光子计数探测器CT中的金属伪影减少:伪影减少技术的定量评估

Metal Artifact Reduction in Photon-Counting Detector CT: Quantitative Evaluation of Artifact Reduction Techniques.
PubMed 文献 · Invest Radiol · 2023-10-09

在临床双源光子计数探测器CT上扫描多能量体模,比较不同采集与重建设置对金属伪影的定量影响。结果显示iMAR联合110 keV虚拟单能图像伪影减少最强,管电压和重建核影响有限。

要点

  • iMAR联合110 keV虚拟单能图像伪影最少(1.2%)。
  • 单独iMAR将伪影从10.5%降至1.4%,单独VMI 110 keV为2.6%。
  • 管电压(120 kV 6.6% vs 140 kV 5.2%)和重建核(Br36 5.5% vs Br56 6.4%)影响有限。

核心结论

  • 在光子计数探测器CT中,iMAR联合110 keV虚拟单能图像在所评估方案中伪影减少最强,而重建核和管电压的影响有限。

研究设计与样本

  • 在临床双源光子计数探测器CT(NAEOTOM Alpha;Siemens Healthcare GmbH)上扫描多能量体模,含4种中心插入物:水等效塑料、铝、钢和钛;采集条件为120 kVp和140 kVp(CTDI vol 10 mGy),重建为虚拟单能图像(110-150 keV)、T3D和标准重建「none」(70 keV VMI),使用不同重建核(Br36、Br56)及有无iMAR;采用重复测量方差分析进行统计。

诊断标准与关键数值

  • 金属伪影相对百分比:钢插入物平均12.6%,钛4.2%,铝1.0%;iMAR有vs无:1.4% vs 10.5%(P < 0.001);VMI 110 keV 2.6%至150 keV 3.3%,T3D 11.0%,none 16.0%(P < 0.001);iMAR联合VMI 110 keV为1.2%;管电压120 kV 6.6% vs 140 kV 5.2%(P = 0.33);重建核Br36 5.5% vs Br56 6.4%(P = 0.17)。

临床应用与阅片要点

  • 作者结论认为,在光子计数探测器CT中,iMAR联合110 keV虚拟单能图像在所评估方案中伪影减少最强,而重建核和管电压的影响有限。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: With the introduction of clinical photon-counting detector computed tomography (PCD-CT) and its novel reconstruction techniques, a quantitative investigation of different acquisition and reconstruction settings is necessary to optimize clinical acquisition protocols for metal artifact reduction.

MATERIALS AND METHODS: A multienergy phantom was scanned on a clinical dual-source PCD-CT (NAEOTOM Alpha; Siemens Healthcare GmbH) with 4 different central inserts: water-equivalent plastic, aluminum, steel, and titanium. Acquisitions were performed at 120 kVp and 140 kVp (CTDI vol 10 mGy) and reconstructed as virtual monoenergetic images (VMIs; 110-150 keV), as T3D, and with the standard reconstruction "none" (70 keV VMI) using different reconstruction kernels (Br36, Br56) and with as well as without iterative metal artifact reduction (iMAR). Metal artifacts were quantified, calculating relative percentages of metal artifacts. Mean CT numbers of an adjacent water-equivalent insert and different tissue-equivalent inserts were evaluated, and eccentricity of metal rods was measured. Repeated-measures analysis of variance was performed for statistical analysis.

RESULTS: Metal artifacts were most prevalent for the steel insert (12.6% average artifacts), followed by titanium (4.2%) and aluminum (1.0%). The strongest metal artifact reduction was noted for iMAR (with iMAR: 1.4%, without iMAR: 10.5%; P < 0.001) or VMI (VMI: 110 keV 2.6% to 150 keV 3.3%, T3D: 11.0%, and none: 16.0%; P < 0.001) individually, with best results when combining iMAR and VMI at 110 keV (1.2%). Changing acquisition tube potential (120 kV: 6.6%, 140 kV: 5.2%; P = 0.33) or reconstruction kernel (Br36: 5.5%, Br56: 6.4%; P = 0.17) was less effective. Mean CT numbers and standard deviations were significantly affected by iMAR (with iMAR: -3.0 ± 21.5 HU, without iMAR: -8.5 ± 24.3 HU; P < 0.001), VMI (VMI: 110 keV -3.6 ± 21.6 HU to 150 keV -1.4 ± 21.2 HU, T3D: -11.7 ± 23.8 HU, and none: -16.9 ± 29.8 HU; P < 0.001), tube potential (120 kV: -4.7 ± 22.8 HU, 140 kV: -6.8 ± 23.0 HU; P = 0.03), and reconstruction kernel (Br36: -5.5 ± 14.2 HU, Br56: -6.8 ± 23.0 HU; P < 0.001). Both iMAR and VMI improved quantitative CT number accuracy and metal rod eccentricity for the steel rod, but iMAR was of limited effectiveness for the aluminum rod.

CONCLUSIONS: For metal artifact reduction in PCD-CT, a combination of iMAR and VMI at 110 keV demonstrated the strongest artifact reduction of the evaluated options, whereas the impact of reconstruction kernel and tube potential was limited.

来源:Invest Radiol,查看原文。版权归原作者所有。

引用

  1. Skornitzke S, Mergen V, Biederer J, et al. Metal Artifact Reduction in Photon-Counting Detector CT: Quantitative Evaluation of Artifact Reduction Techniques. Invest Radiol. 2023. PMID: 37812482. doi:10.1097/RLI.0000000000001036

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。