评估方法对4D flow衍生舒张期二尖瓣口及心肌峰值速度的影响:与超声心动图比较

Impact of the evaluation method on 4D flow-derived diastolic transmitral and myocardial peak velocities: Comparison with echocardiography.
PubMed 文献 · Eur J Radiol · 2023-12-05

前瞻性研究纳入60例无心血管症状受试者,比较3种MR 4D flow评估二尖瓣口舒张早期(E)和晚期(A)峰值速度及比值的方法与超声心动图的一致性。结果显示,从二尖瓣口流入容积取最大速度的方法所测E和E/A与超声无显著偏倚,E/e'及各心肌e'也与超声相关且无显著偏倚。

要点

  • 从二尖瓣口流入容积取最大速度测E和E/A与超声无显著偏倚,可直接沿用超声阈值。
  • 短轴切面二尖瓣尖端法和四腔心切面法均显著低估E和A(p<0.001)。
  • 4D flow所测室间隔、侧壁及平均e'与超声相关(r=0.64-0.81)且无显著偏倚。

核心结论

  • 当二尖瓣口速度从二尖瓣口流入容积取最大值评估时,MR 4D flow可精确准确评估二尖瓣口和心肌峰值速度以表征左室舒张功能,与超声心动图无显著偏倚,从而可使用已验证的超声心动图阈值。

研究设计与样本

  • 前瞻性研究,纳入60例无心血管疾病症状的受试者,行超声心动图和3 T非增强MR 4D flow心脏成像,以超声心动图为参考标准。

诊断标准与关键数值

  • 4D flow所测E、A及E/A与超声的相关性r分别为0.65-0.73、0.75-0.83和0.74-0.86。
  • 短轴法和四腔心法测E和A均显著低于超声(p<0.001);最大速度法测E(p=0.13)和E/A(p=0.07)无显著偏倚。
  • 4D flow测室间隔、侧壁及平均e'与超声相关r=0.64-0.81,最大速度法衍生E/e'与超声无显著偏倚(p=0.26-0.54)。

临床应用与阅片要点

  • 评估左室舒张功能时,二尖瓣口E和A峰值速度应从二尖瓣口流入容积取最大值,以避免显著低估。
  • 按最大速度法测得的E和E/A与超声无显著偏倚,可直接参照已验证的超声心动图阈值进行解读。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To compare agreement of different evaluation methods of magnetic resonance (MR) 4D flow-derived diastolic transmitral and myocardial peak velocities as well as their ratios, using echocardiography as reference.

METHODS: In this prospective study, 60 subjects without symptoms of cardiovascular disease underwent echocardiography and non-contrast 3 T MR 4D flow imaging of the heart. Early- (E) and late-diastolic (A) transmitral peak filling velocities were evaluated from 4D flow data using three different strategies: 1) at the mitral valve tips in short-axis orientation (SA-method), 2) between the mitral valve tips in 4-chamber orientation (4-chamber-method), and 3) as maximal velocities in the transmitral inflow volume (max-velocity-method). Septal, lateral and average early-diastolic myocardial peak velocities (e') were derived from the myocardial tissue in the vicinity of the mitral valve. 4D flow parameters were compared with echocardiography by correlation and Bland-Altman analysis.

RESULTS: All 4D flow-derived E, A and E/A values correlated with echocardiography (r = 0.65-0.73, 0.75-0.83 and 0.74-0.86, respectively). While the SA- and 4-chamber-methods substantially underestimated E and A compared to echocardiography (p < 0.001), the max-velocity-method provided E (p = 0.13) and E/A (p = 0.07) without significant bias. Septal, lateral and average e' from 4D flow as well as the max-velocity-method-derived E/e' correlated with echocardiographic measurements (r = 0.64-0.81) and showed no significant bias (p = 0.26-0.54).

CONCLUSION: MR 4D flow imaging allows precise and accurate evaluation of transmitral and myocardial peak velocities for characterization of LV diastolic function without significant bias to echocardiography, when transmitral velocities are assessed from the transmitral inflow volume. This enables the use of validated echocardiography threshold values.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Reiter C, Reiter G, Kräuter C, et al. Impact of the evaluation method on 4D flow-derived diastolic transmitral and myocardial peak velocities: Comparison with echocardiography. Eur J Radiol. 2023. PMID: 38071909. doi:10.1016/j.ejrad.2023.111247

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。