光谱光子计数CT:使用含金属牛骨标本的图像质量评估

Spectral photon-counting CT: Image quality evaluation using a metal-containing bovine bone specimen.
PubMed 文献 · Eur J Radiol · 2023-09-23

本研究用含不锈钢板和螺钉的牛膝关节标本,在光子计数探测器CT(PCD-CT)和能量积分探测器CT(EID-CT)上比较图像质量和金属伪影严重程度。四名放射科医师评分显示,PCD-CT 140 kV 150 keV图像评分最高,显著优于EID-CT Sn150 kV图像;PCD-CT 72 keV图像评分显著低于其他所有图像。结论认为PCD-CT 140 kV 150 keV图像质量最佳,PCD-CT是减少金属伪影的有前景方法。

要点

  • PCD-CT 140 kV 150 keV图像评分最高,显著优于EID-CT Sn150 kV图像。
  • PCD-CT 72 keV图像评分显著低于其他所有图像,HU值变异较大。
  • ROI分析显示两种扫描仪在伪影严重程度方面无 large difference。

核心结论

  • PCD-CT 140 kV 150 keV含金属牛膝关节标本图像提供最佳图像质量,优于或等同于最佳EID-CT图像,即使没有锡滤过和金属伪影减少算法的 presumed 优势;PCD-CT是减少金属伪影的有前景方法。

研究设计与样本

  • 研究论文;使用含不锈钢板和螺钉的牛膝关节标本,在全身研究型PCD-CT上以120 kV和140 kV成像,在EID双源CT上以Sn150 kV和80/Sn150 kV成像;比较PCD-CT虚拟单能72和150 keV图像与EID-CT有无iMAR处理的图像;四名放射科医师评分;使用Friedman检验和Wilcoxon符号秩检验及Bonferroni校正,P≤0.0001为统计学显著。

关键影像表现/征象

  • PCD-CT 140 kV 150 keV图像评分最高。
  • PCD-CT 72 keV图像评分显著低于其他所有图像。
  • 120 kV和72 keV图像的HU值变异较大。
  • ROI分析显示扫描仪间伪影严重程度无 large difference。

临床应用与阅片要点

  • 作者结论认为PCD-CT 140 kV 150 keV图像质量最佳,PCD-CT是减少金属伪影的有前景方法。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To find the optimal imaging parameters for a photon-counting detector CT (PCD-CT) and to compare it to an energy-integrating detector CT (EID-CT) in terms of image quality and metal artefact severity using a metal-containing bovine knee specimen.

METHODS: A bovine knee with a stainless-steel plate and screws was imaged in a whole-body research PCD-CT at 120 kV and 140 kV and in an EID dual-source CT (DSCT) at Sn150 kV and 80/Sn150 kV. PCD-CT virtual monoenergetic 72 and 150 keV images and EID-CT images processed with and without metal artefact reduction algorithms (iMAR) were compared. Four radiologists rated the visualisation of bony structures and metal artefact severity. The Friedman test and Wilcoxon signed-rank test with Bonferroni's correction were used. P-values of ≤ 0.0001 were considered statistically significant. Distributions of HU values of regions of interest (ROIs) in artefact-affected areas were analysed.

RESULTS: PCD-CT 140 kV 150 keV images received the highest scores and were significantly better than EID-CT Sn150 kV images. PCD-CT 72 keV images were rated significantly lower than all the others. HU-value variation was larger in the 120 kV and the 72 keV images. The ROI analysis revealed no large difference between scanners regarding artefact severity.

CONCLUSION: PCD-CT 140 kV 150 keV images of a metal-containing bovine knee specimen provided the best image quality. They were superior to, or as good as, the best EID-CT images; even without the presumed advantage of tin filter and metal artefact reduction algorithms. PCD-CT is a promising method for reducing metal artefacts.

来源:Eur J Radiol,查看原文。版权归原作者所有。