文献速递神经头颈外周血管CT

暗血成像在头颈部CT血管造影中的临床应用:对图像质量和斑块可见性的影响

Clinical Application of Dark-blood Imaging in Head and Neck CT Angiography: Effect on Image Quality and Plaque Visibility.
PubMed 文献 · Acad Radiol · 2023-12-01

回顾性纳入2021年8月至2023年3月行头颈部三期CT血管造影的患者43例(54个钙化斑块、34个非钙化斑块),采用对比增强提升(CE-boost)技术后处理生成暗血图像。与常规三期相比,暗血图像在钙化和非钙化斑块的图像质量及主观可见性评分上显著提高,斑块负荷的观察者间一致性也更高(ICC 0.870 vs 0.729)。

要点

  • 暗血图像对钙化和非钙化斑块的可见性主观评分均最高(均p<0.001),两名放射科医师评分一致性良好(kappa均>0.7)。
  • 暗血图像钙化斑块CNRplaque-lumen为219.79±159.20、CNRplaque-muscle为181.23±112.12,非钙化斑块分别为30.30±29.11和6.28±4.75,均高于平扫及延迟期(均p<0.001)。
  • 斑块负荷的观察者间一致性暗血图像ICC=0.870,高于动脉期ICC=0.729。

核心结论

  • 与常规三期头颈部CT血管造影相比,CE-boost衍生的暗血成像在钙化和非钙化斑块评估中显著改善了图像质量和可见性。

研究设计与样本

  • 回顾性纳入2021年8月至2023年3月行头颈部三期CT血管造影(平扫期、动脉期、延迟期)的患者43例,平均年龄67.23±10.81岁(范围43-85岁),男性占64.7%。

关键影像表现/征象

  • 暗血图像由对比增强提升(CE-boost)技术后处理生成,用于评估钙化斑块和非钙化斑块的CT值、图像噪声、CNR和SNR。
  • 暗血图像钙化斑块SNR与其他期相比相等或略低(p>0.05或p=0.02)。
  • 暗血图像非钙化斑块SNR较动脉期明显升高(5.56±3.71 vs. 4.23±3.56,p=0.02),与平扫期和延迟期相比无显著差异(p>0.05)。

诊断标准与关键数值

  • 两名经验丰富的放射科医师独立按钙化斑块可见性、非钙化斑块可见性、诊断信心和整体图像质量4项特征以5分制(5分最佳、1分最差)评分,暗血图像各项评分分别为4.99±0.07、4.62±0.48、4.74±0.36和4.81±0.34,均最高(p<0.001)。
  • 暗血图像与动脉期图像在识别狭窄率上一致性良好(p<0.001),暗血图像对狭窄率的诊断准确性以AUC评估。
  • 暗血图像与动脉期图像斑块负荷观察者间一致性分别为ICC=0.870与ICC=0.729。

临床应用与阅片要点

  • 暗血图像可提高钙化斑块和非钙化斑块的可见性及诊断信心,并提高斑块负荷测量的一致性。
  • 暗血图像与动脉期在狭窄率识别上一致性良好,可用于狭窄率评估。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: The aim of this study was to explore the potential of a newly developed dark-blood imaging technique to improve image quality and plaque visibility in head and neck computed tomography (CT) angiography.

MATERIALS AND METHODS: Patients who underwent triphasic head and neck CT angiography scans from August 2021 to March 2023 were retrospectively enrolled (mean age 67.23 ± 10.81 [SD] years, range 43-85 years, 64.7% male). The CT protocol consists of pre-contrast, arterial and delayed phases. Dark-blood images were postprocessed with the contrast-enhancement boost (CE-boost) technique. The quantitative assessment involved evaluating the CT value, image noise, contrast-to-noise ratio (CNR), and signal-to-noise ratio (SNR) of calcified plaque and non-calcified plaque. The plaque CNR relative to the vessel lumen (CNRplaque-lumen), vessel wall (CNRplaque-wall), and adjacent muscle (CNRplaque-muscle) was respectively calculated. Two experienced radiologists independently evaluated the CT images (5, best; 1, worst) by four characteristics including calcified plaque visibility, non-calcified plaque visibility, diagnostic confidence, and overall image quality. Inter-rater variability was also evaluated. The artery stenosis rate and plaque burden on dark-blood images were measured and compared with arterial phases. The intraclass correlation coefficient (ICC) was used for consistency analysis. The diagnostic accuracy of dark-blood images for the stenosis rate was evaluated by the area under the curve (AUC).

RESULTS: A total of 43 patients with 54 calcified plaques and 34 non-calcified plaques were assessed in this study. When compared with pre-contrast and delayed phase, dark-blood images yielded significantly higher CNRplaque-lumen and CNRplaque-muscle of calcified (219.79 ± 159.20 and 181.23 ± 112.12, respectively) and non-calcified (30.30 ± 29.11 and 6.28 ± 4.75, respectively) plaques (all p < 0.001). Calcified plaque SNR of dark-blood showed equal or slightly lower than other phases (p > 0.05 or p = 0.02). A major increase was observed in the non-calcified plaque SNR of dark-blood compared to the arterial phase (5.56 ± 3.71 vs. 4.23 ± 3.56, p = 0.02), although there were no apparent differences compared to pre-contrast and delayed phases (p > 0.05). In subjective analyzes, the calcified plaque visibility (4.99 ± 0.07), non-calcified plaque visibility (4.62 ± 0.48), overall image quality (4.81 ± 0.34), and diagnostic confidence (4.74 ± 0.36) in dark-blood images dominated the highest scores (p < 0.001). The subjective scores of radiologists exhibited good consistency (all kappa value>0.7). The dark-blood image and the arterial phase image exhibited good consistency in identifying the stenosis rate (p < 0.001). In the evaluation of plaque burden, the interobserver agreement for dark-blood images was higher compared to arterial phase images (ICC = 0.870 vs. 0.729).

CONCLUSIONS: Compared to conventional triphasic head and neck CT angiography, the CE-boost derived dark-blood imaging demonstrated a significant improvement in image quality and visibility for both calcified and non-calcified plaque assessment.

来源:Acad Radiol,查看原文。版权归原作者所有。