中位差-1.5% [95% CI: -3.0%, 2.5%];P = .51)。

核心结论

  • 与EID CT相比,使用PCD CT导致视觉估计的狭窄百分比降低。

研究设计与样本

  • 前瞻性研究,2022年1月至3月从门诊连续纳入25例成年患者(中位年龄59岁,范围18-78岁;16例男性),先后行临床双源EID冠状动脉CT血管成像和研究性双源PCD CT检查。

诊断标准与关键数值

  • PCD CT使用多能量数据重建50和100 keV虚拟单能量图像。
  • 两名阅片者独立评估左主干、左前降支、右冠状动脉和回旋支最严重狭窄的直径狭窄百分比,对扫描仪类型非盲。
  • EID CT图像上阅片者1和2分别识别出39和32处直径狭窄百分比大于0%的管腔狭窄。
  • 视觉狭窄百分比PCD CT低于EID CT:阅片者1中位20.6%(IQR 8.8%-61.2%)vs 31.8%(IQR 12.9%-69.7%),P < .001;阅片者2中位6.5%(IQR 0.4%-54.1%)vs 22.9%(IQR 1.8%-67.4%),P = .002。
  • 定量狭窄百分比EID CT与PCD CT无差异(中位差-1.5% [95% CI: -3.0%, 2.5%];P = .51)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Background Compared with energy-integrating detector (EID) CT, the improved resolution of photon-counting detector (PCD) CT coupled with high-energy virtual monoenergetic images (VMIs) has been shown to decrease calcium blooming on images in phantoms and cadaveric specimens. Purpose To determine the impact of dual-source PCD CT on visual and quantitative estimation of percent diameter luminal stenosis compared with dual-source EID CT in patients. Materials and Methods This prospective study recruited consecutive adult patients from an outpatient facility between January and March 2022. Study participants underwent clinical dual-source EID coronary CT angiography followed by a research dual-source PCD CT examination. For PCD CT, multienergy data were used to create VMIs at 50 and 100 keV. Two readers independently reviewed EID CT images followed by PCD CT images after a washout period. Readers visually graded the most severe stenosis in terms of percent diameter luminal stenosis for the left main, left anterior descending, right, and circumflex coronary arteries, unblinded to scanner type. Quantitative measures of percent stenosis were made using commercial software. Visual and quantitative estimates of percent stenosis were compared between EID CT and PCD CT using the Wilcoxon signed-rank test. Results A total of 25 participants (median age, 59 years [range, 18-78 years]; 16 male participants) were enrolled. On EID CT images, readers 1 and 2 identified 39 and 32 luminal stenoses, respectively, with a percent diameter luminal stenosis greater than 0%. Visual estimates of percent stenosis were lower on PCD CT images than EID CT images (reader 1: median 20.6% [IQR, 8.8%-61.2%] vs 31.8% [IQR, 12.9%-69.7%], P < .001; reader 2: 6.5% [IQR, 0.4%-54.1%] vs 22.9% [IQR, 1.8%-67.4%], P = .002). No difference was observed between EID CT and PCD CT for quantitative measures of percent stenosis (median difference, -1.5% [95% CI: -3.0%, 2.5%]; P = .51). Conclusion Relative to using EID CT, using PCD CT led to decreases in visual estimates of percent stenosis. © RSNA, 2023 See also the editorial by Murphy and Donnelly in this issue.

来源:Radiology,查看原文。版权归原作者所有。