文献速递心血管儿科影像MR

高度加速压缩感知四维血流MRI在先天性与获得性心脏病中的应用:儿童及青年主动脉与主肺动脉血流参数与传统四维血流MRI的比较

Highly accelerated compressed sensing 4D flow MRI in congenital and acquired heart disease: comparison of aorta and main pulmonary artery flow parameters with conventional 4D flow MRI in children and young adults.
PubMed 文献 · Pediatr Radiol · 2023-10-26

本研究对27例儿童及青年患者行传统与压缩感知四维血流MRI,比较主动脉与主肺动脉血流参数。压缩感知使平均扫描时间缩短59%(3.4 vs 8.2分钟),净流量与前向流量等效,但收缩期峰值流速被显著低估。

要点

  • 压缩感知四维血流MRI平均扫描时间3.4分钟,较传统8.2分钟缩短59%。
  • 主动脉与主肺动脉净流量、前向流量在两种方法间无显著差异。
  • 压缩感知四维血流MRI显著低估主动脉与主肺动脉收缩期峰值流速。

核心结论

  • 在主动脉与主肺动脉,压缩感知四维血流MRI提供的净流量与前向流量值与传统四维血流MRI等效,但低估收缩期峰值流速;通过缩短扫描时间,可能减少麻醉需求并提高扫描仪产出而不显著损害数据完整性。

研究设计与样本

  • 共27例患者(中位年龄13岁,IQR 9.5岁)在知情同意后接受传统与压缩感知心血管四维血流MRI检查;传统四维血流采用并行成像加速因子2,压缩感知四维血流加速因子7.7;在升主动脉与主肺动脉放置感兴趣区比较血流参数;采用配对Student t检验、Wilcoxon符号秩检验、Bland-Altman图与组内相关系数分析,P<0.05为统计学显著。

诊断标准与关键数值

  • 升主动脉净流量:压缩感知47 ml/beat vs 传统49 ml/beat(P=0.28)。
  • 升主动脉前向流量:压缩感知49 ml/beat vs 传统50 ml/beat(P=0.07)。
  • 主肺动脉净流量:压缩感知49 ml/beat vs 传统51 ml/beat(P=0.18)。
  • 主肺动脉前向流量:压缩感知50 ml/beat vs 传统55 ml/beat(P=0.07)。
  • 收缩期峰值流速:升主动脉压缩感知114 cm/s vs 传统128 cm/s(P<0.001),主肺动脉压缩感知106 cm/s vs 传统112 cm/s(P=0.02),压缩感知显著低估。

临床应用与阅片要点

  • 压缩感知四维血流MRI可缩短扫描时间,可能减少儿童麻醉需求并提高扫描仪产出。
  • 采用压缩感知四维血流MRI时,净流量与前向流量可等效判读,但收缩期峰值流速会被低估。
  • 原文结论未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Four-dimensional flow (4D flow) MRI has become a clinically utilized cardiovascular flow assessment tool. However, scans can be lengthy and may require anesthesia in younger children. Adding compressed sensing can decrease scan time, but its impact on hemodynamic data accuracy needs additional assessment.

OBJECTIVE: To compare 4D flow hemodynamics acquired with and without compressed sensing.

MATERIALS AND METHODS: Twenty-seven patients (median age: 13 [IQR: 9.5] years) underwent conventional and compressed sensing cardiovascular 4D flow following informed consent. Conventional 4D flow was performed using parallel imaging and an acceleration factor of 2. Compressed sensing 4D flow was performed with an acceleration factor of 7.7. Regions of interest were placed to compare flow parameters in the ascending aorta and main pulmonary artery. Paired Student's t-tests, Wilcoxon signed-rank tests, Bland-Altman plots, and intraclass correlation coefficients were conducted. A P-value of < 0.05 was considered statistically significant.

RESULTS: Mean scan acquisition time was reduced by 59% using compressed sensing (3.4 vs. 8.2 min, P < 0.001). Flow quantification was similar for compressed sensing and conventional 4D flow for the ascending aorta net flow: 47 vs. 49 ml/beat (P = 0.28); forward flow: 49 vs. 50 ml/beat (P = 0.07), and main pulmonary artery net flow: 49 vs. 51 ml/beat (P = 0.18); forward flow: 50 vs. 55 ml/beat (P = 0.07). Peak systolic velocity was significantly underestimated by compressed sensing 4D flow in the ascending aorta: 114 vs. 128 cm/s (P < 0.001) and main pulmonary artery: 106 vs. 112 cm/s (P = 0.02).

CONCLUSION: For both the aorta and main pulmonary artery, compressed sensing 4D flow provided equivalent net and forward flow values compared to conventional 4D flow but underestimated peak systolic velocity. By reducing scan time, compressed sensing 4D flow may decrease the need for anesthesia and increase scanner output without significantly compromising data integrity.

来源:Pediatr Radiol,查看原文。版权归原作者所有。

引用

  1. Sodhi A, Markl M, Popescu AR, et al. Highly accelerated compressed sensing 4D flow MRI in congenital and acquired heart disease: comparison of aorta and main pulmonary artery flow parameters with conventional 4D flow MRI in children and young adults. Pediatr Radiol. 2023. PMID: 37882844. doi:10.1007/s00247-023-05788-2

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。