用于高帧率三维心腔内超声的超声矩阵换能器
本文报道了一种用于高帧率三维心腔内超声的超声矩阵换能器原型的研制。该矩阵阵列由16×18个锆钛酸铅元件组成,元件间距160 µm×160 µm,集成于专用集成电路上,采用子阵波束形成与数字化接收及时分复用与脉冲幅度调制数据传输,使线缆数量减少18倍;成像方案采用7个扇形发散发射波束,脉冲重复频率7.7 kHz。原型换能器在5 cm处每元件发射效率28 Pa/V,发射带宽60%,接收动态范围80 dB,每元件最小可检测压力10 Pa,元件良率98%,帧率可达1000 volumes/s,覆盖体积70°×70°×10 cm。结论认为这些先进成像能力有潜力支持复杂介入手术,并实现心脏全容积血流、组织及机电波追踪。
要点
- 矩阵换能器原型元件良率98%,帧率可达1000 volumes/s,覆盖体积70°×70°×10 cm。
- 发射效率28 Pa/V at 5 cm per element,发射带宽60%,接收动态范围80 dB。
- 采用7个扇形发散发射波束、脉冲重复频率7.7 kHz实现高帧率三维成像。
核心结论
- 该矩阵换能器原型具备高帧率三维心腔内超声成像能力,有潜力支持复杂介入手术及心脏全容积血流、组织与机电波追踪。
研究设计与样本
- 研制超声矩阵换能器原型,矩阵阵列由16×18个锆钛酸铅元件组成,元件间距160 µm×160 µm,集成于专用集成电路上,采用子阵波束形成与数字化接收,结合时分复用与脉冲幅度调制数据传输实现18倍线缆减少;成像采用7个扇形发散发射波束,脉冲重复频率7.7 kHz;对原型性能进行表征并验证功能。
诊断标准与关键数值
- 发射效率28 Pa/V at 5 cm per element,发射带宽60%。
- 接收动态范围80 dB,每元件最小可检测压力10 Pa。
- 元件良率98%。
- 帧率可达1000 volumes/s,覆盖体积70°×70°×10 cm。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: Described here is the development of an ultrasound matrix transducer prototype for high-frame-rate 3-D intra-cardiac echocardiography.
METHODS: The matrix array consists of 16 × 18 lead zirconate titanate elements with a pitch of 160 µm × 160 µm built on top of an application-specific integrated circuit that generates transmission signals and digitizes the received signals. To reduce the number of cables in the catheter to a feasible number, we implement subarray beamforming and digitization in receive and use a combination of time-division multiplexing and pulse amplitude modulation data transmission, achieving an 18-fold reduction. The proposed imaging scheme employs seven fan-shaped diverging transmit beams operating at a pulse repetition frequency of 7.7 kHz to obtain a high frame rate. The performance of the prototype is characterized, and its functionality is fully verified.
RESULTS: The transducer exhibits a transmit efficiency of 28 Pa/V at 5 cm per element and a bandwidth of 60% in transmission. In receive, a dynamic range of 80 dB is measured with a minimum detectable pressure of 10 Pa per element. The element yield of the prototype is 98%, indicating the efficacy of the manufacturing process. The transducer is capable of imaging at a frame rate of up to 1000 volumes/s and is intended to cover a volume of 70° × 70° × 10 cm.
CONCLUSION: These advanced imaging capabilities have the potential to support complex interventional procedures and enable full-volumetric flow, tissue, and electromechanical wave tracking in the heart.
引用
- Dos Santos DS, Ossenkoppele B, Hopf YM, et al. An Ultrasound Matrix Transducer for High-Frame-Rate 3-D Intra-cardiac Echocardiography. Ultrasound Med Biol. 2023. PMID: 38036310. doi:10.1016/j.ultrasmedbio.2023.11.001
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。